CN-3000 and CN-6000 Systems Check and Discard Trap Chamber Fluid Job Aid
This job aid describes how to peform the daily maintenance task, check and discard the trap chamber fluid, on the CN-3000 and CN-6000 Systems.
CN-3000 and CN-6000 Systems SIEMENS Healthineers Check and Discard Trap Chamber Fluid Job Aid Check and Discard the Trap Chamber Fluid – Daily Maintenance 1. Open the bottom door 2. Check if any fluid has collected in the trap chamber 3. If water has collected, proceed with the following maintenance steps. If water has not collected, close the bottom door and proceed to step 8 4. Shut down the instrument • Do not execute a probe rinse 5. Remove the trap chamber 6. Dispose of any collected fluid • Do not throw away the float 7. Reinstall the trap chamber to its original position and close the bottom door 8. Document maintenance Note: Document manual maintenance by pressing Maintenance > Manual maint. > Register manual maintenance. Note: If fluid collects inside the trap chamber every day, the main unit may 1 Effective Date: 18-Jun-2021 | HOOD05162003199986 have malfunctioned. Please contact your local Siemens service representative. Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 2 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021
- cn-3000
- cn-6000
- sysmex
- cn