Atellica® 1500 Automated Urinalysis Analyzer Training Workbook
Education resource material designed to support competency mastery on-the-job for the Atellica™ 1500 Automated Urinalysis Analyzer.
SIEMENS Healthineers ---------- SIEMENS Healthingers SIEMENS CLINITEK NowUN Atellico 1500 Atellica® 1500 Automated Urinalysis Analyzer Training Workbook *Not available for sale in the U.S. Product availability varies by country. Page 1 HOOD05162002985708 Effective date: 2/2/2019 Table of Contents SIEMENS Healthineers Hardware Overview……………………….…….….. 3 Software Overview……………………………….….. 10 Routine Operations Daily Maintenance.….…………………………. 14 System Preparation.……..………………….…. 24 Sample Processing………….……………..……. 32 Reviewing Results................................... 33 Printing Results....................................... 36 Calibration (CLINITEK Novus® Analyzer)...... 39 Quality Control….......................................... 42 Maintenance………………………………….………… 47 Monthly Maintenance................................. 49 Periodic Maintenance................................ 51 Database Backup......................................... 53 2 © Siemens AG 2013. All rights reserved. Page 2 Hardware Overview – Atellica® 1500 Analyzer SIEMENS Healthineers Atellica® 1500 Automated Urinalysis Analyzer is composed of: 1. Atellica® UAS 800 Analyzer 2. CLINITEK Novus® Automated Urine Chemistry Analyzer 3. Sample rack handler 2 1 3 3 Hardware Overview © Siemens AG 2013. All rights reserved. Page 3 Hardware Overview – Atellica® UAS 800 Analyzer SIEMENS Healthineers Atellica® UAS 800 Analyzer External Components 3 1. Sample rack handler • Up to 10 racks can be loaded SIEMENS 2. Sampling area Healthineers 7 • Samples are mixed and aspirated 3. Cuvette waste bin 5 2 6 1 • Holds used cuvette cartridges Abillar wage 4. Wash and waste bottles • Utilized for cleaning of the analyzer 5. On/off button (on left side of analyzer) WASTE 4 WASH 6. PC, monitor, keyboard, mouse (PC not shown) • Used to navigate the software 7. Serial number (on back of instrument, not shown) 4 © Siemens AG 2013. All rights reserved. Page 4 Hardware Overview – Atellica® UAS 800 Analyzer SIEMENS Healthineers Atellica® UAS 800 Analyzer Internal Components Cuvette carousel • holds up to 12 cuvette cartridges, each contains up to 50 cuvettes Not visible in this image: Pipette • Used to mix, aspirate and dispense urine sample IT Front and rear cuvette guides Cuvette • Used to transport cuvette Centrifuge carousel • Spins cuvette to bank the particles on one plane Microscope and high-resolution camera • Used to analyze the urine sample A 5 © Siemens AG 2013. All rights reserved. Page 5 Hardware Overview – CLINITEK Novus® Analyzer SIEMENS Healthineers CLINITEK Novus® Automated Urine Chemistry Analyzer External Components 1. Sample rack handler • Up to 10 racks can be loaded 2. Display • Used to navigate the software Healthincers CONHEMENS 2 3 3. Sampling area 5 • Aspirates urine sample Barcode reader 4 6 • 4. Card waste drawer 1 Atellica 1500 • Holds used cards 5. Waste bottle 6. Rinse bottle 7. On button (on side of instrument, not shown) 8. Serial number (on back of instrument, not shown) 6 © Siemens AG 2013. All rights reserved. Page 6 Hardware Overview – CLINITEK Novus® Analyzer SIEMENS Healthineers CLINITEK Novus® Automated Urine Chemistry Analyzer Internal Components 1. Pipette (not shown) • Dispenses urine sample on test pads and into SG well 2. Cassette compartment door • Holds a reagent cassette and keeps it away from moisture 3. Card Platform 4 • Holds test card during processing 4. Syringe • Aspirates and dispenses urine sample through pipette 3 2 5. SG well (not shown) • Used to measure specific gravity and report clarity 6. Color camera and LED lights (not shown) • To take images of test pads for analysis 7 © Siemens AG 2013. All rights reserved. Page 7 Hardware Overview Hardware Overview Review Questions Review Questions 1. Where are the instrument serial numbers located? 2. How many racks can be loaded onto the Atellica® 1500 Analyzer? 3. Which instrument uses cuvettes? 4. What is the function of the SG well on the CLINITEK Novus® Analyzer? 8 © Siemens AG 2013. All rights reserved. Page 8 Software Overview Hardware Overview Review Questions Review Questions Answer Key 1. On the back of each instrument. 2. 10 3. Atellica® UAS 800 4. Measure specific gravity and report clarity 9 © Siemens AG 2013. All rights reserved. Page 9 Software Overview – Atellica® UAS 800 Analyzer SIEMENS Healthineers 1 Atellica® UAS 800 Analyzer Screen Overview Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 V ? i 1. Title bar 2 Measure Database Quality Control Help Info Logout Shut Down Displays device name and software version Measure List | 25 samples • Date & Time Rack Tube ID Name status 2. Menu bar 3/27/2017 1:56:22 PM 02 01 8001010 V Completed • Displays icons used to perform a task, including Shut 3/27/2017 1:56:42 PM V Completed Down and Logout 3/27/2017 1:57:11 PM 02 8001006 3 V Completed 3/27/2017 1:57:42 PM 04 8001025 Completed 3. Workspace 3/27/2017 1:58:15 PM 02 05 8001007 Completed 3/27/2017 1:58:48 PM 02 06 8001022 V Completed • Displays information and messages for the current task 3/27/2017 1:59:20 PM 02 07 8001015 V Completed 3/27/2017 1:59:54 PM 02 08 8001005 Completed 4. System Function bar 3/27/2017 2:00:26 PM 09 8001021 V Completed • Used to perform