Resolving Sample Tip Tray Loader Jams on the Atellica® IM Analyzer Job Aid

Resolving Sample Tip Tray Loader Jams on the Atellica® IM Analyzer Job Aid

This job aid describes how to resolve sample tip tray loader jams on the Atellica® IM Analyzer.

Atellica® IM Analyzer: Resolving Common Errors SIEMENS Healthineers Sample Tip Tray Loader Jam Job Aid Issue: Sample Tip Tray Loader Jams Description: Tip Tray is jammed somewhere within the Tip Loader mechanism - usually at the staging area or while ejecting to the waste. Online Help Reference: Manually Removing IM Sample Tip Trays Troubleshooting Steps: 2 1. Ensure the analyzer is in the Stopped state 2. Unlock and open the IM Analyzer front cover, then lift up and open the IM tip tray loader cover 3. Lift up the jammed tip tray to remove it A. Ensure no additional trays are left in the tip tray loading area 4. Close the IM tip tray loader cover and the front cover 2 5. Resume the analyzer, select in the System Status area A. A new tray will load during start-up 6. If problem persists: Tip tray loader cover A. Check Tip Tray Waste and Tip Tray Storage areas for other potential jams and remove 2 IM tip tray loader manually B. Contact the local technical support provider for assistance 1 Template Effective Date: 10 Aug 2021| HILS 2223 Effective Date 08 June 2023 | HOOD05162003355898 Unrestricted © Siemens Healthineers, 2023 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. © Siemens Healthcare GmbH 2023 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 2 91052 Erlangen, Germany Effective Date 08 June 2023 | HOOD05162003355898 Telephone: +49 9131 84-0 Unrestricted © Siemens Healthineers, 2023 siemens-healthineers.com

  • Atelica
  • Attellica
  • Atellica