This job aid describes how to view sample results on the Joblist and Browser [Reaction Curve] screens of the CN-3000 and CN-6000 Systems.

CN-3000 and CN-6000 Systems SIEMENS Healthineers Review Sample Results Job Aid Assay parameter Num Dilution ratio Analysis status Fbg sec 1 1/1 19.9 View Sample Results from the Joblist Fbg C. 1 1/1 81.7 Sample details and error flags 1. Press Joblist on the Menu PT sec 1 1/1 19.8 2. Select a sample from the Joblist 3. Press to display the Details display area PT R. 1 1/1 * 1.80 4. Review analysis results and any associated error flags • Red – analysis result cannot be calculated • Yellow – analysis result requires confirmation • Grey – another error has occurred, or there is no error 4000 5. Select an assay parameter to display the reaction curve below 3000 6. If an error has occurred, the error with the 2000 Selected assay highest priority will display below the reaction curve reaction curve and the total number of errors 1000 will display to the right of the error mark 0 O 10 20 30 40 7. Double click the reaction curve to display the Browser [Reaction Curves] screen OR press Vial QC has not performed. 1 Total number of errors Browser in the Toolbar 1 Highest priority Effective Date: 18-Jun-2021 | HOOD05162003199986 error message Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Review Sample Results Job Aid Assessment Flags Masked Results Display Meaning Display Priority Display Meaning Analysis result could not be obtained due to • Analysis result error **** * • Replication difference error . an error * • MDA slope ratio error 1 • Low-reliability flag related to Calculated parameter and mean value could - - - - - . not be calculated HIL check Dilution analysis Calibration curve not validated, preventing XXXX.X d performed 2 calculated parameter The value below/above ++++.+ Number of displayed digits exceeded < / > report limit OR calibration 3 curve range exceeded / / / /. / Error occurred in the mean calculation due to + / - The value above/below an error in one of the duplicate analysis results mark limit 4 2 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Review Sample Results Job Aid View Sample Results from the Browser X [Reaction Curves] Screen Menu Reagent Calib. curve QC chart Order calib. Order QC Order STAT Joblist Maintenance Shutdown Edit Validate CC/QC Previous Next Output Search Inquiry Close Error Micro CP Remeasure 1945 B ............................ Lip Vol A HOST # 000001-5 ........... Validated Hem @ 2020/05/20 15:58 1. Press Joblist on the Menu 4 A APTT AT INN Fbg PT 2. Select a sample from the Joblist Dilution ratio 1/1 3. Press Browser OR double click on the sample A 1st Early Reaction Error: Slow Reaction 4. View the reaction curve shapes for each analysis 4000-1 Assay Value Units APTT sec ***** * 0008.0128.0001 sec 3600- This error is detected when an abnormal reaction is indicated in parameter by viewing each tab APTT R. **** ** the early stage of a coagulation reaction and the reaction speed is 3200- slow in the vicinity of coagulation detection point (50%). Clotting point An error icon 2800-| Target assay: APTT • is displayed on the tab of Analysis range: Measurement time (main) 0 to 180 sec, Measurement time (sub) 0 to 360 sec analysis parameters that contain errors 2400- 2000- 5 Explanation 5. Review errors in the Error information display area 1600- Error info: A 4 Code/Message Countermeasures 6. Select an error code and press Show error 1200 0008.0128.0001 description, if applicable 800- Early reaction error: Slow reaction 400- A 0008.0128.0016 Onboard error support, including error Early reaction error: Early % • 20 40 60 80 100 120 A 0032.0002.0000 explanation and countermeasures, is displayed in ( secl Flat curve ? the Details display area Close error explanation • Select Close error explanation to return to the original Details display area CN-6000 II Ready 6 Stop Procedure continued on the next slide 3 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Review Sample Results Job Aid View Sample Results from the Browser [Reaction