actions on the current screen 3/27/2017 2:01:00 PM 02 800.001 V Completed 3/27/2017 2:16:48 PM 03 07 8001010 V Completed 5. Status bar 3/27/2017 2:17:09 PM 03 02 8001003 Completed Displays task and system indicators 3/27/2017 2:17:37 PM 03 8001006 V Completed • 3/27/2017 2:18:10 PM 03 04 8001025 V Completed 486 4 O A M Initialize Rack Out Clear List Cuvette Reg STAT Test Automatic Manual Start Online Operator 1 Particle Classes: 14 Recognition Threads: Waiting 0 Waste bin: 114| Waste tank Ok Water tank Ok 3/27/2017 3:57:24 PM 5 10 Software Overview © Siemens AG 2013. All rights reserved. Page 10 Software Overview – CLINITEK Novus® Analyzer SIEMENS Healthineers 1 CLINITEK Novus® Analyzer Screen Overview SIEMENS Status: Ready CLINITEK Novus 2 Home 09 51 AM 05/17/2012 1. Status bar Displays system status indicators Tests Remaining 368 Status Log • 3 Cassette 10 Test Page Cassette Lot 1022001 2. Title bar 7 Onboard Stability 13 Days 1/1 Next Sample 0-00172 • Displays current screen name, date, and time Next STAT S-00001 A 3. Cassette Status Displays status of loaded cassette V • 4. Off button • Shuts down software and powers down hardware 5. Navigation bar 8 • Displays task and system indicators 20% 100% 6. System function bar 4 • Displays menus to complete tasks 7. Status log System Setup Results Start • Displays system messages 8. System indicators 5 6 • Displays status of waste and rinse 11 Software Overview © Siemens AG 2013. All rights reserved. Page 11 Hardware Overview Software Overview Review Questions Review Questions 1. What area of the system software on both analyzers displays system status information? 2. On both analyzers, where is the Shut Down button located? 3. On the Atellica® UAS800 analyzer, which icon do you select to switch users? 12 © Siemens AG 2013. All rights reserved. Page 12 Software Overview Software Overview Review Questions Review Questions Answer Key 1. Status bar 2. Right side of Menu bar for Atellica® UAS 800, Left side of screen above Navigation bar for CLINITEK Novus® Analyzer. 3. Logout in the Menu bar 13 © Siemens AG 2013. All rights reserved. Page 13 Daily Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Empty and clean the cuvette waste bin Clean the sample rack handler 1. Remove the cuvette waste bin. 1. Wipe the sample rack handler with a cloth moistened with a 2. Empty cuvette waste bin according to your facility’s disposal guidelines. 1:10 bleach solution. 3. Clean the cuvette waste bin with 1:10 solution of bleach. 2. Wipe the sample rack handler with water and dry. 4. Rinse with water and dry. 5. Replace the cuvette waste bin. Note: The cuvette counter resets to 0 when it is replaced. G CLINITEK NovUs A UAS 800 rack handler ********************** 14 Daily © Siemens AG 2013. All rights reserved. Page 14 Maintenance Daily Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Clean the front cuvette guide 1. Open the front cover. 2. Remove the front cuvette guide holding it at a slight angle to avoid bumping the cuvette carousel. 3. Wipe the guide with 70% isopropyl alcohol. 4. Wipe the guide with water and dry. 5. Clean the supporting plate and the pins where the guide is installed with a cloth moistened with 70% isopropyl alcohol. 6. Replace the front cuvette guide at a slight angle. Align the holes on the guide with the supporting pins. Confirm there is no gap when the guide is seated on the supporting pins. 7. Close the front cover. 15 Daily © Siemens AG 2013. All rights reserved. Page 15 Maintenance Daily Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Clean the rear cuvette guide 2 1. Open the front cover. 2. Turn the microscope arm toward the front of the analyzer. Do not remove the securing screw. 3. Hold the rear cuvette guide at a slight angle and remove it from the analyzer. 4. Wipe the guide with 70% isopropyl alcohol, wipe with water and dry. 5. Clean the supporting plate and the pins where the guide is installed with a cloth moistened with 70% isopropyl alcohol. #### 6. Replace the rear cuvette guide at a slight angle. Align the holes on the guide with the supporting pins in the analyzer. 7. Turn the microscope arm back to its original location. 8. Close the front cover. 16 © Siemens AG 2013. All rights reserved. Page 16 Daily Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Check the washing liquid bottle 1. Check the washing liquid bottle level in the status bar or by visual Waste bin: 114 Waste tank Ok Water tank Ok inspection. 2. Refill the bottle if it is almost empty. Do not perform this task when the analyzer is running. BERNSTEIN 3. Remove the cap with attached sensor and place it on a clean surface. 4. Fill the bottle with deionized water. 5. Replace the cap on the bottle. Note: Ensure the tubing is inserted into the cap until the “Push Until” label on the tubing is almost covered. WASH 17 © Siemens AG 2013. All rights reserved. Page 17 Daily Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Empty and clean the waste water bottle 1. Check the waste water bottle level in the status bar or by visual Waste bin: 114 Waste tank Ok Water tank Ok inspection. 2. Remove the cap with attached sensor. 3. Empty the bottle according to the lab’s waste disposal guidelines. BERNSTEIN 4. Clean the bottle with a 1:10 solution of bleach. 