
Ysio Max - Detector Calibration Job Aid
In this PDF, you will learn about how to calibrate the detector of the Ysio Max system.
Target group: All users
Audio: Yes
Recommended to be viewed on the following devices: Laptop, desktop computer (sufficiently large display required)
SIEMENS Healthineers Ysio Max - Detector Calibration (Fluorospot Compact) Template Effective Date: 26 Mar 2025 | QR700020784 1 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Ysio Max - Detector Calibration SIEMENS Healthineers (Fluorospot Compact) Preconditions for successfull detector calibration 3 Step-by-Step guide: Detector calibration portable detector 7 Step-by-Step guide: Detector calibration fixed detector wall stand 19 Step-by-Step guide: Detector calibration fixed detector table 30 Template Effective Date: 26 Mar 2025 | QR700020784 2 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Preconditions for successfull detector calibration Template Effective Date: 26 Mar 2025 | QR700020784 3 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 General detector calibration information SIEMENS Healthineers • Always follow the instructions on the FLC • The detector temperature must be constant exactly. (this is checked by the system) before the calibration is started. • All mobile detectors of the same type (MAX wi- D or MAX mini) have to be disconnected except • All mobile detectors have to be calibrated in the one to be calibrated. Combinations free positions. allowed: 1 MAX wi-D or 1 MAX mini or 1 MAX wi-D + 1 MAX mini. • No objects may be in the beam path. • After successful calibration of the detector, the complete system including the imaging system has to be rebooted. Template Effective Date: 26 Mar 2025 | QR700020784 4 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Preconditions for successful detector calibration SIEMENS Healthineers Portable detectors • The portable detector must be warmed up for at least 1 hour (battery inserted). • The detector may not have been exposed to radiation for at least 7 hours prior to calibration. • The portable detector must be free of any soiling. SIEMENS Healthineers MAX mini SIEMENS Healthineers MAX wi-D • The detector must be calibrated outside the trays with an SID of 150 cm in landscape or portrait format. • The detector edges must be overexposed slightly (about 2 cm; check with the light field). Template Effective Date: 26 Mar 2025 | QR700020784 5 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Preconditions for successful detector calibration SIEMENS Healthineers Fixed RAD detectors • The fixed Rad detector must be warmed up for at least 1 hour (system switched on). SIEMENS Healthineers • The detector may not have been exposed to radiation for at least 7 hours prior to calibration. This also applies for previous calibration procedures. • If calibration fails, a time period of at least 2 hours must expire before calibration can be attempted again. • The tabletop must be centered in the transverse direction. Template Effective Date: 26 Mar 2025 | QR700020784 6 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Detector calibration: Portable detector Template Effective Date: 26 Mar 2025 | QR700020784 7 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector (free position) SIEMENS Healthineers System setup SIEMENS A Template Effective Date: 26 Mar 2025 | QR700020784 8 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector – Service user SIEMENS Healthineers - X SERVICE • Enter the service as @ Siemens Healthcare Service Software - Internet Explorer O Tune Up Home Event Log Reports Help Logout uGy*m2 Service service level 7 user. Tube Load Component Adjust FD Calibration 0.00 TUI 1 Stand Select the detector for calibration Centering Detector Navigate to Repair > System Adjust Calibrate FD Detector Serial Number Last calibration • Untersuchung Inverter Integrated Detector in Table 123499 23-06-2022 Tube Tune-Up > Detector mAs Integrated Detector in Wall stand 1249G6 27-09-2021 OO Dose lontomat 1 MAX wi-D 161637 31-08-2021 FD-Calibration Calibration. Control Fluoro 1 Additional Tools Manual Calibration # 42.6 cm x 42.6 cm Error Suppress List 114 cm Ist-Wert Soll-Wert Nachverarbeitung phil ohne ohne • Select the particular detector and start the KV 73 Calibration menu with mAs 20 Dokumentation Next. 2 2 ms Dosis S mm Cu 0.0 000 Fokus % 80 Bild Extras After selection, click 'Next" Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 9 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector – Admin user SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer • Enter the service as clinical X Tune Up Home Logout admin user. User Password Component Adjust FD Calibration Centering System Adjust Select the detector for calibration FD Calibration 1 3D Calibration Calibrate FD Detector Last calibration • Navigate to Repair > Tune-Up Projection Geometry Serial Number Integrated Detector in Table 123499 23-06-2022 O Integrated Detector in Wall stand 1249G6 27-09-2021 > Detector Calibration. 