Performing Automated CI Daily Maintenance on the Atellica® CI Analyzer Job Aid

Performing Automated CI Daily Maintenance on the Atellica® CI Analyzer Job Aid

This job aid provides instructions on how to manually perform Automated CI Daily Maintenance on the Atellica® CI Analyzer.

Performing Automated CI Daily Maintenance on the SIEMENS Atellica® CI Analyzer Job Aid Healthineers Performing Automated CI Daily Maintenance Manually 1 Troubleshooting The CI Daily Maintenance contains these automated maintenance System DL CI-01 activities: Current Status Day Needs Tools • System cleaning: Probes, mixers, and CH cuvettes STATS Sample Interventions 2 Supply Needs Worklist Assay Evaluation • Cleans IM wash block water lines 1 WU 2 0 • Refresh the CH Reaction Bath Additive 1 Inprocess 0 Calibration Reagent Needs Calibration Results Orders IMT Daily Cleaning 1 Intervention Needed • 0 QC 1 Patient (non-STAT) 5 L • CI Daily Autocheck The system automatically performs Automated CI Daily Maintenance QC Statistics Calibration Due Assay Status QC Scheduler 3 Maintenance at a configured time, or it can be manually performed 5 % 5 KG on demand for troubleshooting purposes only. Inprocess: CH-02 To perform the Daily Maintenance activity manually: 0 Errors QC Needs Guided Workflows Utilities 0 Activity Falled 3 Warnings 8 Activities Due 5 00 1. Ensure the CI Analyzer is in Ready or Standby state: A. In the System Status area, select the CI module < 1 2 3 > B. Review the status of the CI analyzer Note: The CI Daily Maintenance is not required on the same scheduled day the analyzer Verify the Water Bath Additive (WBA) and RPC4 are onboard by performs the CI Weekly Maintenance. 2. navigating to Supply Needs > View All Supplies Note: The WBA is loaded and unloaded onto the system the same as all other reagents. The 3. Navigate to Maintenance > All Activities WBA will automatically be off loaded from the system when the Onboard Stability and/or expiration date passes. *This procedure is continued on the next slide. Template Effective Date: 10 Aug 2021 | HILS 2223 Effective Date: 21-JUL-2022 | HOOD05162003285288 Unrestricted © Siemens Healthineers, 2022 Performing Automated CI Daily Maintenance on the SIEMENS Atellica® CI Analyzer Job Aid Healthineers Performing Automated CI Daily Maintenance 8 DL-C11 LabManager Maintenance Manually (continued) Filters All Activities ? HX 2022 4. From the Filters Module drop-down menu, View By Filters Tx Clear Module : CI-01 Activity Type: All Frequency : All select the appropriate CI Analyzer module By 4 Module Module Activity Last Completed CI-01 View Details 5. Select CI Daily Maintenance Activity Type CI-01 Checking the Lamp Coolant May 25, 2022 All 6 Perform Sunday Mond CI-01 CI Autocheck June 1, 2022 6. 26 Frequency Select Perform > OK 27 All Mark as Completed CI-01 CI Daily Maintenance 5 June 1, 2022 Q Search Activity Procedure Details 2 Wait for maintenance to complete (up to 60 CI-01 CI Weekly Maintenance 7. Inprocess (4 minutes remaing) W minutes) 10 CI-01 Cleaing the Exterior of the Reagent Probe May 26, 2022 CI-01 Cleaning the CH Cuvettes May 31, 2022 8. The Maintenance Log records and stores 16 17 CI-01 Cleaning the Cuvette Waste Area May 31, 2022 this activity as completed 23 24 CI-01 Cleaning the IM Primary Waste Reservoir May 30, 2022 30 31 CI-01 Hardware Check Maintenance May 11, 2022 Note: The analyzer checks for In-process samples CI-01 IMT Advanced Clean June 1, 2022 prior to the start of this daily maintenance. The Frequency All CI-01 IMT Daily Cleaning May 31, 2022 analyzer waits for these samples to process and Module All CI-01 Inspecting and Emptying the IM Water Trap May 25, 2022 then starts the activity. Activity Type All CI-01 Refreshing the Water Bath Additive May 31, 2022 Activity Status All Note: If any part of this maintenance fails, the Close Related Cancel an Autostart All Activities analyzer automatically repeats that activity when CI Daily Maintenance is performed again. The operator would need to access the failed Note: The Procedure Details button on the Maintenance screens (highlighted above) can be maintenance activity and select to 'Perform.' used to access Online Help instructions on how to perform the maintenance task. 2 Effective Date: 21-JUL-2022 | HOOD05162003285288 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 3 siemens-healthineers.com Effective Date: 21-JUL-2022 | HOOD05162003285288 Unrestricted © Siemens Healthineers, 2022

  • atelica
  • attellica
  • attelica