5. Replace the cap with attached sensor. Note: Allowing the floater to stay in contact with the tubes inside the bottle could result in an erroneous waste status. WASTE 18 © Siemens AG 2013. All rights reserved. Page 18 Daily Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Perform daily disinfecting rinse Shut Down (49) 1. Confirm the analyzer is in the Online status. 2. Remove all racks from the rack handler. 5 Are you sure you want to shut down the Atellica UAS 800 Automated Urine Sediment Analyzer? 3. Pour 6 mL of 2% bleach solution into a test tube. Place the test tube in Position 1 of an empty rack. × 4. Select the Shut Down button on the software. 5. Select the check to confirm you want to shut down the analyzer. Do you want to begin the daily disinfecting rinse? (35) 6. Select the check to confirm you want to start the procedure. 6 7. Open the front cover. X 8. Remove the front cuvette guide holding it at a slight angle to avoid bumping the cuvette carousel. Starting Daily Disinfection Wash 9. Close the front cover and select the check to start the rinse. - Place rack with a test tube of 6 ml 2% bleach solution - Remove front cuvette guide 10. After the disinfecting rinse is finished and the instrument is powered off - Close instrument door - Click green check to continue. (120) automatically, replace the front cuvette guide at a slight angle. Align the 9 holes on the guide with the supporting pin. × 11. Close the front cover. 19 © Siemens AG 2013. All rights reserved. Page 19 Daily Maintenance – CLINITEK Novus® Analyzer SIEMENS Healthineers Clean the SG well SIEMENS À Status: Not Ready CLINITEK Novus 1. Pour at least 2 mL of 5.25% sodium hypochlorite into a tube labeled System 01:30 PM 07/29/2013 Bleach. Calibration Diagnostics 2. Place the tube into the first position of a sample rack. Control System Management 3. Load the rack onto the system. Clean SG Well About the System 4. Select System. Prime Pump 5. Select Clean SG Well. 6. Select Start. Load & Unload Clean the sample rack handler ? 1. Use a germicidal wipe to clean the sample rack handler. System Setup Results Start 20 © Siemens AG 2013. All rights reserved. Page 20 Daily Maintenance – CLINITEK Novus® Analyzer SIEMENS Healthineers Empty the waste bottle (External bottle) Empty the waste bottle (Internal bottle) 1. Disconnect the tubing from the bottle. 1. Remove racks from the sample rack handler. 2. Empty the waste according to your facility’s disposal guidelines. 2. Press the latch and pull the waste bottle drawer out of the 3. Reconnect the tubing. slot. 3. Turn the bottle cap counterclockwise and remove it. 4. Empty the waste according to your facility’s disposal guidelines. 5. Replace the cap and turn it clockwise, finger tight. 6. Wipe off any spillage on the bottle. 7. Slide the waste bottle drawer into the slot until it is latched. 21 © Siemens AG 2013. All rights reserved. Page 21 Hardware Overview Daily Maintenance Review Questions Review Questions 1. On the Atellica® UAS 800 analyzer, where do you check the status of the waste water bottle and washing liquid bottle? 2. What cleaning solution is used to clean the front and rear cuvette guides on the Atellica® UAS 800 analyzer? 3. What concentration of sodium hypochlorite is used in the cleaning of the CLINITEK Novus® SG well? 22 © Siemens AG 2013. All rights reserved. Page 22 Software Overview Daily Maintenance Review Questions Review Questions Answer Key 1. Status bar 2. 70% isopropyl alcohol 3. 5.25% 23 © Siemens AG 2013. All rights reserved. Page 23 SIEMENS System Preparation – Atellica® UAS 800 Analyzer Healthineers Logging on 1. Press the power button on the analyzer. 2. Turn on the personal computer (PC). 3. When the dialog box appears, type in login name and SIEMENS Healthineers password. Atellica UAS 800 4. Select OK. Automated Urine Sediment Analyzer Username Password Logging off Ok 1. Select Logout in the Menu bar. Version 4.0.72 | @ 2015 Siemens AG 24 System © Siemens AG 2013. All rights reserved. Page 24 Preparation SIEMENS System Preparation – Atellica® UAS 800 Analyzer Healthineers Loading cuvettes CE IVD 1. Open the front cover. CE [IVDI CE [IVD] CE LIVO LOT 2. Rotate wing screw to the left on top of cuvette carousel. 2019-04-23 8 2019-04-23 LOT] 100429516 160429512 3 2019-04-23 LOT] 160429616 LOT 2019-04-23 160429616 Atellica URS 800 3. Lift and remove top of cuvette carousel. TEZ Ovettes Areilica URS 800 Eso queries Atellica UAS 800 Areilica UAS 200 50 4. Open cuvettes carton and remove registration card. Ouvettes 5. Load a single cuvette cartridge into the carousel. Image of cuvettes carton 6. Turn the carousel clockwise to the next empty slot. 7. Load cuvette cartridges until the carousel is full. 8. Replace top on carousel and tighten wing screw. 9. Remove tape at the bottom of each cuvette cartridge. 10. Close the front cover. Note: Cuvettes must be registered with the system before processing samples. 25 © Siemens AG 2013. All rights reserved. Page 25 SIEMENS System Preparation – Atellica® UAS 800 Analyzer Healthineers Cuvette Registration (44) Registering cuvettes You must register cuvettes before using them on the analyzer. 1. Select the Measure icon. Remove the registration card from the front of the cuvette box and 2. Select the Cuvette Reg button. then click the green check button. Insert the cuvette registration card into the barcode scanner slot 3. Press the check. when prompted. Insert the registration card in the sampling area with barcode . To cancel registration click the red X button. 