1 MAX wi-D 161637 31-08-2021 • Select the particular detector and start the Calibration menu with Next. 2 2 After selection, click 'Next' Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 10 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers The calibration switch must be SIEMENS switched on. STISe For Ysio Max systems the calibration O switch can be found on the back of the RCM – Exposure box. 1 1 Template Effective Date: 26 Mar 2025 | QR700020784 11 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers All mobile detectors have to be calibrated on the floor. (Do not place the detector directly on the floor!) Center the tube unit to the detector and set the SID specified by the FLC. Calibration of the portable detector is performed in landscape or portrait format. It is MAX wi-D SIEMENS Healthineers mandatory to overexpose the detector (circa 2 1 cm / 0.8 inch). 1 Template Effective Date: 26 Mar 2025 | QR700020784 12 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers • Follow the instructions on the imaging system exactly. Once a condition is fulfilled, the red X changes to a green checkmark. 1 • If all conditions are met, the Next button appears in the status bar. 2 Siemens Healthcare Service Software - Internet Explorer SERVICE - X SERVICE siemens Healthcare Service Software - Internet Explorer X O Service Tune Up | Home | Event Log | Reports | Help | Logout uGy*m2 Tune Up Home | Event Log | Reports | Help | Logout uGy*m2 0 o Service 0.00 FD Calibration 0.00 Tube Load FD Calibration Tube Load Component Adjust Component Adjust TUI 1 TUI 1 Detector: MAX wi-D Serial number: 2013C6 Stand Detector: MAX wi-D Serial number: 2013C6 Centering Centering Detector Detector System Adjust Step 1 of 4 - check FD parameters for calibration System Adjust Step 1 of 4 - check FD parameters for calibration Untersuchung Inverter Inverter Untersuchung Tube Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! Einstell. Service mAs Please wait while systems verifies detector parameters Einstell. Service mAs Please wait while systems verifies detector parameters Dose lontomat Dose lontomat FD-Calibration FD-Calibration Control Fluoro Parameter Ready Current value Remarks Control Fluoro Parameter Ready Current value Remarks Additional Tools Additional Tools Manual Calibration Manual Calibration Error Suppress List SID X 114.4 cm Please set SID to 150,0 cm Error Suppress List SID 150,0 cm Nachverarbeitung Nachverarbeitung Grid 1 Grid 1 DO Temperature constant 35.6 ºC Please wait ... (00:23) Temperature constant 35.7 ºC < X Dokumentation Dokumentation 2 System is verifying detector parameters ... Manual Back ----- Exit Please click 'Next' to select the calibration mode Manual Back Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 13 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers @ siemens Healthcare Service Software - Internet Explorer SERVICE X • A status dialog appears when Tune Up Home Event Log Reports Help Logout uGy*m2 Service Next is selected. Tube Load FD Calibration O 0.00 Component Adjust TUI 1 Detector: MAX wi-D Stand Serial number: 2013C6 Centering Detector System Adjust Step 2 of 4 - select the calibration mode(s) • Select the modes to be Instructions: Please select the mode you want to calibrate Untersuchung Inverter Tube calibrated. 1 mAs Mode Select Status Progress Reset Einstell. Service Dose lontomat FD-Calibration Control Fluoro DR Radiography Additional Tools 1 Last calibration: 23-6-2022 Manual Calibration DR Tomography / Long Time Exposure Last calibration: 23-6-2022 Error Suppress List • To start the calibration, select Nachverarbeitung DO Go. 2 • Follow the instructions on the Dokumentation imaging system. 