Performing Automated CH Daily Maintenance on the Atellica® CH Analyzer Job Aid Version 1.26 and Above
This job aid provides instructions on how to perform the daily maintenance task of Automated CH Daily Maintenance for the Atellica® IM Analyzer.
Performing Automated CH Daily Maintenance on the SIEMENS Atellica® CH Analyzer Job Aid Healthineers Performing Automated CH Daily Maintenance 1 2 Troubleshooting Manually System SH-01 SH-02 CH-01 CH-02 IM-01 CI-01 The CH Analyzer has automated daily maintenance Current Status Day Needs Tools Samples activities to decontaminate the analyzer. The system Supply Needs 2 STATS Sample Interventions Worklist automatically performs Automated CH Daily Maintenance at configured times or it can be manually 1 0 2 2 01 performed on demand for troubleshooting purposes 1 STATS Intervention Needed 1 Routine Patients Reagent Needs Calibration Results JOOC only. If any part of this maintenance fails, the analyzer 1 STATS Inprocess 0 Calibration 0 automatically repeats that maintenance. QC 5 mm To perform the Daily Maintenance activity manually: 3 Maintenance QC Statistics Calibration Due Assay Status 1. Ensure the CH Analyzer is in Ready, Processing or 10 ¡ 3 K 5 KG Standby state: Inprocess: CH-02 QC Needs Guided Workflows A. In the System Status area, select the 0 Errors 0 Activity Falled 3 Warnings appropriate CH Module 8 Activities Due 5 SH-01 SH-02 B. Review the status of the CH analyzer < 1 2. Verify the Water Bath Additive (WBA) and RPC4 are onboard by navigating to Supply Needs > View All Note: The CH Daily Maintenance is not required on the same scheduled day the analyzer Supplies performs the CH Weekly Maintenance. 3. Navigate to Maintenance > All Activities Note: The WBA is loaded and unloaded onto the system the same as all other reagents. The WBA will automatically be off loaded from the system when the Onboard Stability and/or *This procedure is continued on the next slide. expiration date passes. Template Effective Date: 27-MAY-2019 | HOOD05162003010284 1 Effective Date: 28-Jun-2021 | HOOD05162003201947 Unrestricted © Siemens Healthineers, 2021 Performing Automated CH Daily Maintenance on the SIEMENS Atellica® CH Analyzer Job Aid Healthineers Maintenance 6 LabManager Performing Automated CH Daily All Activities ? ¿ X Filters 021 Maintenance Manually (continued) Filters Module : CH-02 Activity Type: All Frequency : All View By 4. From the Filters Module drop-down menu, 4 Module Module Activity Last Completed By CH-02 select the appropriate CH Analyzer module CH-02 IMT Advanced Clean April 14, 2021 View Details Activity Type All 6 Perform CH-02 IMT Daily Cleaning April 14, 2021 5. Select CH Daily Maintenance Sunday Mond Frequency All Mark as Completed 26 27 CH-02 Cleaning the Cuvette Waste Area March 15, 2021 Q Search Activity 5 Procedure Details 6. Select Perform > Yes CH-02 CH Daily Maintenance April 14, 2021 W CH-02 Inspecting th washer Probes April 8, 2021 7. Select the option to confirm the next 10 CH - 02 Hardware Check Maintenance March 24, 2021 scheduled maintenance time 16 17 CH-02 Replacing the IMT Peristallic Pump Tubing April 10, 2021 8. Wait for maintenance to complete 23 24 9. The Maintenance Log records and stores 30 31 this activity as completed Frequency All Module All Activity Type All Activity Status All Close Related Review & Approve Maintenance Log All Activities 7 Thursday 4/15/2021 1:6 PM Note: The analyzer checks for In-process samples prior to the start of this daily maintenance. The analyzer waits for these samples to process and then starts the activity. 2 Effective Date: 28-Jun-2021 | HOOD05162003201947 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 3 Effective Date: 28-Jun-2021 | HOOD05162003201947 Unrestricted © Siemens Healthineers, 2021
- attelica
- attellica
- atelica