4. facing the reader. 3 × 5. Select the check to return to the Measure screen. Barcode scan successful. (69) Cuvette registration complete. 5 Cuvette Reg 26 © Siemens AG 2013. All rights reserved. Page 26 System Preparation – CLINITEK Novus® Analyzer SIEMENS Healthineers Loading reagent cassette 1. Select System. 2. Select Load & Unload. 3. Select Yes. 4. Open the front cover. 5. Unlock and open the cassette door. 6. Remove the cassette. 7. Load the new cassette. 8. Close and lock the cassette door. 9. Close the front cover. Note: The system reads the lot number and expiration date of the reagent cassette when the front cover is closed. 27 © Siemens AG 2013. All rights reserved. Page 27 SIEMENS System Preparation – CLINITEK Novus® Analyzer Healthineers Empty and fill rinse bottle 1. Remove racks from the sample rack handler. 2. Press the latch forward and pull the rinse drawer out of the slot. 3. Turn the bottle cap counter clockwise to remove it. 4. Empty the rinse solution according to your facility’s disposal guidelines. 5. Fill the rinse bottle with 1000 mL of distilled or deionized water. 6. Add 2 mL of CLINITEK Novus Rinse Additive. 7. Gently swirl the bottle six times. 8. Replace the cap and turn it clockwise, finger tight. 9. Slide the rinse drawer into the slot until it is latched. 10. Select System. 11. Select Prime Pump. 28 © Siemens AG 2013. All rights reserved. Page 28 System Preparation – CLINITEK Novus® Analyzer SIEMENS Healthineers Empty card waste drawer 1. Remove racks from the sample rack handler. 2. Press the latch forward and pull the card waste drawer out of the slot. 3. Empty the cards according to your facility’s disposal guidelines. 4. Remove and clean the internal liner with isopropyl alcohol. 5. Replace the internal liner. 6. Slide the card waste drawer into the slot until it is latched. 29 © Siemens AG 2013. All rights reserved. Page 29 Hardware Overview System Preparation Review Questions Review Questions 1. After loading cuvettes onto the Atellica® UAS 800 analyzer, which screen do you need to access to register cuvettes? 2. On the CLINITEK Novus® analyzer, which screen provides access to the Load & Unload button for loading reagent cassettes? 3. Which task must be performed after filling the rinse bottle on the CLINITEK Novus® analyzer? 30 © Siemens AG 2013. All rights reserved. Page 30 Software Overview System Preparation Review Questions Review Questions Answer Key 1. Measure screen 2. System screen 3. Prime pump 31 © Siemens AG 2013. All rights reserved. Page 31 Sample Processing – Atellica® 1500 Analyzer SIEMENS Healthineers Loading barcoded sample tubes 1. Place sample into the rack starting with position 1. 2. Tilt the rack at an angle to the right, align the rack slot with the retaining rail and load on the CLINITEK Novus® Analyzer sample rack handler. Ensure the barcodes are facing the system. 3. Select Start on the CLINITEK Novus screen. When sampling on the CLINITEK Novus Analyzer is complete, the rack moves along the sample rack handler to the Atellica® UAS 800 Analyzer. Note: The minimum volume required to run the sample on both systems is 2.6 mL Samples to avoid for CLINITEK Novus® analyzer: visibly bloody, mucoid or foamy samples Samples to avoid for Atellica® UAS 800 analyzer: foamy samples 32 Sample © Siemens AG 2013. All rights reserved. Page 32 Processing Sample Processing – Atellica® 1500 Analyzer SIEMENS Healthineers Reviewing Results Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 F Date Range 3/27/2016 3/27/2017 V ? i - 1. Select Database on the Atellica® UAS 800 Measure Database Quality Control Help Info Logout Shut Down Sediment Result Sample List | 42 samples Selected Sample | Date: 3/27/2017 Name: - ID: 00102 software. Not Invalid Ft Review Y Filter Dilution: 1.0 Status: + Positive omments: Fl Invalid F Has Valid Image Date&Time ID Name /- Manual Eval Particl Category Ref. (p/p Not Empty Cuvette F. No Valid Image 3/27/2017 11:05:12 AM 2. Select a sample from the Sample List. N/A Invalid No 00/00 RB 0 .. 5 Empty Cuvette E Negative 3/27/2017 11:23:03 AM 8001004 N/A Invalid No 00/00 WBC <3 1 .. 9 ET Low Sample Level El Positive 3/27/2017 11:25:50 AM 00102 Positive No 15/15 WBCC O .. 9 3. Results are displayed in the Selected Sample El Normal Sample Level QC Failed 3/27/2017 11:26:03 AM 8001013 + Positive No 15/15 Has Result QC Passed CRY O .. 3/27/2017 11:35:15 AM 8001010 + Positive No 15/15 area. FI No Result F Validated HYA O .. 3/27/2017 11:35:35 AM 8001003 + Positive No 15/15 FI Measured Not Validated PAT D .. 1.5 3/27/2017 11:36:05 AM 8001006 Positive No 15/15 4. Use the Filter to narrow results. ET Not Measured ET To Confirm NEC o .. 3/27/2017 11:36:36 AM 8001025 Positive NO 15/15 Not Review Not To Confirm EPI o .. 5 Image for review Checked Valid 3/27/2017 11:37:10 AM Positive No 15/15 YEA O .. 01 No No Microscopy mode 3/27/2017 11:37:44 AM 8001022 + Positive No 15/15 BAC 0 .. 130 O No No No Manual El Automatic 3/27/2017 11:38:14 AM 8001015 + Positive No 15/15 .BACr 0 .. 130 03 N No No 3/27/2017 11:38:48 AM 8001005 + Positive No 15/15 0 .. 130 04 No Vo Chemical Result BACK NO 3/27/2017 11:39:22 AM 8001021 + Positive No 15/15 MUC O .. 264 05 No No El Abnormal ET No Result 3/27/2017 11:39:55 AM + Positive No 15/15 SPRM 0 .. 