3 Message from webpage × X-ray will be automatically released during calibration. Ensure that nobody is in the examination room. You can stop the calibration at any time by clicking "Abort' 2 3 OK Click 'Go' to start the calibration Manual Back Go Exit Template Effective Date: 26 Mar 2025 | QR700020784 14 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers Follow the instructions on the imaging e Siemens Healthcare Service Software - Internet Explorer X SERVICE Tune Up Home Event Log Reports Help Logout uGy*m2 Service system. Insert the 0.6 mm Cu filter and Tube Load FD Calibration 0.00 O Component Adjust calib fd4343rc 30 table press OK. 1 TUI 1 Stand Detector: MAX wi-D Serial number: 2013C6 Centering Detector System Adjust Step 3 of 4 - system calibrates selected mode(s) Untersuchung Inverter Instructions: Please wait while the system performs acquisitions Tube The system then will process the acquired images The Cu filter can be found in the blue mAs Dose lontomat Mode Select Status Progress Reset FD-Calibration accessory folder. Control Fluoro Additional Tools DR Radiography Started Preparation, please wait ... Manual Calibration 46.9 cm x 46.9 cm Error Suppress List DR Tomography / Long Time Exposure Last calibration: 23-6-2022 8 130 cm Ist-Wert Soll-Wert Nachverarbeitung ohne ohne Message from webpage X KV 70 ? Is the following filter inserted: - 0.6 mm Cu mAs Dokumentation Click 'Cancel' to abort the calibration ms 100 Dosis 1 OK Cancel mm Cu 0.0 teatthineers 000 Fokus % 80 SNEWS Bild Extras 1 System is performing acquisitions ... Abort Template Effective Date: 26 Mar 2025 | QR700020784 15 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers SERVICE O • Wait until the calibration is has e Siemens Healthcare Service Software - Internet Explorer X Tune Up Home Event Log Reports | Help | Logout uGy*m2 Service finished. Do not enter the room or O 4 User Password FD Calibration 3346.0 Tube Load calib fd3543ezh 31 free Component Adjust interrupt the calibration! At the end TUI 1 Detector: MAX wi-D Serial number: 2013C6 TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) of the calibration, an indicator is Stand Centering Instructions: Please wait while the system performs acquisitions Detector The system then will process the acquired images Examination displayed showing whether or not System Adjust Rad Inverter Mode Select Status Progress Fluoro Reset Tube mAs the calibration was successful. 1 Dose lontomat FD-Calibration DR Radiography Accepted Finished (not yet saved) Dose DEXI Water Equivalent DR Tomography / Long Time Exposure Started 53 % processing # 47.0 cm x 47.0 cm Control Fluoro 150 cm Postprocessing Collimation 3D Calibration Current Target • Press OK to save the calibration Projection Geometry none Reference 10 2D-Topo Calibration Message from webpage X results. 2 Reference 10 1 Additional Tools Manual Calibration Error Suppress List ? The calibration has finished successfully. kV 70 Save the calibration results now? mAs Documentation Click 'Cancel' to discard all calibration results or save them later. ms Dose OK Cancel mm Ci 0.0 Focus 2 % 80 Image Tools System is processing acquired images ... Abort S/N:2013C6 Ready 2013C6 O 7%HU Template Effective Date: 26 Mar 2025 | QR700020784 16 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers e Siemens Healthcare Service Software - Internet Explorer X SERVICE • Wait until the results have Tune Up Home Event Log Reports Help Logout uGy*m2 Service O 4 been saved. The progress FD Calibration 1723.0 User Password Tube Load calib fd3543ezh 30 free Component Adjust TUI 1 Detector: MAX wi-D Serial number: 2013C6 area will be updated TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) Stand accordingly. Centering Instructions: 1 Examination 1 Detector System Adjust Mode Select Status Progress Reset Settings Service Inverter Tube mAs DR Radiography Successful Dose lontomat Saved • Confirm with OK. 2 FD-Calibration DR Tomography / Long Time Exposure Successful javed Dose DEXI Water Equivalent Control Fluoro Collimation Postprocessing 3D Calibration Projection Geometry Reference 10 2D-Topo Calibration Reference 10 Additional Tools Message from webpage X Manual Calibration Error Suppress List Calibration results have been saved sucessfully Documentation 2 OK Synchronizing new calibration data, please wait .. Abort S/N:2013C6 Ready 2013C6 O 10%HU Template Effective Date: 26 Mar 2025 | QR700020784 17 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer X • Press Exit to leave the SERVICE Tune Up Home Event Log Reports Help Logout uGy*m2 Service detector calibration. 