Atellica® UAS 800 Analyzer Sample Processing Job Aid
Process barcoded samples. Process STAT samples. Review results. Rerun a sample. Print sample results.
Sample Processing - AtellicaTM UAS 800 Analyzer SIEMENS Healthineers Processing barcoded sample tubes in Automatic mode 1. Place sample into the rack starting with position 1. 2. Tilt the rack at an angle to the right, align the rack slot with the retaining rail and load on the sample rack handler. Ensure the barcodes are facing the system. 3. Select Measure. 4. Select Automatic. 5. Select Start. Note: The minimum volume required is 2.0 mL Atellica UAS 800 Analyzer Samples to avoid: foamy samples 1 HOOD05162002602819 © Siemens AG 2013. All rights reserved. Page 1 Effective Date: 06/22/2017 Sample Processing SIEMENS Healthineers Processing barcoded sample tubes in STAT mode 1. Place sample into the rack starting with position 1. 2. Tilt the rack at an angle to the right, align the rack slot with the retaining rail and load on the sample rack handler. Ensure the barcodes are facing the system. 3. Select Measure. 4. Select STAT Test. 5. Select the check to confirm the STAT test. The system will complete testing any sample in progress and then eject that rack 6. Remove the ejected rack and select the check. Atellica UAS 800 Analyzer 7. Load the STAT rack. 8. Select the check. Note: The minimum volume required is 2.0 mL Samples to avoid: foamy samples 2 © Siemens AG 2013. All rights reserved. Page 2 Sample Processing SIEMENS Healthineers Reviewing Results Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 FI Date Range 3/27/2016 3/27/2017 V ? 1. Select Database on the Atellica™ UAS 800 Measure Database Quality Control Help into Logout Shut Down Sediment Result Selected Sample | Date: 3/27/2017 Name: - ID: 00102 software. Not invalid Sample List | 42 samples Review T Filter Dilution: 1.0 Status: + Positive omments: Fl Invalid Has Valid Image Date&Time ID Name Manual Eval Particle Category Ref. (pib) FT Not Empty Cuvette ET No Valid Image 2. Select a sample from the Sample List. 3/27/2017 11:05:12 AM N/A Invalid NO 00100 Empty Cuvette Negative RBC 3/27/2017:11:23:03 AM 8001004 N/A Invalid No WBC <3 0 9 En Low Sample Level E Positive 3/27/2017 11:25:50 AM 00102 + Positive No 15/15 .WBCC 3. Results are displayed in the Selected Sample Normal Sample Level QC Failed 3/27/2017 11:26:03 AM 8001013 + Positive No 15/15 " Has Result E QC Passed CRY 02.6 3/27/2017 11:35:15 AM 8001010 Positive Vo 15/15 area. F No Result Validated HYA O .. 3/27/2017 11:35:35 AM 8001003 + Positive 15/15 PAT 0 _ 1.5 Measured F Not Validated 3/27/2017 11:36.05 AM + Positive No 15/15 4. Use the Filter to narrow results. ET Not Measured F To Confirm NEC 0 .. 2 3/27/2017 11:36:36 AM 8001023 + Positive NO 15/15 ET Not Review Not To Confirm EPI 0. Image For review Checked Modified Valld 3/27/2017: 11:37:10 AM 8001007 Positive No 15/15 YEA 0 3 No No NO Microscopy mode 3/27/2017 11:37:44 AM 8001022 Positive NO 15/15 BAC 0 .. 130 02 No No No Manual FT Automatic 3/27/2017 11:38:14 AM 8001015 + Positive 15/15 .BACK 0 .. 130 No 3/27/2017 11:38:48 AM 8001005 + Positive No 15/15 BACK 0 .. 130 Chemical Result No NO 3/27/2017 11:39:22 AM 8001021 + Positive No 15/15 MUC 3. 264 05 No NO El Abnormal ET No Result 3/27/2017 11:39:55 AM 8001001 + Positive No 15/15 SPRM 0 .. 12 06 No No Normal Has Result 3/27/2017 1:56:22 PM 8001010 Positive No 15/15 El Error 07 No No No 3/27/2017 1:56:42 PM 8001003 + Positive NO 15/75 N No 3/27/2017 1:57:11 PM 8001006 + Positive No 15/15 Edit Recalculate Add note Reset Apply Delete Modify Comment Validate Outputs ID Errors Worklist Renun Data View Gallery Online Operator Particle Classes: 14 Recognition Threads: Waiting 0 Waste bin: 114| Warte tank Ok Water tank Ok 3/27/2017 3:58:19 Pm 3 © Siemens AG 2013. All rights reserved. Page 3 Sample Processing SIEMENS Healthineers Result Flags Flag Description Review Review is necessary Invalid Invalid sample Empty Empty cuvette ! Low level sample Skip Skipped To confirm. Sample requires