YSIO X.pree Detector Calibration (syngo XR) Job Aid - USA
In this PDF, you will learn how to calibrate the detector of the YSIO X.pree system.
SIEMENS Healthineers YSIO X.pree Detector Calibration (syngo XR) Job Aid Version 2023.xx QR700009680 I Effective: 06/12/2024 Template Effective Date: 10 Aug 2021| HILS 2223 Unrestricted © Siemens Healthineers AG, 2024 YSIO X.pree detector calibration (syngo XR) job aid SIEMENS Healthineers Table of contents General detector calibration information 3 Detector calibration: Portable detector 7 Detector calibration: Fix detector BWS 16 SIEMENS Healthingers ..... 2 Unrestricted © Siemens Healthineers AG, 2024 SIEMENS Healthineers General detector calibration information 3 Unrestricted © Siemens Healthineers AG, 2023 General detector calibration information SIEMENS Healthineers • Always follow the instructions on syngo XR exactly. • Before calibration of a MAX mini Pixium 2430EZ or X.wi- D 24 detector, the Default Type for Free Exposure in • All mobile detectors of the same type the myExam Cockpit - Detector Management must be (MAX wi-D, MAX mini or X.wi-D) except the one to be set to MAX mini or calibrated have to be disconnected. X.wi-D 24. myExam Cockpit • Allowed combinations: 1 MAX wi-D or 1 MAX mini or 1 Detector Management Result List Attributes Viewing Workflow Editor Local Data Configurations Client Settings Devices MAX wi-D + 1 MAX mini or 1 X.wi-D Default Type for Free Exposure: MAX mini O MAX wi-D Details: X X • No objects may be in the beam path. + MAX static MAX wi-D MAX mini • After successful calibration of the detector, the complete system including the imaging system has to • All mobile detectors must be calibrated in free be rebooted. positions. • The detector temperature must be constant (this is checked by the system) before the calibration is started. 4 Unrestricted © Siemens Healthineers AG, 2024 Preconditions for a successful detector calibration (1) SIEMENS Healthineers Portable detectors • The portable detector must be warmed up for at YSTO X.pree least 1 hour (battery inserted). • The detector must not have been exposed to radiation for at least 7 hours prior to calibration. • The portable detector must be free of any soiling. • The detector must be calibrated on the floor with a SID of 150 cm in the landscape or portrait format. SIEMENS .. lealthineers • The detector edges must be overexposed slightly (about 2 cm; check with the light field). 5 Unrestricted © Siemens Healthineers AG, 2024 Preconditions for a successful detector calibration (2) SIEMENS Healthineers Fixed detectors • The fixed Rad detector must be warmed up for at least 1 hour (system switched on). • The detector must not have been exposed to radiation for at least 7 hours before calibration. This also applies to previous calibration procedures. • If calibration fails, a period of at least 2 hours must expire before calibration can be attempted again. • The tabletop must be centered in the transverse direction. 6 Unrestricted © Siemens Healthineers AG, 2024 SIEMENS Healthineers Detector calibration: Portable detector (X.wi-D, MAX wi-D or MAX mini) 7 Unrestricted © Siemens Healthineers AG, 2023 Calibration of a portable detector SIEMENS Healthineers System setup This image displays the setup for the detector calibration YSIO X.pree of an MAX wi-D detector. SIEMENS - The same setup also applies for any X.wi-D detector. === V 8 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a portable detector – Step 1 of 7 SIEMENS Healthineers Enter the administration SIEMENS .. admin2 Healthineers ? portal as a clinical admin user. Healthineers XP-TC-12 / admin2 Free (Restricted Remote Access) Repair Installation Technical Configuration Clinical Configuration Tune-Up Detector Calibration 1. 1 Navigate to Repair Configuration Export-Import 1 FD-Calibration Tune-Up Detector Calibration. Select detector for calibration Calibrate FD Location Serial Number Last Calibration 2. 2 Select the detector and Free 23061A 12-07-2023 10:23:23 start the Calibration alibrate FD Location Serial Number Message menu by selecting Next. Wallstand 1527N4 Portable detector can be calibrated only in free! 2 Next Exit o Click a button to proceed V 10:43 13.07.2023 9 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a portable detector – Step 2 of 7 SIEMENS Healthineers 1 The calibration switch must be switched on. (Backside of the Exposure box) - 2 All mobile detectors must be calibrated on the floor. Do not place the detector directly on the floor! - Center the tube unit to the detector and set the SID specified by the UIS. Calibration of the portable detector is performed in landscape or portrait format. It is mandatory to overexpose the detector (circa 2 cm / 0,8 inch). MAX wi-D Healthingers STOR LIDDUINSTOPH - SNawais O 2 1 CAL 10 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a portable detector – Step 3 of 7 SIEMENS Healthineers 1. 