1 4 User Password FD Calibration 1723.0 O Tube Load calib fd3543ezh 30 free Component Adjust Detector: MAX wi-D Serial number: 2013C6 TUI 1 TUI 2 TUI 3 Step 4 of 4 - system has calibrated the selected mode(s) • An information message Stand Centering Instructions: Examination Detector System Adjust Mode Select Status Progress Reset Settings ervice comes up to deactivate the Inverter Tube mAs DR Radiography Successful Saved calibration switch. Dose lontomat 2 FD-Calibration Dose DEXI DR Tomography / Long Time Exposure Successful Saved Water Equivalent Control Fluoro Collimation Postprocessing 3D Calibration • Disengage the calibration Projection Geometry Reference 10 2D-Topo Calibration TEST switch and reboot the Reference 10 Additional Tools Message from webpage × Manual Calibration Error Suppress List system. Please deactivate the listed switch(es): Calibration Switch Documentation 2 OK 1 Click 'Next' to select another detector. Click 'Go' to repeat the calibration Manual Next| Go Exit S/N:2013C6 Ready 2013C6 10%HU Template Effective Date: 26 Mar 2025 | QR700020784 18 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Detector calibration: Fixed detector BWS Template Effective Date: 26 Mar 2025 | QR700020784 19 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers (Wall stand or patient table) - System setup Stringers Healthineers 73 20 Template Effective Date: 26 Mar 2025 | QR700020784 20 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector – Service user SIEMENS Healthineers e Siemens Healthcare Service Software - Internet Explorer X SERVICE • Enter the service as service Tune Up Home Event Log Reports Help O Logout uGy*m2 Service level 7 user. O Tube Load FD Calibration 0.00 o Component Adjust TUI 1 Stand Select the detector for calibration Centering Detector Navigate to Repair > Tune-Up System Adjust Calibrate FD Detector Serial Number Last calibration • Inverter Integrated Detector in Table 123499 23-06-2022 IT Untersuchung Tube Integrated Detector in Wall stand 1249G6 27-09-2021 > Detector Calibration. mAs 1 Dose lontomat 1 MAX wi-D 1616:37 31-08-2021 FD-Calibration Control Fluoro Additional Tools Manual Calibration 42.6 cm x 42.6 cm Error Suppress List 114 cm • Select the particular detector 8 Ist-Wert Soll-Wert Nachverarbeitung ohne ohne and start the Calibration menu with Next. 2 kV 73 2 mAs 20 Dokumentation ms Dosis mm Cu 0.0 Fokus % 80 Bild Extras After selection, click 'Next' T Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 21 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector – Admin user SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer O X • Enter the service as clinical Tune Up Home Logout admin user. User Password Component Adjust FD Calibration Centering System Adjust 1 Select the detector for calibration FD Calibration 3D Calibration Calibrate Last calibration • Navigate to Repair > Tune-Up Projection Geometry FD Detector Serial Number Integrated Detector in Table 123499 23-06-2022 Integrated Detector in Wall stand 1249G6 27-09-2021 > Detector Calibration. 1 MAX wi-D 161637 31-08-2021 OOO • Select the particular detector and start the Calibration menu with Next. 2 2 After selection, click 'Next' Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 22 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers The calibration switch must be SIEMENS switched on. STOP For Ysio Max systems the calibration O switch can be found on the backside of the RCM – Exposure box. 1 1 Template Effective Date: 26 Mar 2025 | QR700020784 23 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers • Follow the instructions on the imaging system exactly. Once a condition is fulfilled, the red X changes to a green checkmark. 1 • If all conditions are met, the Next button appears in the status bar. 2 Siemens Healthcare Service Software - Internet Explorer - X SERVICE Siemens Healthcare Service Software - Internet Explorer - X SERVICE Tune Up Home Event Log Reports Help Logout uGy*m2 Tune Up Home | Event Log | Reports | Help Logout uGy*m2 Service Service 0 O Tube Load FD Calibration 0.00 Tube Load Component Adjust FD Calibration 0.00 ok Component Adjust TUI 1 TUI 1 Stand Detector: Integrated Detector in Wall stand Serial number: 1249G6 Stand Detector: Integrated Detector in Wall stand Serial number: 1249G6 Centering Centering Detector Detector System Adjust Step 1 of 4 - check FD parameters for calibration System Adjust Step 1 of 4 - check FD parameters for calibration Untersuchung Untersuchung Inverter Inverter Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! Tube Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! Einstell