manual review based on custom rule. QC passed QC failed 4 © Siemens AG 2013. All rights reserved. Page 4 Sample Processing SIEMENS Healthineers Atellica UAS 800: Automated urine Sediment Analyzer 4.0 Rerunning a sample Measure Databar Quality Control Help Logout Shut Down Sample List | 42 sample Selected Sample | Date: 3/27/2017 Name: - ID: 0010) T Filter Dilution: 1.0 Status: + Poudre Comments 1. Select a sample on the Database screen. Date& Time ID Name Manual Eval Particle Category pul Ref. (pipD 3/27/2017 11:05:12 AM N/A invalid No RBC 0.5 3:27/2017 11:23:03 8001004 N/A Invalid No 00100 2. Select Rerun. 3/27/2017 11:25:50 AM 00102 + Positive No 15/15 .WaCe 327/2017 11:26:03 AM 8001013 + Positive NO 15/15 CRY 3. If diluting the sample, choose dilution factor. 3/27/2017 11:35:15 AM 8001010 + Positive No 15/15 HYA 0_,2 3/27/2017-11:35:35 AM E001008 + Positive No 15/15 PAT 8001006 Positive No MY SOSELL ZLOULE 15/15 NEC 0.42 4. Place the sample on the sample rack handler. 27/2017-11:36:36 AM Positive No 15/15 EPI 0 ., 5 Image For review Checked Modified Valld 12017 :37:10AM 8001007 Positive No YEA 01 No No No 3/27/2017 11:37:44 AM 8001022 Positive No 15/15 5. Select Measure. BAC ) .. 130 3/27/2017 11:38:14 AM 8001015 Positiv No 15/15 .BACF D: 130 3/27/2017 11:38.48 AM 8001005 Positive No .BACC o .. 130 No NO No 6. Select Automatic. 3/27/2017 11:39:22 AM 8001021 + Positive MUC 0 264 No 3/27/2017 11:39:55 AM 8001001 Positive No 15/15 SPRM 0 .. 12 No 3/27/2017 1:56:22 PM Positive 15/15 No No No 7. Select Start. 1227/2017 1:56:42 PA 8001003 Positive No 15/15 08 No N No 2/27/2017 1:57:11 PM 8001006 Positive No 15/15 Edit Recalculate Add note Delete Modity Comment Validate Outputs ID Errors Worklist Rerun Data View Gallery Online Operator 1 Particle Classer: 14| Recognition Threads: waiting O waste bint 114 @ Waite tank Ok Water tank Ok 3/27/2017 3:58:19 PM Note: The minimum volume required to run the sample is 2.0 mL Rerun Date: 3/27/2017 Name: - ID: 8001004 Dilution Factor: 1.0 V × 5 © Siemens AG 2013. All rights reserved. Page 5 Sample Processing SIEMENS Healthineers Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 Printing sample and QC results ? Measure Database Quality Control Help Info Logout Shut Down 1. Select sample(s) on the Atellica™ UAS 800 Database screen. Sample List | 42 samples Selected Sample | Date: 3/27/2017 Name: - ID: 00102 T Filler Dilution: 1.0 Status: + Positive Comments: 2. Select Outputs. Date&Time £ Name Manual Eval Particle Category Rel. (p/pl) 3/27/2017 11:05:12 AM NIA invalid NO RBC 0.5 3. Select Print. 3/27/2017 11:23:03 AM 8001004 N/A Invalid No WBC <3 0 .. 9 3/27/2017 11:25:50 AM 00102 + Positive No 15/15 .WBCC 0 .. 9 Outputs 3/27/2017 11:26:03 AM 8001013 + Positive No 15/15 CRY 0 .. 6 4. Select the check. Date: 3/27/2017 3/27/2017 11:35:15 AM 8001010 + Positive No 15/15 HYA 0 .2 3/27/2017 11:35:35 AM 8001003 + Positive No 15/15 PAY 0 15 Selected Sample: 3/27/2017 11:36:05 AM 8001006 + Positive No 15/15 NEC 0 .. 2 3/27/2017 11:36:36 AM 8001025 + Positive No 15/15 EP 0 .. 5 Image e For review Checked Modified Valid 3/27/2017 11:37:10 AM 8001007 + Positive No 15/15 YEA 0.3 NO No No 15/15 Save Print Send To 3/27/2017 11:37:44 AM 8001022 + Positive No BAC 0 .. 130 02 No No No 3/27/2017 11:38:14 AM + Positive No 15/15 .BACI .0 .. 130 05 NO NO Database Backup: 3/27/2017 11:38:48 AM 8001005 + Positive NO .BACC 0 .. 130 04 No No No 3/27/2017 11:39:22 AM 8001021 + Positive NO 15/15 MUC ) .. 264 Vo No No 3/27/2017 11:39:55 AM 8001001 + Positive NO 15/15 SPRM 0. 17 No S 3/27/2017 1:56:22 PM 8001010 + Positive No 15/15 Full Anonymized 07 NO No Z 3/27/2017 1:56:42 PM 8001003 Positive No 15hs 08 No No No 3/27/2017 1:57:11 PM 8001006 Positive No 15/15 Edit Recalculate Add note × Delete Modify Comment Validate Outputs ID Error Worklist Rerun Data View Gallery Online Operator 1 Particle Classes: 14 Recognition Threads: Waiting O. Waste bin: 114 . Waste tank Ok Water tank Ok 3/27/2017 3:58:19 PM 6 © Siemens AG 2013. All rights reserved. Page 6