1 Follow the instructions on the imaging system exactly. Once a condition is fulfilled, the red X changes to a green checkmark. 2. 2 If all conditions are met, the Next button appears in the status bar. Select Next then. FD-Calibration FD-Calibration Detector: fd4343ezx Detector position: Free Serial number: 23061A Detector: fd4343ezx Detector position: Free Serial number: 23061A Step 1 of 4 - check FD parameters for calibration Step 1 of 4 - check FD parameters for calibration Align tube to detector and Check that light field is at least 2cm larger than the detector size or fully open collimator Align tube to detector and Check that light field is at least 2cm larger than the detector size or fully open collimator Instruction: Insert 0.6mm Copper into the collimator and assure that no additional objects are in the beam path Instruction: Insert 0.6mm Copper into the collimator and assure that no additional objects are in the beam path Wait while the system verifies all parameters and all checkmarks are green Wait while the system verifies all parameters and all checkmarks are green (2) Parameter Ready Current value Nominal value Remarks Parameter Ready Current value Nominal value Remarks SID n.a. 150.00 cm Align tube and detector with required SID SID 149.41 cm 150.00 cm Align tube and detector with required SID Grid 1 X n.a None Assure that no grid is inserted Grid 1 n.a. None Assure that no grid is inserted Temperature constant X n.a 01:00 Wait for the displayed time until the detector temperatur is stable Temperature constant 29.4 ºC 00:00 Wait for the displayed time until the detector temperatur is stable Calibration switch X n.a. On Activate the required switch(es) Calibration switch On On Activate the required switch(es) 2 Exit Next Exit 11 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a portable detector – Step 4 of 7 SIEMENS Healthineers A status dialog appears when FD-Calibration Next is selected. Detector: fd4343ezx Detector position: Free Serial number: 23061A 1 1. Select the modes to be Step 2 of 4 - select the calibration mode(s) calibrated. Instruction: Select the mode you want to calibrate Mode Select Status Progress Reset Defect Map 2 2. To start the calibration, fd4343ezx_31 1 12-07-2023 10:20 select Go. 3. 3 Follow the instructions on the imaging system. Alert X 2 X-ray will be automatically released during calibration. Go Exit Ensure that nobody is in the examination room. OK 3 12 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a portable detector – Step 5 of 7 SIEMENS Healthineers 1. Follow the instructions on the imaging system. Insert the 0.6 mm Cu-filter and select OK. SIEMENS XP-TC-12 / admin2 Free (Restricted Remote Access) Healthineers Repair Installation " Technical Configuration Clinical Configuration HECLEMENS Tune-Up Detector Calibration Configuration Export-Import FD-Calibration 150 mm 0000 0000 Detector: fd4343ezx Detector position: Free Serial number: 23061A Step 3 of 4 - system calibrates selected mode(s) Instruction: Confirm X H ? Is the following filter inserted: 0.6 mm Cu 1 Mode Progress OK Cancel fd4343ezx_31 Series 0 of 0 Abort 13 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a portable detector – Step 6 of 7 SIEMENS Healthineers SIEMENS 1 Free (Restricted Remote Access) 1. Wait until the calibration is XP-TC-12 / admin2 Healthineers Repair Installation Technical Configuration Clinical Configuration done. Don‘t enter the room Tune-Up Detector Calibration or interrupt the calibration! Configuration Export-Import FD-Calibration At the end of the calibration, an indicator Detector: fd4343ezx Detector position: Free Serial number: 23061A displays whether the Step 4 of 4 - system has calibrated the selected mode(s) Instruction: Calibration finished, if you want to save the results calibration was successful or Mode Status Progress fd4343ezx_31 Accepted Acceptance check passed 1 not. Saving results to detector. 2 2. Select Save to save the calibration results. 