Service mAs Please wait while systems verifies detector parameters Einstell. Service mAs Please wait while systems verifies detector parameters Dose lontomat Dose lontomat FD-Calibration ED Calibration Control Fluoro Parameter Ready Current value Remarks Control Fluoro Parameter Ready Current value Remarks Additional Tools Additional Tools Manual Calibration Manual Calibration Error Suppress List SID 114.4 cm Please set SID to 180,0 cm Error Suppress List SID 180,0 cm Nachverarbeitung X Nachverarbeitung Grid 1 V Grid 1 Temperature constant X 35.6 ºC Please wait ... (00:23) Temperature constant 35.7 º Dokumentation Dokumentation 2 System is verifying detector parameters . Manual Back ----- Exit Please click 'Next' to select the calibration mode Manual Back| Next |Exit Template Effective Date: 26 Mar 2025 | QR700020784 24 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer - X • A status dialog appears when SERVICE Tune Up Home Event Log Reports Help Logout O uGy*m2 Service Next is selected. Tube Load FD Calibration 0.00 Component Adjust TUI 1 Detector: Integrated Detector in Wall stand Stand Serial number: 1249G6 Centering Detector System Adjust Step 2 of 4 - select the calibration mode(s) • Select the modes to be Untersuchung Inverter Instructions: Please select the mode you want to calibrate Tube calibrated. 1 mAs Mode Select Status Progress Reset Einstell. Service Dose lontomat FD-Calibration Control Fluoro DR Radiography Last calibration: 23-6-2022 Additional Tools 1 Manual Calibration DR Tomography / Long Time Exposure Last calibration: 23-6-2022 • To start the calibration, select Error Suppress List Nachverarbeitung pill DO Go. 2 • Follow the instructions on the imaging system. Dokumentation 3 Message from webpage X X-ray will be automatically released during calibration. Ensure that nobody is in the examination room. You can stop the calibration at any time by clicking 'Abort" 3 2 OK Click 'Go' to start the calibration Manual Back Go Exit Template Effective Date: 26 Mar 2025 | QR700020784 25 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers Follow the instructions on the imaging @ Siemens Healthcare Service Software - Internet Explorer X SERVICE Tune Up Home Event Log Reports Help Logout Service system. Insert the 0.6 mm Cu filter and uGy*m2 Tube Load FD Calibration 0.00 O O Component Adjust calib fd4343rc 30 table press OK. 1 TUI 1 Stand Detector: Integrated Detector in Wall stand Serial number: 1249G6 Centering Detector System Adjust Step 3 of 4 - system calibrates selected mode(s) Untersuchung Inverter Instructions: Please wait while the system performs acquisitions The Cu filter can be found in the blue Tube The system then will process the acquired images mAs Dose lontomat Mode Select Status Progress Reset FD-Calibration accessory folder. Control Fluoro Additional Tools DR Radiography Started Preparation, please wait ... Manual Calibration 46.9 cm x 46.9 cm Error Suppress List DR Tomography / Long Time Exposure Last calibration: 23-6-2022 130 cm Ist-Wert Soll-Wert Nachverarbeitung ohne ohne Message from webpage X KV 70 ? s the following filter inserted: - 0.6 mm Cu mAs 1 Dokumentation Click 'Cancel' to abort the calibration ms 100 Dosi Sco 1 OK Cancel Healthineers mm Cu 0.0 000 Fokus SNOWHIS % 80 NOunvD ........... Bild Extras 1 System is performing acquisitions ... Abort Template Effective Date: 26 Mar 2025 | QR700020784 26 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer • Wait until the calibration has X SERVICE Tune Up Home Event Log Reports Help Logout uGy*m2 Service finished. Do not enter the room 4 O User Password FD Calibration 3346.0 Tube Load calib fd3543ezh 31 free or interrupt the calibration! At Component Adjust TUI 1 Detector: Integrated Detector in Wall stand Serial number: 1249G6 TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) the end of the calibration, an Stand Centering Please wait while the system performs acquisitions Detector Instructions: The system then will process the acquired images Examination indicator is displayed showing System Adjust Fluoro Inverter Mode Select Status Progress Reset Rad Tube mAs whether or not the calibration Dose lontomat FD-Calibration DR Radiography Accepted Finished (not yet saved) Dose DEXI DR Tomography / Long Time Exposure Started 47.0 cm x 47.0 cm was successful. Water Equivalent 53 % processing 0 1 Control Fluoro 150 cm Collimation 8 Postprocessing 3D Calibration Current Target Projection Geometry none Reference 