Resolving Secondary Vacuum Out of Range Errors on the Atellica® IM Analyzer Job Aid

Resolving Secondary Vacuum Out of Range Errors on the Atellica® IM Analyzer Job Aid

This job aid describes how to resolve secondary vacuum out of range errors on the Atellica® IM Analyzer.

Atellica® IM Analyzer: Resolving Common Errors SIEMENS Healthineers Secondary Vacuum Out of Range Job Aid Issue: Secondary Vacuum Pressure Out of Range Description: Secondary waste vacuum (for the aspirate probes) is out of range. This will stop the system. Waste bottles may or may not have liquid waste. Online Help Reference: Inspecting IM Waste Reservoirs for Vacuum Leaks Troubleshooting Steps: 4 3 1. Open the water and liquid waste door 2 2 1 2. Locate the appropriate waste reservoir 1 3. Ensure that the reservoir is seated correctly, and that there is no strain on any of the tubing 4. Ensure that the cap is tight and that there are no cracks in the cap 5. Ensure the connections to that waste reservoir are secure 7 6 5 6. Close the water and liquid waste door Primary waste reservoir 5 Dryer 7. If problem persists, contact the local technical support provider for assistance 2 Tertiary waste reservoir 6 Secondary waste reservoir 3 Water trap 7 Special reagent water 4 Water and liquid waste door 1 Template Effective Date: 10 Aug 2021| HILS 2223 Effective Date 08 June 2023 | HOOD05162003355898 Unrestricted © Siemens Healthineers, 2023 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. © Siemens Healthcare GmbH 2023 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 2 91052 Erlangen, Germany Effective Date 08 June 2023 | HOOD05162003355898 Telephone: +49 9131 84-0 Unrestricted © Siemens Healthineers, 2023 siemens-healthineers.com

  • Atelica
  • Attellica
  • Atellica