2 Save Next Exit O Click a button to proceed 14 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a portable detector – Step 7 of 7 SIEMENS Healthineers 1 1. Wait until the saving is done. The progress area will be FD-Calibration updated accordingly. Detector: fd4343rec Detector position: Wallstand Serial number: 1920F4 2 2. Press Exit to leave the Step 4 of 4 - system has calibrated the selected mode(s) detector calibration. Instruction: Calibration finished,if you want to save the results Mode Status Progress 1 fd4343ezx_31 Accepted Results successfully saved 3 3. Disengage the calibration Saving results to detector switch and reboot the system. Alert X 2 Please disable calibration switch and for the changes to take effect, reboot the system Save Next Exit OK 15 Unrestricted © Siemens Healthineers AG, 2024 SIEMENS Healthineers Detector calibration: Fixed detector (MAX static) 16 Unrestricted © Siemens Healthineers AG, 2023 Calibration of a fixed detector (bucky wall stand) SIEMENS Healthineers System setup SIEMENS weithineers so YSIO X pree TTTT A STENTENS 17 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a fixed detector – Step 1 of 7 SIEMENS Healthineers Enter the administration SIEMENS ... Meduser Healthineers ? # 2 portal as a clinical admin SIEMENS .. : Healthineers XP-TC-73 / Service Technician Siemens (Restricted Remote Access) Diagnosis Repair Le Installation (ª Technical Configuration Quality Assurance Clinical Configuration user. SSD/HDD Repair Access Point Configuration 1 Download FD-Calibration Navigate to Repair System Configuration Check 1 - System Version Check Tune-Up Detector Delete Detector Data Fixed FD Assignment Select detector for calibration Calibration Tune-Up Stand Adjustments Calibrate FD Location Serial Number Last Calibration Centering Wallstand 1920F4 21-08-2023 14:32:00 Select the detector and Camera Calibration 2 - Delete Tube Load Data start the Calibration Inverter Adjustment Ca prate FD Location Serial Number Message Tube Adjustment Table 23071D Portable detector can be calibrated only in free! menu by selecting Next. mAs Adjustment Dose lontomat Detector Calibration 2 Detector Sensitivity & EXI Manual Calibration Configuration Export-Import Export Reports Next Exit 0 Click a button to proceed Service Key Expiry Date: 2023. 09. 26. XX1 1:02 PM 8/29/2023 18 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a fixed detector – Step 2 of 7 SIEMENS Healthineers The calibration switch must be switched on. 1 For Ysio X.pree systems, the SIEMENS - Healthineers swawais calibration switch can be STOR found on the backside of the 1 RCM – Exposure box. CAL 19 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a fixed detector – Step 3 of 7 SIEMENS Healthineers 1 Follow the instructions on the imaging system exactly. Once a condition is fulfilled, the red X changes to a - green checkmark. 2 If all conditions are met, the Next button appears in the status bar. Select Next then. - SIEMENS ? Meduser Meduser Healthineers SIEMENS HR Service / IQAP ealthineers 8/29/1983 ? ? calthingers XP-TC-73 / Service Technician Siemens (Restricted Remote Access) lealthineers XP-TC-73 / Service Technician Siemens (Restricted Remote Access) - Diagnosis / Repair Installation Technical Configuration Quality Assurance Clinical Configuration Diagnosis Repair Installation Technical Configuration Quality Assurance Clinical Configuration >SSD/HDD Repair SSD/HDD Repair Access Point Configuration Access Point Configuration Download FD-Calibration Download FD-Calibration System Configuration Check System Configuration Check System Version Check System Version Check Delete Detector Data Wallstand 1920F4 Delete Detector Data Detector: fd4343rec Detector position: Serial number Fixed FD Assignment Fixed FD Assignment Detector: fd4343rec Detector position: Wallstand Serial number: 1920F4 Tune-Up Step 1 of 4 - check FD parameters for calibration Tune-Up Stand Adjustments Stand Adjustments Step 1 of 4 - check FD parameters for calibration Align tube to detector and Check that light field is at least 2cm larger than the detector size or fully open collimator Align tube to detector and Check that light field is at least 2cm larger than the detector size or fully open collimator Centering Centering Camera Calibration Instruction: Insert 0.6mm Copper into the collimator and assure that no additional objects are in the beam path Camera Calibration Instruction Insert 0.6mm Copper into the collimator and assure that no additional objects are in the beam path Delete Tube Load Data Wait while the system verifies all parameters and all checkmarks are green Delete Tube Load Data Wait while the system verifies all parameters and all checkmarks are green Inverter Adjustment Inverter Adjustment Tube Adjustment Parameter Ready Current value Nominal value Remarks Tube Adjustment Parameter Ready Current value Nominal value Remarks mAs Adjustment SID I.a. 180.00 cm Align tube and detector with required SID mAs Adjustment SID 180.34 cm 180.00 cm Align tube and detector with required SID Dose lontomat Dose lontomat Grid 1 n.a. None Assure that no grid is inserted Grid 1 None None Assure that no grid is inserted Detector Calibration Detector Calibration Detector Sensitivity & EXI Temperature constant n.a. 01:00 Wait for the displayed time until the detector temperatur is stable Detector Sensitivity & EXI Temperature constant 32.8 ºC 00:00 Wait for the displayed time until the detector temperatur is stable Manual Calibration Calibration switch n.a. On Activate the required switch(es) Manual Calibration Calibration switch On Activate the required switch(es) x X X X Configuration Export-Import Configuration Export-Import Export Reports Export Reports 2 Exit Next Exit Loading UI Service Key Expiry Date: 2023. 