10 2D-Topo Calibration Message from webpage Press OK to save the calibration Reference 10 1 X • Additional Tools Manual Calibration Error Suppress List ? The calibration has finished successfully. results. KV 70 2 Save the calibration results now? mAs Documentation Click 'Cancel' to discard all calibration results or save them later. ms Dose 2 OK Cancel mm Cu 0.0 Focus % 80 Image Tools System is processing acquired images ... Abort S/N:2013C6 Ready 2013C6 7%HU Template Effective Date: 26 Mar 2025 | QR700020784 27 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fix RAD detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer X SERVICE • Wait until the saving is done. The Tune Up Home Event Log Reports Help Logout uGy*m2 Service progress area will be updated 1723.0 User Password FD Calibration O 4 Tube Load calib fd3543ezh 30 free Component Adjust accordingly. TUI 1 Detector: Integrated Detector in Wall stand Serial number: 1249G6 1 TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) Stand Centering Instructions: Examination Detector System Adjust Mode Select Status Progress Reset Settings Service Confirm with OK. Inverter • 2 Tube mAs DR Radiography Successful 1 Saved Dose lontomat FD-Calibration Save Dose DEXI DR Tomography / Long Time Exposure V Successful Water Equivalent Control Fluoro Postprocessing Collimation 3D Calibration Projection Geometry Reference 10 2D-Topo Calibration Reference 10 Additional Tools Message from webpage X Manual Calibration Error Suppress List Calibration results have been saved sucessfully Documentation 2 OK Synchronizing new calibration data, please wait ... Abort S/N:2013C6 Ready 2013C6 10%HU Template Effective Date: 26 Mar 2025 | QR700020784 28 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers Siemens Healthcare Service Software - Internet Explorer SERVICE • Press Exit to leave the detector X Tune Up Home Event Log Reports Help Logout uGy*m2 Service calibration. 1 User Password FD Calibration 4 1723.0 O Tube Load calib_fd3543ezh_30 free Component Adjust TUI 1 Detector: Integrated Detector in Wall stand Serial number: 1249G6 TUI 2 TUI 3 Step 4 of 4 - system has calibrated the selected mode(s) Stand An information message comes up to Centering Instructions: • Examination Detector System Adjust Mode Select Status Progress Reset Settings Service deactivate the calibration switch. Inverter 2 Tube mAs DR Radiography Successful Saved Dose lontomat FD-Calibration DR Tomography / Long Time Exposure Successful Saved Dose DEXI Water Equivalent Control Fluoro • Disengage the calibration switch and Collimation Postprocessing 3D Calibration Projection Geometry reboot the system. Reference 10 2D-Topo Calibration Reference 10 Additional Tools Message from webpage X Manual Calibration Error Suppress List Please deactivate the listed switch(es): Calibration Switch Documentation 2 OK 1 Click 'Next' to select another detector. Click 'Go' to repeat the calibration Manual Next Go Exit S/N:2013C6 Ready 2013C6 10%HU Template Effective Date: 26 Mar 2025 | QR700020784 29 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Detector calibration: Fixed detector table Template Effective Date: 26 Mar 2025 | QR700020784 30 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector (Table) SIEMENS Healthineers System setup SIEMENS Template Effective Date: 26 Mar 2025 | QR700020784 31 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector – Service user SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer 0 SERVICE X • Enter the service as service Tune Up Home Event Log Reports Help Logout Service O O uGy*m2 level 7 user. Tube Load Component Adjust FD Calibration 0.00 TUI 1 Stand Select the detector for calibration Centering Detector System Adjust Calibrate FD Detector Serial Number Last calibration • Navigate to Repair > Tune-Up Inverter Integrated Detector in Table 123499 23-06-2022 Untersuchung Tube Integrated Detector in Wall stand 1249G6 27-09-2021 O mAs > Detector Calibration. 1 Dose lontomat 1 MAX wi-D 1616:37 31-08-2021 O FD-Calibration Control Fluoro Additional Tools Manual Calibration # 42.6 cm x 42.6 cm Error Suppress List 8 114 cm • Select the particular detector Ist-Wert Soll-Wert Nachverarbeitung ohne ohne and start the Calibration menu with Next. 2 KV 73 2 mAs 20 Dokumentation ms Dosis S mm Cu 0.0 000 Fokus % 80 Bild Extras L After selection, click 'Next" Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 32 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector – Admin user SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer X • Enter the service as clinical Tune Up Home Logout admin user. User Password Component Adjust FD Calibration Centering System Adjust Select the detector for calibration 3D Calibration 1 • Navigate to Repair > Tune-Up > Projection Geometry Calibrate FD Detector Serial Number Last calibration Integrated Detector in Table 123499 23-06-2022 Integrated Detector in Wall stand 1249G6 27-09-2021 Detector Calibration. 