12 06 No Nc Normal F Has Result 3/27/2017 1:56:22 PM 8001010 + Positive No 15/15 07 No El Error No 3/27/2017 1:56:42 PM 8001003 + Positive NO 15/15 NO No 3/27/2017 1:57:11 PM 800100 Positive No 15/15 Edit Recalculate Add note Reset Apply ARME Delete Modify Comment Validate Outputs ID Errors Worklist Rerun Data View Gallery Online Operator Particle Classes: 14 Recognition Threads: Waiting 0 Waste bin: 114| Waste tank Ok Water tank Ok 3/27/2017 3:58:19 PM 33 © Siemens AG 2013. All rights reserved. Page 33 Sample Processing – Atellica® 1500 Analyzer SIEMENS Healthineers Result Flags Flag Description Review Review is necessary Invalid Invalid sample Empty Empty cuvette ! Low level sample Skip Skipped To confirm. Sample requires manual review based on custom rule. QC passed QC failed 34 © Siemens AG 2013. All rights reserved. Page 34 Sample Processing – Atellica® 1500 Analyzer SIEMENS Healthineers Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 V Rerunning a sample Measure Database Quality Control Help Sample List | 42 samples Shut Down Selected Sample | Date: 3/27/2017 Name: - ID: 00102 Q Filter Dilution: 1.0 Status: + Positive 1. Select a sample on the Database screen. Comments: Date&Time ID Name Manual Eval 3/27/2017 11:05:12 AM Particl Category Ref. (p/ul) N/A Invalid No 00/00 RBC 3/27/2017 11:23:03 AM 0 .. 2. Select Rerun. 8001004 N/A Invalid NO NBC 3/27/2017 11:25:50 AM 00102 Positive No 15/15 3/27/2017 11:26:03 AM .WBC 8001013 Positive ... No 15/15 CRY 3. If diluting the sample, choose dilution factor. 3/27/2017 11:35:15 AM 8001010 + Positive o .. NO 15/15 3/27/2017 11:35:35 AM HYA 8001003 0. Positive No 15/15 PAT 3/27/2017 11:36:05 AM 0 .. 1.5 8001006 Positive 15/15 NEC • As a shortcut, hold the left/right arrow on keyboard to move through the 3/27/2017 11:36:36 AM D ... 2 8001025 - Positive NO 15/15 3/27/2017 11:37:10 AM + Positive EPI 15/15 O .. 5 ge|For review Checked 8001007 Modified Valid NO dilution factor. For example, to choose a dilution factor of 5.0, you don't have to YEA 3/27/2017 11:37:44 AM 8001022 07 Positive 0 .. No No NO 15/15 No BAC 0 .. 130 3/2712017 11:38:14 AN 8001015 + Positive No No No click from 1.0 to 5.0. You can hold the arrow and it will quickly move forward. No 15/15 .BACK 3/27/2017 11:38:48 AM 0 .. 130 6001005 + Positive 03 NO NO 15/15 No BACC NO 3/27/2017 11:39:22 AM 8001021 Positive 0 .. 130 04 15/15 No No No MUC Place the sample on the CLINITEK Novus® Analyzer sample rack handler. 3/27/2017 11:39:55 AM 8001001 Positive 0 .. 264 NO 5/15 FINO 4. NO SPRM 3/27/2017 1:56:22 PM O .. 12 8001010 + Positive No No 15/15 06 No 3/27/2017 1:56:42 PM 07 8001003 . Positive NO 15/15 Note: Diluted samples should be loaded on the Atellica® UAS 800 rack handler. 3/27/2017 1:57:11 PM 8001006 positive No M 15/15 Edit Recalculate Add note See more information on running diluted samples on the next slide. Modify Comment Validate Outputs ID Errors Norklis Rerun Data View Online Operator 1 Particle Classes: 14 Recognition Threads: Waiting 0 Waste bin: 114 @ Waste tank Ok 5. Select Start on the CLINITEK Novus screen. Water tank OK 3/27/2017 3:58:19 PM Rerun Note: if running a diluted sample on the Atellica® UAS 800 system, select Start Date: 3/27/2017 Name: - ID: 8001004 on the Atellica® 1500 Analyzer. Dilution Factor: 1.0 When sampling on the CLINITEK Novus system is complete, the rack moves along the sample rack handler to the Atellica® UAS 800 Analyzer. × Note: The minimum volume required to run the sample on both systems is 2.6 mL. 35 © Siemens AG 2013. All rights reserved. Page 35 Sample Processing – Atellica® 1500 Analyzer SIEMENS Healthineers Diluting Samples in a Connected System • Dilution factors are not entered on the CLINITEK Novus® Analyzer; therefore, diluted samples should not be run on the CLINITEK Novus® Analyzer. • If the Atellica® UAS 800 Analyzer is not able to determine results due to an overcrowded sample, the Rerun option may be used to dilute and re-run the sample to obtain a result. In this case, the diluted sample(s) should only be run on the Atellica® UAS 800 Analyzer. A dilution factor should be entered to calculate the undiluted sample result. Caution: As is common on sediment analyzers, dilution may cause deterioration of sediment particles. Sediment particles present in low number may be missed upon dilution. Each laboratory should establish its own protocols to verify results accuracy for diluted samples. 36 © Siemens AG 2013. All rights reserved. Page 36 Sample Processing – Atellica® 1500 Analyzer SIEMENS Healthineers Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 Printing sample and QC results V ? i Measure Database Quality Control Info Logout Shut Down 1. Select sample(s) on the Atellica® UAS 800 Database screen. Sample List | 42 samples Selected Sample | Date: 3/27/2017 Name: - ID: 00102 Dilution : 1.0 Status : + Positive Q Filter Comments : 2. Select Outputs. Date&Time A ID Name Manual Eval Particle Category pul ef. (p/ul) 3/27/2017 11:05:12 AM N/A Invalid NO RBC <4 0 .. 5 3/27/2017 11:23:03 AM 8001004 N/A Invalic N 00/00 3. Select Print. WBC <3 0 .. 9 3/27/2017 11:25:50 AM 0010: + Positive No 15/15 .WBCC 0 .. 9 Outputs 3/27/2017 11:26:03 AM 8001013 + Positive No 15/15 CRY 0 .. 6 4. Select the check. Date: 3/27/2017 3/27/2017 11:35:15 AM 8001010 + Positive No 15/15 HYA 0 .. 2 3/27/2017 11:35:35 AM 8001003 + Positive 15/15 PAT O .. 1.5 3 Selected Sample: 3/27/2017 11:36:05 AM 8001006 + Positive NO 15/15 NEC O .. 