Biograph PET Systems - How to Correct Parameters – Injected Dose
This job aid provides instructions on how to correct injected dose information
Biograph PET Systems - How to Correct Parameters – Injected Dose Open the Patient Browser and select the Patient with the wrong data at the Study Level. Next select Patient -> Reconstruction. O Patient Browser LOX Patient | Applications Transfer Edit PET View Filter Evaluation Sort Options Help Register Num Sent Send to Simulation TaskCard Emergency ... Load to Viewing to Load Report Load Last Report Reconstruction Update Reference LITIes PATIENT Mode Reference Lines Export CT Data HT, THISWASIT Import CT Data. Export PET/CT Data CORRECT_ PET Onco (Adult)_ (502] Patient Protocol Import PET/CT Data Filming Layout ... Vision600_WB_CE [8] PET Dose Report 11111 : Film Task Status /ision600_MBF Film Preview .. [7] PET Dose Report SR Expose Film Task Ctrl+P Vision600_Brain [501] Dose Report Copy to Film Sheet Num Enter 11111 1 Show MPPS dules [6] PET Statistics Complete MPPS Print Preview STMBF3 ([3] CT 3.0 Br38 Print The patient will open into the chronicle. Select the PET scan, then click on the Routine tab and select the Syringe icon to access the dose information. Eauent Applications FET all uselt view setup mmage Qpuons system help TRULY, CORRECT H TRULY, CORRECT Examination 123456 Hospital +01-Jan-1980, M, 44Y- Biograph1 28_Vision 600 Edge 123456 Biograph1 28_Vision 600 Edge 20-Sep-2024 1-2 1-1 VO80A :"01-Jan-1980, M, 44Y VGBOA H-SP :20-Sep-2024 00:09:06.02 00:11:40.00 1 IMA 1 5 IMA 351 TOP 1 AC 2007 SP -750.5 SP -759.5 Viewing Filming KV 120 GT 0.0 2048 0/0 Tr60.fon.UOPO .. 11 .. T 100% B Onco (Adult) TRULY, CORRECT 1051.5 mm 21 sec 3 mCi 3D Topogram Isotope F-18 Pharm CT Dose 3.00 mc CH Pause Date 20-Sep-2024 Time Edit dose parameters Simulation PEI Scan range Match CT Range No. of beds Scan duration 3 sec TrueD Range: Begin End Table: Position Height Scan direction HOLO Recor -1811.0 -759.5 -1513.5 125.0 Caudocranial Routine Scan Recon Auto Tasking 9% 19% 20-Sep-2024 00:48:57 Page 1 of 6 Effective Date: 07/11/2024 | QR700015851 SIEMENS Unrestricted © Siemens Healthcare Pty Ltd. 2024 Healthineers Biograph PET Systems – How to Correct Parameters – Injected Dose If the injected dose information was entered incorrectly, click on the Edit check box, then re- enter the correct dose information. Radiopharmaceutical Dose X Isotope Ga-68 Pharmaceutical Edotreotide (DOTATOC) Volume (mL) 0.00 Pre-injection Assay Dose 3.00 mCi Date 20-Sep-2024 Time 00:00:55 Injection Assay Dose 3.00 mCi Start Date 20-Sep-2024 v Start Time 00:00:55 Stop Date 20-Sep-2024 Stop Time 00:00:55 Post-injection Residual Assay Dose mCi Date 20-Sep-2024 Time 00:50:02 DICOM Information Edit Blood Glucose Level Radionuclide Total Dose 3.00 mci Radiopharmaceutical Start Time 00:00:55 0.00 mmol/l Radiopharmaceutical Injection Duration 0 S Date 20-Sep-2024 OK Cancel Click OK. You will receive this pop up indicating the data will need to be reprocessed. Click OK. Examination TRULY, CORRECT Hospital TRULY, CORRECT Hospital 123456 +01-Jan-1980, M, 44Y- Biograph1 28_Vision 600 Edge :123456 VG80A "01-Jan-1980, M, 44Y Biograph1 28_Vision 600 Edge VG80A 20-Sep-2024 H-SP :20-Sep-2024 H-SP-CR 00:09:06.02 00:11:40.00 1 IMA 1 5 IMA 351 TOP 1 AC 2007 SP -750.5 SP -759.5 Viewing 10 cm PET Examination x Filming KV 120 mA 35 PET Raw data information has changed. TI 9.0 Please reprocess the data and recalculate GT 0.0 SUV. 2048 0/0 Tr60.fOn.UOPO .. 