1 MAX wi-D 161637 31-08-2021 OO • Select the particular detector and start the Calibration menu with Next. 2 2 After selection, click 'Next" Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 33 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers The calibration switch must be SIEMENS switched on. STISe For Ysio Max systems the calibration switch can be found on the back of the RCM – Exposure box. 1 1 Template Effective Date: 26 Mar 2025 | QR700020784 34 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers • Follow the instructions on the imaging system exactly. Once a condition is fulfilled, the red X changes to a green checkmark. 1 • If all conditions are met, the Next button appears in the status bar. 2 Siemens Healthcare Service Software - Internet Explorer - SERVICE Siemens Healthcare Service Software - Internet Explorer - X SERVICE 0 X Tune Up Home | Event Log Reports | Help Logout uGy*m2 Tune Up Home Event Log Reports | Help Logout Service 0.00 uGy*m2 Service Tube Load Tube Load o Component Adjust FD Calibration FD Calibration 0.00 TUI 1 Component Adjust Stand Detector: Integrated Detector in Table TUI 1 Serial number: 123499 Stand Detector: Integrated Detector in Table Serial number: 123499 Centering Centering Detector Detector System Adjust Step 1 of 4 - check FD parameters for calibration System Adjust Step 1 of 4 - check FD parameters for calibration Untersuchung Untersuchung Tube Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! Tube Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! mAs Please wait while systems verifies detector parameters Einstell. Service mAs Please wait while systems verifies detector parameters Einstell. Service Dose lontomat Dose lontomat FD-Calibration Parameter Ready Current value Remarks FD-Calibration Control Fluoro Control Fluoro Parameter Ready Current value Remarks Additional Tools Additional Tools Manual Calibration Manual Calibration Error Suppress List SID X 114.4 cm Please set SID to 130.0 cm Error Suppress List SID 129.8 cm Nachverarbeitung Nachverarbeitung Grid 1 V Grid 1 Temperature constant x 35.6 ºC Please wait ... (00:23) Temperature constant V 35.7 ºC Dokumentation Dokumentation 2 System is verifying detector parameters ... Manual Back ----- Exit Please click 'Next' to select the calibration mode Manual Back | Next | Exit Template Effective Date: 26 Mar 2025 | QR700020784 35 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer SERVICE X • A status dialog appears when Tune Up Home Event Log Reports Help | Logout Service O uGy*m2 0.00 Next is selected. Tube Load FD Calibration Component Adjust TUI 1 Stand Detector: Integrated Detector in Table Serial number: 123499 Centering Detector System Adjust Step 2 of 4 - select the calibration mode(s) Instructions: Please select the mode you want to calibrate Untersuchung • Select the modes to be Inverter Tube mAs Mode Select Status Progress Reset Einstell. Service calibrated. 1 Dose lontomat FD-Calibration Control Fluoro DR Radiography Last calibration: 23-6-2022 Additional Tools 1 Manual Calibration DR Tomography / Long Time Exposure Last calibration: 23-6-2022 Error Suppress List • To start the calibration, select Nachverarbeitung DO Go. 2 • Follow the instructions on the Dokumentation imaging system. 3 Message from webpage X X-ray will be automatically released during calibration. Ensure that nobody is in the examination room. You can stop the calibration at any time by clicking 'Abort" 3 OK 2 Click 'Go' to start the calibration Manual Back Go Exit Template Effective Date: 26 Mar 2025 | QR700020784 36 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fix RAD detector SIEMENS Healthineers Follow the instructions on the imaging @ Siemens Healthcare Service Software - Internet Explorer X SERVICE Tune Up Home Event Log Reports Help Logout uGy*m2 Service system. Insert 0.6 mm Cu filter and Tube Load FD Calibration 0.00 O O Component Adjust calib fd4343rc 30 table press OK. 1 TUI 1 Stand Detector: Integrated Detector in Table