3/27/2017 11:36:36 AM 8001025 + Positive NO 15/15 EPI 0 .. 5 Image For review Checked Modified Valid 3/27/2017 11:37:10 AM 8001007 + Positive NO 15/15 YEA 0 .. 3 01 No No No + Positive NO 15/15 Save Print Send To 3/27/2017 11:37:44 AM 8001022 BAC 0 .. 130 02 No No No 3/27/2017 11:38:14 AM 8001015 + Positive NO 15/15 .BACr 0 .. 130 03 No NO N NO Database Backup: 3/27/2017 11:38:48 AM 8001005 + Positive 15/15 . BAC 0 .. 130 04 No No No 3/27/2017 11:39:22 AM 8001021 + Positive NO 15/15 MUC 0 .. 264 05 No No No 3/27/2017 11:39:55 AM 8001001 Positive NO 15/15 SPRM 0 .. 12 06 No 3 3/2/12017 1:56:22 PM 8001010 + Positive NO 15/15 Full Anonymized 07 No NO NO 3/27/2017 1:56:42 PM 8001003 Positive No 15/15 08 NO NO NO 3/27/2017 1:57:11 PM 8001006 + Positive 15/15 Edit Recalculate Add note anmans × V Delete Modify Comment Validate Outputs ID Errors Worklist Rerun Data View Gallery Online Operator 1 Particle Classes: 14 Recognition Threads: Waiting 0 Waste bin: 114| Waste tank Ok Water tank Ok 3/27/2017 3:58:19 PM Note: Only validated results can be printed, or saved, or send to LIS. 37 © Siemens AG 2013. All rights reserved. Page 37 Hardware Overview Sample Processing Review Questions Review Questions 1. After racks are loaded onto the sample rack handler, how do you start the run? 2. What is the minimum volume required to run a sample on both analyzers? 3. Where are sample results viewed? 4. How are QC results printed? 38 © Siemens AG 2013. All rights reserved. Page 38 Software Overview Sample Processing Review Questions Review Questions Answer Key 1. Select Start on the CLINITEK Novus® analyzer software 2. 2.6 mL 3. Atellica® UAS 800 analyzer Database screen 4. Atellica® UAS 800 analyzer Database screen, select QC result(s), select Outputs, select Print 39 © Siemens AG 2013. All rights reserved. Page 39 Calibration – CLINITEK Novus® Analyzer SIEMENS Healthineers Preparing Calibrator Rack Performing Calibration Rack Position Calibrator 1. Select System. SIEMENS Status: Ready CLINITEK Novus 2. Select Calibration. System > Calibration 10 25 AM 03/27/2012 1 Bleach The calibration passed. 3. To add a new lot select Change. Calibration Date & Time 03/27/2012 10:20 AM 2 CAL #4 Select Add. CAL 1 Lot 2106 • CAL 2 Lot 2106A CAL 3 Lot • Enter a lot. 2106B 3 CAL #1 CAL 4 Lot 2106C Print Select an expiration date. Cassette 10 Test • Cassette Lot 1022001 Calibration Temperature Send 4 CAL #2 23.2 ºC 4. Select Next. 5 CAL #3 5. Select Start. Exit 6. The calibration status of passed or Note: Allow calibrators to equilibrate failed will display at the end of the run. to room temperature before using. Reviewing Calibration Results 1. Select Results. 2. Select Calibration. 3. Select All Results. 40 © Siemens AG 2013. All rights reserved. Calibration Page 40 Hardware Overview Calibration Review Questions Review Questions 1. The calibrators should reach which temperature before use? 2. What is the proper placement of bleach and calibrators in a rack? 41 © Siemens AG 2013. All rights reserved. Page 41 Software Overview Calibration Review Questions Review Questions Answer Key 1. Room temperature 2. Position 1 – Bleach, Position 2 – CAL #4, Position 3 – CAL #1, Position 4 – CAL #2, Position 5 – CAL#3 42 © Siemens AG 2013. All rights reserved. Page 42 Quality Control – Atellica® UAS 800 Analyzer SIEMENS Healthineers Adding new QC lot information Running a QC test 1. Select Quality Control. 1. Follow QC vendor instruction to mix the QC material in its container. 2. Select Add in the Low or High Level Lots List pane. 2. Pour at least 2 mL of low and high level control solutions into 3. Select the Liquid type. appropriately labeled test tubes. 4. Enter the lot number. 3. Place the test tubes in a rack. Low level control solution in position 1 and high level control solution in position 2. 5. Select the Expiration date. 4. Place the rack on the sample rack handler. 6. Enter the RBC and WBC limit ranges. 5. On the QC Settings screen, select Low Level Lot and High Level Lot 7. Select the check. checkboxes for the appropriate lots. 6. Select Start QC. 7. Select the check. 8. When testing is complete, a message appears indicating QC Passed or QC Failed. 43 © Siemens AG 2013. All rights reserved. Quality Control Page 43 Quality Control – Atellica® UAS 800 Analyzer SIEMENS Healthineers Reviewing QC results Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 V Select Database. Measure Database Quality Control Settings Help Logout hut Down 1. QC Settings QC Graphs 2. Select Data view or Gallery on the lower-right for more details. QC Conditions O Last 8 weeks @ Between 8/16/2016 8/26/2016 Update Upper and Lower Limits Category Low Level High Level Low level RBC High level RBC Lot ID 79581 79582 320 300 iquid Type Bio-Rad qUAntify Plus Bio-Rad q 280 Expiration Date 5/13/2017 5/13/2017 35 260 243.76 30 28.16 240224.4 218.24 221.76 227.04 First - Last Meas . Date 8/16/2016 - 8/23/2016 8/16/2016 22 21.12 24.64 220 201 62 208 28 25 20.24 195.36 200 194.48 Reviewing QC graphs Measure Count 11 10 20 13.2 14.96 180 160 RBC Limits (p/ul) 0-50 105-316 140 RBC Average (p/pl) 20.00 213.58 120 1. Select Quality Control. RBC SD (p/ul) 4.97 15.18 12:00: 140:00 12:01 18 400:00 12:0/19 400:00 12:00:00 12:00. 00:00 2.022 AM O 12:003 12:007 00:00 12.8/18 :00:00 AMO 400:00 12021 419:00 2:00:00 2:00:00 AM BC CV (%) 24.83 7.11 WBC Limits (p/ul) Low level WBC High level WBC 0 -25 55 - 167 2. Select QC Graphs. 