Curves] Screen (continued) Q X Reagent Calib curve QC chart Order calib Order QC Order STAT Joblist Margorance Valdice Predous Next Search Close Complete Micro CP Remeasure 0 9M Validated A HOST # 000018-3 7. Press the following tabs in the Details display area: O 19/06/2020 16:25 Evaluation info. PT INN APTT FS Fbg DD • Displays the data used for calculating the Dilution ratio 1/2 • Redilution 1st A Initial 1st, Redilution 1st Evaluation Info Measurement info Evaluation data analysis results Asun Value Urats tritoemation Value Measurement info. Fbg sec d 7.6 sec bH 3339 • Fbg C. 4.93 g/L BH point time [sec] 1.9 Displays traceability information, including the 3200- dH 344 • End point time [sec] 32.7 results of the HIL check and sample volume Coag% 50 2400- Wave 405 check dOD 0.0472 1600- Error info: A • Evaluation data* Code Message 0000.0000.0000 • Displays 1%-100% time for the reaction curve 800- Dilution ratio was changed. * Only appears when the evaluation algorithm is Show error description the percentage detection method. CN-6000 Ready 3D 7 4 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Review Sample Results Job Aid View HI*L Check and Tube Volume Check 1 6 Information AAA Menu Reagent Calib. curve QC chart Order calib. Order QC Order STAT Joblist Maintenance Shutdown Graph Browser Waveform profile Edit data Validate Font size Customize Output Delete 1. Press Joblist on the Menu All sample Specify date 5 (2/2) 2. Verify the Sample info tab is selected Status Date Start time V S Mode Sample No. Rack No. -pos. End time Ict Lip Volume R CP Output Seq. Sample sequence Edit Patient ID 3. View the Hem, Ict*, and Lip columns On Hold 2021/05/21 12: 21 M est3 M 000002-1 12:34 GH Error 2021/05/21 14:28 V 123 M 000002-1 14.36 GH 4 Results are displayed 0 to 5, with 5 being the 5 • On Hold 2021/05/21 14:45 Lisa M 000002-1 14:53 4 GH 5 9 highest concentration 3 4 4. View the Volume column • If the sample volume exceeds or falls below the set range, a plus/minus numeric value is displayed • If the check is not performed, [-] is displayed 5. Select a sample from the Joblist 6. Press Browser Sample Analysis results Filter Sort Analysis start date and time(Ascending) Procedure continued on the next slide 2 Hardware check is completed. (OK) 00m 00s * Icterus check for use outside of the United States only 5 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Review Sample Results Job Aid 7 View HI*L Check and Tube Volume Check X Menu Reagent Calib. curve QC chart Order calib. Order Q Order STAT Joblist Maintenance Shutdown Edt Validate ccoC Previous Next Output Search Inquiry Cose Information (continued) On Hold Micro CP Remeasur 2900300 M Lip Vol 000006-4 Not Validated Hem A HOST 2020/09/11 11:27 7. View the Sample information display area APTT Fbg 8 • Hem, I* and/or Lip is displayed if the HI*L Dilution ratio 1/1 A 1st Evaluation info Measurement info Evaluation data check result is equal to or exceeds the set 4000- Assay Value Units Name Value value APTT sec 24.4 sec Lip level 4 3600- APTT R. *xxx.XX - Sample vol. level 8. Press the Measurement info. tab to view the 1200- [mm] Surface of Sample 2800- flagging level values [mm] 2400- Measurement time 180 9. View the error information display area 2000 Cuvette lot no. 9 Error info: A 2 SB cuvette lot no. 1000.1000.0000 Hemolyzed sample – results 1600- • Code/Message CN-Wash lot no. 1200- 1000.2000.0000 ExClean Il Lot no. could be affected. Follow laboratory protocol 800- Icteric sample 1000.2000.000 Icteric* sample – results could "Chatting point 400- 0000.4000.0030 Reagent name Lot no Vial position Elapsed time • Validated calibration curve is not found. APTT FS 538559 A2- 8 O be affected. Follow laboratory protocol CaC12 - A1- 8 O [sec) • 1000.3000.000 Lipemic sample – results could Show error description be affected. Follow laboratory protocol CN-3000 Ready * Icterus check for use outside of the United States only 6 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 7 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021

  • cn-3000
  • cn-6000
  • sysmex
  • cn