Atellica® HEMA 570 580 SMS Replace a Reagent Bottle on the Analyzer Job Aid
This job aid provides instructions for replacing a reagent bottle on the on the Atellica® HEMA 570/580 analyzer.
Atellica® HEMA High Volume 570, 580, SMS SIEMENS Healthineers Maintenance Replace a Reagent Bottle - Analyzer You need the following elements to perform this procedure: Atellica® HEMA Cleaner or • Emergency 123456 5 = ? Atellica® HEMA HGB or Stop Home > Reagents Lab Manager • Atellica® HEMA Diff or Cycles Remaining DIF DIR 0 Slides 149 • O Atellica® HEMA RET/PTLO • 3 Analyzer Slidemaker HOME 1. Before replacing any reagent, you must stop the analyses 5 by selecting Pause racks on the vertical bar Supply Remaining Status 2. From the home screen, navigate to Reagents and HEMA RET/PLTO 83 % Change consumables HEMA HGB 85 % Change Start racks 3. Select the Analyzer tab HEMA Diff 0 % 4 Low Change 4. The reagent indicator shows one of the following statuses: HEMA Cleaner 72 % Change EJect racks EXPIRED, LOW or EMPTY HEMA Diluent 99 % Change (Continued on next page) STAT mode HEMA Baso 93 % Change HEMA Waste EJect slides Notes: Ejection tray • Wear protective gear, gloves and a lab coat. Staln slide • Always keep a reagent bottle close to your analyzer. This allows you to replace empty reagents with new ones that BACK are already at operating conditions temperature. Reagents V QC Tubes POWER SAVING I READY OFF 06/22/2021 HEMA580 1.0.0 lidemaker Transport 09:45:33 AM O 1 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021 Atellica® HEMA High Volume 570, 580, SMS SIEMENS Healthineers Maintenance MEMENS : .. Replace a Reagent Bottle - Analyzer (Continued) 5. Open the compartment door of the empty bottle (push in and release) 6. Unplug the reagent bottle 7. Select the Change button corresponding to the reagent 8. Scan the barcode label of the new container The reagent information fields are automatically filled • (lot number, opening date, expiration date, initial volume, and remaining volume) 9. Select the Prime reagent checkbox 10. Select OK (Continued on next page) 5 Note: Do not pour a reagent from one container into another. Particles at the bottom of the old container can contaminate the new reagent and will cause unacceptable background results especially for Platelets. 2 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021 Atellica® HEMA High Volume 570, 580, SMS SIEMENS Healthineers Maintenance Replace a Reagent Bottle - Analyzer Emergency Home > Reagents 123456 1 6 ? (Continued) Stop Lab Manager Cycles Remaining DIF 678 DIR 83 Slides 149 11. Insert the new bottle into the compartment and press down on it to ensure it is seated properly Analyzer Slidemaker HOME 5 12. Shut the reagent compartment Supply Remaining Status 13. Press Back HEMA RET/PLTO Confirmation Change 17 14. A confirmation window will appear. Select OK HEMA HGB Change Start racks 15. The analyzer primes the reagents HEMA Diff Change 16. Select Start racks from the vertical bar ? Reagents need to be primed. HEMA Cleaner Please make sure that the new reagents are in Change EJect racks position, then press "OK" to prime. HEMA Diluent Change STAT mode HEMA Baso ? Change 15 V × HEMA Waste Help Ok Cancel EJect slides Ejection tray V Stain slide Note: If you do not select Start racks after changing a BACK .................................... reagent, the analyzer will not run any analyses. Reagents V QC 0 Tubes READY RI READY OFF 06/22/2021 HEMA580 1.0.0 Slidemaker Transport 09:11:20 AM 3 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 4 Job Aid Effective Date: 11-Aug-2021| HOOD05162003214634 Unrestricted © Siemens Healthineers, 2021
- hema
- heme