170- WBC Average (p/jul) 0.00 107.58 24 160 22 150 WBC SD (p/ul) 0.00 20 8.31 18 140 3. Select the date range from the drop-down calendar. VBC CV (%) 0.00 7.72 16 180124.74 14 120 114.18 112.86 111.54 105.6 109.56 12 110 103.62 58:340 10 100 96.36 8 90 6 80 4. Select Update. 70 2 0 0 O 60 O O O 8/17 8/18 8/19 8/22 8/23 8/18 12:00:00 12:00:00 12:00:00 2:00:0 12:00:00 12:00:00 2.000 12:00:00 12:00-0 Output 12:020 12:0000 12:00:00 AM AM AM AM AM 2.00:00 00:00 Online 800Service article Classes: 14 Recognition Threads: Waiting 0 Waste bin: 46 | Waste tank Ok Water tank Ok 4/13/2017 5:35:36 AM 44 © Siemens AG 2013. All rights reserved. Page 44 Quality Control – CLINITEK Novus® Analyzer SIEMENS Healthineers Running a QC test Reviewing QC results at the end of a run 1. Pour at least 2 mL of each control into appropriately labeled test 1. Results automatically appear when the QC run is complete. tubes. 2. Use the scroll buttons to review details at the top of the screen. 2. Place the test tubes in a rack. Control 1 in position 1, Control 2 in position 2, Control 3 in Position 3 (if applicable). 3. Use the navigation bar to proceed to the next control. 3. Place the rack on the sample rack handler. Reviewing QC results for monthly review 4. Select System. 1. Select Results from the Home screen. 5. Select Control. 2. Select Control. 6. Verify the onscreen control lots are correct. 3. Enter search criteria. 7. To add a new lot, select Change. Reviewing QC Graphs for monthly review 1. Select Results from the Home screen. • Select Add. Enter the control lot and select Enter. 2. Select Control. • • Enter the expiration date and select Enter. 3. To graph all results, select All Results and then Graph. 8. Select Next. 4. To graph specific analytes or controls, select the analyte/control and select Enter. 9. Select Start. 45 © Siemens AG 2013. All rights reserved. Page 45 Hardware Overview Quality Control Review Questions Review Questions 1. Which limit ranges need to be entered for quality control material on the Atellica® UAS 800 analyzer? 2. What screen do you have to be in to start QC on the Atellica® UAS 800 analyzer? 3. How do you start a QC run on the CLINITEK Novus® analyzer? 46 © Siemens AG 2013. All rights reserved. Page 46 Software Overview Quality Control Review Questions Review Questions Answer Key 1. RBC and WBC 2. Quality Control from the Menu bar 3. Select System from the Home screen, select Control, make the necessary lot selections and press Start. 47 © Siemens AG 2013. All rights reserved. Page 47 SIEMENS Maintenance – Atellica® 1500 Analyzer Healthineers Interval Maintenance Task Instrument User Guide Daily Empty and clean cuvette bin Atellica™ UAS 800 Analyzer Chapter 7 Clean the sample rack handler Both Chapter 7 Clean the front and rear cuvette guides Atellica™ UAS 800 Analyzer Chapter 7 Check the washing liquid bottle Atellica™ UAS 800 Analyzer Chapter 7 Empty and clean the waste water bottle Atellica™ UAS 800 Analyzer Chapter 7 Perform daily disinfecting rinse Atellica™ UAS 800 Analyzer Chapter 7 Clean the SG well CLINITEK Novus® Analyzer Chapter 4 Empty the liquid waste bottle CLINITEK Novus® Analyzer Chapter 4 Monthly Clean the microscope objective Atellica™ UAS 800 Analyzer Chapter 7 Clean the centrifuge door and centrifuge arm Atellica™ UAS 800 Analyzer Chapter 7 48 Maintenance Maintenance © Siemens AG 2013. All rights reserved. Page 48 SIEMENS Maintenance – Atellica® 1500 Analyzer Healthineers Interval Maintenance Task Instrument User Guide Periodic Clean the analyzer exterior Atellica™ UAS 800 Analyzer Chapter 7 Clean the drip tray Atellica™ UAS 800 Analyzer Chapter 7 Rinse the washing liquid bottle Atellica™ UAS 800 Analyzer Chapter 7 Remove empty cuvette cartridges Atellica™ UAS 800 Analyzer Chapter 7 Clean the washing liquid bottle Atellica™ UAS 800 Analyzer Chapter 7 Prime the pump CLINITEK Novus® Analyzer Chapter 4 Clean the card platform CLINITEK Novus® Analyzer Chapter 4 Clean the card grippers CLINITEK Novus® Analyzer Chapter 4 Clean the racks CLINITEK Novus® Analyzer Chapter 4 Clean the rack sensors CLINITEK Novus® Analyzer Chapter 4 Clean the pad sensor CLINITEK Novus® Analyzer Chapter 4 49 Maintenance © Siemens AG 2013. All rights reserved. Page 49 Monthly Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Clean the microscope objective 2 3 & 4 1. Open the front cover. 2. Turn the microscope arm toward the front of the analyzer. Do not remove the securing screw. 3. Gently wipe the microscope objective with a cloth dipped in 70% isopropyl alcohol. 4. Dry the objective with a clean cloth. 5. Turn the microscope arm back to its original location. 6. Close the front cover. 50 Monthly © Siemens AG 2013. All rights reserved. Page 50 Maintenance Monthly Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Clean the centrifuge cover and centrifuge arm 1. Open the front cover. 2. Remove the front cuvette guide holding it at a slight angle to avoid bumping the cuvette carousel. 3 3. Pull the black knob on the centrifuge cover to release the cover. 4. Remove the centrifuge cover. 5. Grasp the center of the centrifuge arm and carefully pull to remove it. 6. Wipe the centrifuge cover and arm with 70% isopropyl alcohol. 