1i .. Onco (Adult) 1051.5 mm 21 sec 3 mCi 3D Topogram OK Dose 3.00 mCi Pause Date 20-Sep-2024 Time 00:00:55 Simulation PET Scan range Match CT Range No. of beds 7 Scan duration 3 sec TrueD Range: Begin End Table: Position Height Scan direction Hold Recor Recon -1811.0 -759.5 1513.5 125.0 Caudocranial Routine Recon Auto Tasking 20-Sep-2024 00:50:51 Page 2 of 6 Effective Date: 07/11/2024 | QR700015851 SIEMENS Unrestricted © Siemens Healthcare Pty Ltd. 2024 Healthineers Biograph PET Systems – How to Correct Parameters – Injected Dose Click on the Recon tab. Select the first recon, then click on 8 or the next available recon spot. Applications FET Edit isell View S TRULY, CORRECT H Hospital TRULY, CORRECT Examination 123456 *01-Jan-1980, M. 44Y Biograph1 28_Vision 600 Edge 123456 Biograph1 28_Vision 600 Edge VG80A-01-Jan-1980, M, 44Y 20-Sep-2024 2A17-1 H-SP 20-Sep-2024 H-SP-CR 00:09:06.02 00:11:40.00 1 IMA 1 4 IMA 351 TOP 1 AC 2007 SP -750.5 SP -759.5 Viewing Filming KV 120 mA 35 TI 9.0 GT 0.0 2048 0/0 Tr60.fOn.UOPO .. 11 .. AC CT 5.0 HD_FOV Onco (Adult) TRULY, CORRECT 1051.5 mm 21 sec 3 mCi 3D Topogram Recon job D 12 3 4 5 6 8 Series description PET AC CT Recon range Bed 1 7 Attenuation correction CT Pause Output Image type Corrected AC CT 5.0 HD_FoV (1-1) Scatter Correction Simulation PET Recon method TrueX+TOF (ultraHD-PET) Relative PET AC Iterations 4 Subsets 5 Registration Matrices Default Image size 440 Zoom 1.0 | Filter All-pass TrueD Offset (mm) x 0 Y F Additional Settings Hold Recor Comments Routine Scan Recon Auto Tasking 20-Sep-2024 00:53:48 This will copy all the parameters of the PET AC recon to a new reconstruction box. Edit the Series Description to reflect this is the corrected reconstruction. Also check the Auto Tasking tab to ensure this recon will be transferred to the desired network nodes, ie PACS. Click Recon. TRULY, CORRECT Examination 123456 Biograph1 28_Vision 600 Edge 01-Jan-1980, M, 44Y VOBOA 20-Sep-2024 00:09:06.02 1 IMA 1 TOP 1 SP -750.5 Viewing Filming KV 120 mA 35 TI 9.0 GT 0.0 2048 0/0 Tr60.fOn.UOPO .. 1i .. Onco (Adult) TRULY, CORRECT 1051.5 mm 21 sec 3 mCi 3D Topogram Recon job 1 2 3 4 5 6 7 8 Series description PET AC Corrected CT Recon range Bed 1 Attenuation correction CT Pause Output Image type Corrected AC CT 5.0 HD_FoV (1-1) Simulation Scatter Correction PET Recon method TrueX+TOF (ultraHD-PET) Relative terations 4 Subsets Registration Matrices Default Image size 440 Zoom 1.0 Filter All-pass TrueD Offset (mm) × 0 YO Additional Settings Hold Recor Recon Comments Routine Scan Recon |Auto Tasking 20-Sep-2024 00:51:53 Page 3 of 6 Effective Date: 07/11/2024 | QR700015851 SIEMENS Unrestricted © Siemens Healthcare Pty Ltd. 2024 Healthineers Biograph PET Systems – How to Correct Parameters – Injected Dose Repeat this process for any other reconstructions you need. For example, select the PET NAC, click on 8, adjust the Series Description and check the Auto Tasking tab for transfer nodes. Seup wage aprons TRULY, CORRECT Biograph1 28_Vision 600 5004 Examination H Hospital TRULY, CORRECT 123456 Biograph1 28_Vision 600 Edge 123456 *04-Jan-1980, M, 44Y VG80A "01-Jan-1980, M, 44Y 20-Sep-2024 H-SP VI 20-Sep-2024 00:09:06.02 00:11:40.00 1 IMA 1 5 IMA 351 TOP 1 AC 2007 SP -750.5 SP -759.5 Viewing VIV KV 120 mA 35 TI 9.0 GT 0.0 Tr60.fon.UOPO.11 .. Onco (Adult) TRULY, CORRECT mci 3D Topogram Recon job 1 2 3 4 5 1 8 Series description PET NAC CT Recon range Bed 7 Attenuation correction CT Pause Output Image type Uncorrected AC CT 5.0 HD_FoV (1-1) Scatter Correction Simulation PET Recon method Iterative+TOF None PET NAC Iterations 4 Subsets 5 Registration Matrices Default Image size 440 Zoom 1.0 Filter Gaussian FWHM (mm) 2.0 TrueD Offset (mm) × 0 =Y 0 Additional Settings Recor Comments Routine Scan Recon |Auto Tasking 20-Sep-2024 00:52:59 Click Recon. These new reconstructions will reflect the correct dose information. TRULY, CORRECT Examination | Viewing 123456 Blographi 28_Vision 600 Edge 01-Jan-1980, M, 44Y- 20-Sep-2024 H.sp 00:09:06.02 1 IMA 1 TOP 1 SP -750.5 R 10cm Filming KV 120 mA 35 TI 9.0 GT 0.0 2048 0/0 Tr60.fon.UOPO .. 