09. 26. O Performing preparation checks. Service Key Expiry Date: 2023. 09. 26. V 3/292 PM LAY V 1:03 PM 129/2023 20 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a fixed detector – Step 4 of 7 SIEMENS Healthineers A status dialog appears when Next is selected. SIEMENS ... Meduser Healthineers Service / IQAP 8/29/1983 ? SIEMENS Healthineers XP-TC-73 / Service Technician Siemens (Restricted Remote Access) - Diagnosis Repair Installation ₹" Technical Configuration O Quality Assurance º Clinical Configuration SSD/HDD Repair 1 Select the modes to be calibrated. Access Point Configuration - Download FD-Calibration System Configuration Check System Version Check Delete Detector Data Fixed FD Assignment Detector fd4343rec Detector position: Wallstand Serial number 1920F4 Tune-Up 2 To start the calibration, select Go. Stand Adjustments Step 2 of 4 - select the calibration mode(s) - Select the mode you want to calibrate Centering Instruction: Camera Calibration Delete Tube Load Data Mode Select Status Progress Reset Defect Map Inverter Adjustment fd4343rec_31 1 21-08-2023 14:31 Tube Adjustment 3 Follow the instructions on the imaging mAs Adjustment Dose lontomat - Detector Calibration system. Detector Sensitivity & EXI Manual Calibration Configuration Export-Import Export Reports 2 Go Exit Click a button to proceed Service Key Expiry Date: 2023. 09. 26. V 2 1:05 PM 139/202 Alert X X-ray will be automatically released during calibration. Ensure that nobody is in the examination room. OK 3 21 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a fixed detector – Step 5 of 7 SIEMENS Healthineers 1- Follow the instructions on the imaging system. Insert the 0.6 mm Cu-filter and select OK. SIEMENS .. hineers AR Service / IQAP Meduser 8/29/1983 ? SIEMENS Healthineers". XP-TC-73 / Service Technician Siemens (Restricted Remote Access) Diagnosis Repair Installation * Technical Configuration O Quality Assurance Clinical Configuration SSD/HDD Repair Access Point Configuration Download FD-Calibration System Configuration Check System Version Check Delete Detector Data Fixed FD Assignment Detector: fd4343rec Detector position: Wallstand Serial number 1920F4 Tune-Up Stand Adjustments Step 3 of 4 - system calibrates selected mode(s) Centering Instruction: Confirm X Camera Calibration Delete Tube Load Data ? Is the following filter inserted: 0.6 mm Cu Inverter Adjustment Mode rogres Tube Adjustment fd4343rec_31 OK Cancel 1 eries 0 of 0 mAs Adjustment Dose lontomat Detector Calibration Detector Sensitivity & EXI Manual Calibration Configuration Export-Import Export Reports Abort D Setting CP step Service Key Expiry Date: 2023. 09. 26. V 2 1:07 PM 8/29/2023 22 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a fixed detector – Step 6 of 7 SIEMENS Healthineers 1 Wait until the calibration is - done. Don‘t enter the room FD-Calibration or interrupt the calibration! At the end of the calibration, Detector: fd4343rec Detector position: Wallstand Serial number: 1920F4 an indicator displays whether Step 4 of 4 - system has calibrated the selected mode(s) the calibration was successful Instruction: Calibration finished,if you want to save the results Mode Status or not. Progress fd4343rec_31 Accepted Acceptance check passed 1 Saving results to detector: 2 Select Save to save the - calibration results. 2 Save Next Exit 23 Unrestricted © Siemens Healthineers AG, 2024 Calibration of a fixed detector – Step 7 of 7 SIEMENS Healthineers 1 Wait until the saving is done. FD-Calibration - The progress area will be updated accordingly. Detector: fd4343rec Detector position: Wallstand Serial number: 1920F4 Step 4 of 4 - system has calibrated the selected mode(s) Instruction: Calibration finished, if you want to save the results 2 Select Exit to leave the Mode Status Progress - detector calibration. fd4343rec_31 Accepted Results successfully saved 1 Saving results to detector: 3 • Disengage the calibration switch and reboot the system. 1 Alert X 2 Save Next Exit Please disable calibration switch and for the changes to take effect, reboot the system OK 24 Unrestricted © Siemens Healthineers AG, 2024 SIEMENS Healthineers Please note this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthineers AG 2023 Siemens Healthineers Headquarters Siemens Healthineers AG Siemensstr. 3 91301 Forchheim, Germany Phone: +49 9191 18-0 siemens-healthineers.com 25 Unrestricted © Siemens Healthineers AG, 2024
- detector
- xray