7. Rinse the centrifuge cover and arm with water and dry them. 8. Replace the centrifuge arm by aligning it with the magnet. The magnet will pull the arm into place. KIDS 9. Replace the centrifuge cover by aligning the pin on the knob with the clot on the door. 10.Replace the front cuvette guide. 11.Close the front cover. 5 51 © Siemens AG 2013. All rights reserved. Page 51 Periodic Maintenance – Atellica® UAS 800 Analyzer SIEMENS Healthineers Performing periodic maintenance tasks Refer to the Atellica® UAS 800 Analyzer User Guide for detailed instructions for each task. 1. Clean the analyzer exterior. 2. Clean the drip tray (if necessary). 3. Rinse the washing liquid bottle. 4. Remove empty cuvette cartridges. 5. Clean the washing liquid bottle. 52 Periodic © Siemens AG 2013. All rights reserved. Page 52 Maintenance Periodic Maintenance – CLINITEK Novus® Analyzer SIEMENS Healthineers Performing periodic maintenance tasks Refer to the CLINITEK Novus® Analyzer Operator’s Guide for detailed instructions for each task. 1. Prime the pump. 2. Clean the card platform. 3. Clean the card grippers. 4. Clean the racks. 5. Clean the rack sensors. 6. Clean the pad sensor. 7. Clean the barcode reader. 53 © Siemens AG 2013. All rights reserved. Page 53 Database Backup – Atellica® 1500 Analyzer SIEMENS Healthineers Atellica UAS 800: Automated Urine Sediment Analyzer 4.0 Perform database backup V ? i Outputs Measure Database Quality Control Info Logout Shut Down 1. Select Database. Sample List | 42 samples Selected Sample | Date: 3/27/2017 Name: - ID: 00102 Date: 3/27/2017 Dilution : 1.0 Status : + Positive Q Filter Comments : 2. Select Outputs. Date&Time ID Name Manual Eval Particle Category pul ef. (p/ul) Selected Sample: 3/27/2017 11:05:12 AM N/A Invalid NO RBC <4 0 .. 5 3/27/2017 11:23:03 AM 8001004 N/A Invalic N 00/00 3. Select Full or Anonymized. WBC <3 0 .. 9 3/27/2017 11:25:50 AM 0010: + Positive No 15/15 .WBCC 0 .. 9 Save Print Send To V 3/27/2017 11:26:03 AM 8001013 + Positive No 15/15 CRY 0 .. 6 4. Select the target location. 3/27/2017 11:35:15 AM 8001010 + Positive No 15/15 HYA 0 .. 2 Database Backup: 3/27/2017 11:35:35 AM 8001003 + Positive 15/15 PAT O .. 1.5 3 5. Select the check to confirm. 3/27/2017 11:36:05 AM 8001006 + Positive NO 15/15 NEC O .. 3/27/2017 11:36:36 AM 8001025 + Positive NO 15/15 EPI 0 .. 5 Image For review Checked Modified Valid 8001007 6. Select the check to save. Full Anonymized 3/27/2017 11:37:10 AM + Positive NO 15/15 YEA 0 .. 3 01 No No No 3/27/2017 11:37:44 AM 8001022 + Positive NO 15/15 BAC 0 .. 130 02 No No No 3/27/2017 11:38:14 AM 8001015 + Positive NO 15/15 .BACr 0 .. 130 03 No NO N 7. Select the x to close the window. 3/27/2017 11:38:48 AM 8001005 + Positive NO 15/15 × . BAC 0 .. 130 04 No No No 3/27/2017 11:39:22 AM 8001021 + Positive NO 15/15 MUC 0 .. 264 05 No No No 3/27/2017 11:39:55 AM 8001001 Positive NO 15/15 SPRM 0 .. 12 06 No 3 312/12017 1:56:22 PM 8001010 + Positive NO 15/15 07 No NO NO 3/27/2017 1:56:42 PM 8001003 Positive No 15/15 08 NO NO NO 3/27/2017 1:57:11 PM 8001006 + Positive 15/15 Edit Recalculate Add note anmans V Delete Modify Comment Validate Outputs ID Errors Worklist Rerun Data View Gallery Online Operator 1 Particle Classes: 14 Recognition Threads: Waiting 0 Waste bin: 114 @ Waste tank Ok Water tank Ok 3/27/2017 3:58:19 PM 54 © Siemens AG 2013. All rights reserved. Page 54 Hardware Overview Maintenance Review Questions Review Questions 1. Which component needs to be removed from the Atellica® UAS 800 analyzer before the centrifuge cover and arm can be cleaned? 2. Which cleaning solution is used to clean the Atellica® UAS 800 analyzer microscope objective? 3. Which button on the Database screen provides access to the back up procedure? 55 © Siemens AG 2013. All rights reserved. Page 55 Software Overview Maintenance Review Questions Review Questions Answer Key 1. Front cuvette guide 2. 70% isopropyl alcohol 3. Outputs 56 © Siemens AG 2013. All rights reserved. Page 56 SIEMENS Training Checklist – Atellica® 1500 Analyzer Healthineers To document that a facilitation event occurred, work with the learners to complete the following checklist for your records only. This does not serve as the final sign off for the customer. It is up to the facility to determine competency for their employees. Competency Objectives Complete Competency Objectives Complete System Overview Identify major hardware components Maintenance Perform daily maintenance Navigate the software Perform monthly maintenance Routine Operations Log into the software Perform periodic maintenance Determine the status of supplies Perform database backup Load supplies and empty waste Calibration (CLINITEK Prepare calibrators Novus® Analyzer) Process patient samples Process a calibration View patient results Review calibration results Print sample results Quality Control Add new QC lot information Process QC samples View QC data Observer: ______________________________________ Operator: ______________________________________ Date: _________________________________________ Training Checklist Page 57 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, CLINITEK Novus, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability may vary from country to country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2019 Global Siemens Healthineers Headquarters Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84 0 siemens.com/healthineers 58 Effective Date: 2/2/2019 | HOOD05162002985708 Unrestricted © Siemens Healthcare GmbH, 2019