1i .. Onco (Adult) TRULY, CORRECT 1051.5 mm 21 sec 3 mCi 3D Topogram Recon job 1 2 3 4 6 7 8 Series description PET NAC Corrected CT Recon range Bed 1 Attenuation correction CT Pause Output Image type Uncorrected AC CT 5.0 HD_FoV (1-1) Scatter Correction Simulation PET Recon method Iterative+TOF None Iterations 4 Subsets 5 Registration Matrices Default Image size 440 Zoom 1.0 Filter Gaussian FWHM (mm) 2.0 Offset (mm) X 0 HY 0 TrueD Additional Settings Recon Comments Routine Scan Recon |Auto Tasking 20-Sep-2024 00:54:41 Page 4 of 6 Effective Date: 07/11/2024 | QR700015851 SIEMENS Unrestricted © Siemens Healthcare Pty Ltd. 2024 Healthineers Biograph PET Systems – How to Correct Parameters – Injected Dose This can also be used to correct a reconstruction if the wrong Isotope has been selected. Repeat the steps above to open the patient into the chronicle, then click on the Syringe icon to open the dose dialog. Choose the correct isotope from the drop down menu options, along with the appropriate pharmaceutical. Radiopharmaceutical Dose x Isotope F-18 Pharmaceutical Fluorodeoxyglucose Volume (mL) 0.00 Ga-66 Pre-injection Assay Ga-68 Ge-68 Dose 3.00 1-124 K-38 Date 20-Sep-2024 Time 00:00:55 Mn-51 Mn-52 Injection Assay X N-13 Radiopharmaceutical Dose Dose 3.00 Na-22 Nb-90 Start Date 20- 0-14 Isotope Ga-68 Pharmaceutical Edotreotide (DOTATOC) O-15 Stop Date 20 Rb-82 Volume (mL) 0.00 Sc-43 Post-injection Residual As Sc-44 Se-73 Pre-injection Assay Tb-152 Dose mCi Date 20 Dose 3.00 mCi Date 20-Sep-2024 Time 00:00:55 DICOM Information Edit Radionuclide Total Dose 3.00 Injection Assay Radiopharmaceutical Start Time 00:00:55 Dose 3.00 mCi Start Date 20-Sep-2024 Start Time 00:00:55 Radiopharmaceutical Injection Duration Stop Date 20-Sep-2024 Stop Time 00:00:55 Date 20-Sep-2024 Post-injection Residual Assay OK Can Dose mCi Date 20-Sep-2024 Time 00:50:02 DICOM Information Edit Blood Glucose Level Radionuclide Total Dose 3.00 mc 0.00 mmol/L Radiopharmaceutical Start Time 00:00:55 Radiopharmaceutical Injection Duration O S Date 20-Sep-2024v OK Cance Click OK. Hospital TRULY, CORRECT Examination Viewing Filming 8_Vision 600 Edge 123456 VO904 "01-Jan-1980, M, 44Y H-SP 20-Sep-2024 00:11:40.00 5 IMA 351 AC 2007 SP-759.5 PET Examination PET Raw data information has changed . Please reprocess the data and recalculat 2048 0/0 Tr60.fon.UOPO.11 .. Onco (Adult) 1051.5 mm 21 sec 3 mCi Topogram OK CT Dose 3.00 mei Pause Date 20-Sep-2024 Time 00:00:55 PET Scan range Match CT Range Scan duration sec Range: Begin -1811.0 -759.5 Table: Position -1513.5~ Scan direction Caudocranial You will receive an error message stating that the data needs to be reprocessed. Repeat the reconstructions as per the instructions above to reflect the correct isotope. Page 5 of 6 Effective Date: 07/11/2024 | QR700015851 SIEMENS Unrestricted © Siemens Healthcare Pty Ltd. 2024 Healthineers Biograph PET Systems – How to Correct Parameters – Injected Dose *Images courtesy of photo library Siemens Healthcare Pty Ltd Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare Pty Ltd, 2024 Siemens Healthcare Pty Ltd Level 3, 141 Camberwell Rd Hawthorn East, VIC 3123 AUSTRALIA Page 6 of 6 Effective Date: SIEMENS 07/11/2024 I QR700015851 Unrestricted © Siemens Healthcare Pty Ltd. 2024 Healthineers
- Biograph
- PET-CT