Serial number: 123499 Centering Detector System Adjust Step 3 of 4 - system calibrates selected mode(s) Inverter Instructions: Please wait while the system performs acquisitions Untersuchung The system then will process the acquired images The Cu filter can be found in the blue Tube mAs Dose lontomat Mode Select Status Progress Reset FD-Calibration accessory folder. Control Fluoro Additional Tools DR Radiography Started Preparation, please wait ... Manual Calibration 46.9 cm × 46.9 cm K Error Suppress List DR Tomography / Long Time Exposure Last calibration: 23-6-2022 & 130 cm Ist-Wert Soll-Wert Nachverarbeitung HIT ohne ohne Message from webpage X KV 70 ? Is the following filter inserted: - 0.6 mm Cu MAS 1 Dokumentation Click 'Cancel' to abort the calibration ms 100 Dosis 1 OK Cancel S Healthineers mm Cu 0.0 SIEMENS I Fokus % 80 NOUOVO Bild Extra 1 System is performing acquisitions ... Abort Template Effective Date: 26 Mar 2025 | QR700020784 37 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers X SERVICE • Wait until the calibration is has @ Siemens Healthcare Service Software - Internet Explorer Tune Up Home Event Log | Reports Help | Logout uGy*m2 Service finished. Do not enter the room or 4 3346.0 O User Password FD Calibration Tube Load calib fd3543ezh 31 free Component Adjust interrupt the calibration! TUI 1 Detector: Integrated Detector in Table Serial number: 123499 TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) At the end of the calibration, an Stand Centering Instructions: Please wait while the system performs acquisitions Detector The system then will process the acquired images Examination System Adjust indicator is displayed showing Inverter Mode Select Status Progress Reset Fluoro Rad Tube mAs whether or not the calibration was Dose lontomat FD-Calibration DR Radiography Accepted Finished (not yet saved) Dose DEXI Water Equivalent DR Tomography / Long Time Exposure Started 53 % processing 47.0 cm x 47.0 cm successful. 1 Control Fluoro 8 150 cm Collimation Postprocessing 3D Calibration Current Target Projection Geometry none Reference 10 2D-Topo Calibration Message from webpage 1 X Reference 10 • Press OK to save the calibration Additional Tools Manual Calibration Error Suppress List ? The calibration has finished successfully. KV 70 results. 2 Save the calibration results now? mAs Documentation Click 'Cancel' to discard all calibration results or save them later. ms Dose S 2 OK Cancel mm Cu 0.0 Focus % 80 Image Tools System is processing acquired images ... Abort S/N:2013C6 Ready 2013C6 7%HU O Template Effective Date: 26 Mar 2025 | QR700020784 38 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a fixed RAD detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer X SERVICE • Wait until the results have Tune Up Home | Event Log | Reports | Help Logout uGy*m2 Service FD Calibration 4 1723.0 O been saved. The progress area User Password Tube Load calib fd3543ezh 30 free Component Adjust TUI 1 Detector: Integrated Detector in Table Serial number: 123499 will be updated accordingly. 1 TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) Stand Centering Instructions: Examination Detector System Adjust Mode Select Status Progress Reset Settings Service Inverter • Press Exit to leave the detector Tube mAs DR Radiography Successful Dose lontomat DR Tomography / Long Time Exposure M Successful 1 Saved calibration. FD-Calibration Save 2 Dose DEXI Water Equivalent Control Fluoro Postprocessing Collimation 3D Calibration Projection Geometry Disengage the calibration Reference 10 • 2D-Topo Calibration Reference 10 Additional Tools Message from webpage X switch and reboot the system. Manual Calibration Error Suppress List Calibration results have been saved sucessfully Documentation 2 OK Synchronizing new calibration data, please wait ... Abort S/N:2013C6 Ready 2013C6 10%HU Template Effective Date: 26 Mar 2025 | QR700020784 39 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Please note that this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product-related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration-dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers Headquarters Siemens Healthineers AG Siemensstr. 3 91301 Forchheim, Germany Phone: +49 9191 180 siemens-healthineers.com Template Effective Date: 26 Mar 2025 | QR700020784 40 Job Aid Effective Date: 25 Jun 2025 | QR700025014 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025
- Ysio Max Detector Calibration