Atellica® HEMA 570 580 SMS Run Control Blood Samples Job Aid
This job aid provides instructions for running control blood samples on the Atellica® HEMA570 580 SMS.
Atellica® HEMA High Volume 570, 580, SMS SIEMENS Healthineers Quality Control Run a Control Blood Sample in Rack Mode 1. Prepare your control blood according to the specific instructions detailed in the control blood leaflet (temperature, mixing, etc.) 2. Place the tube in a rack Make sure that the barcode label is visible for the internal • barcode reader 3 Do not place control samples with patient samples on the • same rack 3. Place the rack on the loading tray of the analyzer 002213 The rack of control samples must always be loaded on the • analyzer but never on the Atellica® HEMA Transport 4. Press Start racks The analyzer runs automatically a second sampling for the • PLTOPT test (580 only) 5. When the analysis is completed, the QC results are automatically added to the QC results screen. The rack is ejected on the analyzer ejection tray 6. The QC indicator of the Status Bar turns red if one of the enabled Westgard rules of the analyzer has been violated or the manufacturer's target range in the QC target file, is not met 1 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021 Atellica® HEMA High Volume 570, 580, SMS SIEMENS Healthineers Quality Control Run a Control Blood Sample in STAT Mode 1. Prepare your control blood according to the specific instructions detailed in the control blood leaflet (temperature, mixing, etc.) 2. From the vertical bar, select STAT mode. The STAT mode door opens 3. Use the keyboard or barcode scanner to enter the lot number in the Sample ID box 4. Select the DM query button. The rest of the information will be filled in automatically 5. Gently mix the control sample 6 6. Place the tube in the tube holder Rotate the tube holder so the tube is in the front • position closest to you. Only one label per tube 7. Make sure that the tube cap can be pierced. If not, remove the cap (Continued on next page) 2 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021 Atellica® HEMA High Volume 570, 580, SMS SIEMENS Healthineers Quality Control Run a Control Blood Sample in STAT Mode Emergency (Continued) STAT mode 123456 5 5 ? Stop Lab Manager 8. If running a body fluid control, select Prime Analysis order 9. Select Start Sample ID AD430H HOME 5 10. When the door opens, remove the tube from the Mode Control tube holder and recap it if needed Sample type HEMA CONTROL 11. Select New order if you need to rerun the same control sample or a new sample Test DIF Start racks 12. Select Exit STAT mode when the analysis is Order HEMA CONTROL | AD430 DM query information completed EJect racks 13. When the analysis is completed, the QC results are Exit STAT automatically added to the QC results screen Comments Edit mode Information EJect slides Tube AD430H: Please mix and insert the tube, then press "Start". Note: When running an Atellica® HEMA Control BF sample type 10 Staln slide after a prolonged stop, the prime cycle may fail due to a background noise check error. If so, try to run another prime ++ cycle to solve the problem. If the problem persists, please contact Prime Start Reset New order BACK your local Siemens Healthineers representative. Reagents QC 11 Tubes READY M READY OFF 06/16/2021 HEMA580 1.0.0 Slidemaker Transport 09:49:49 AM 3 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 4 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021
- hema
- heme