Syva® Viva-ProE™ System Weekly Maintenance Job Aid

This job aid provides instructions on how to perform weekly maintenance on the Syva® Viva-ProE™ System.

Weekly Maintenance Rinsing the Probes 1. Fill the 10% cleaning solution bottle on the reagent rotor and the tube in the W position on the sample rotor, with fresh 10% cleaning solution 2. On the Home screen, select the Tasks icon 3. Select Maintenance > Clean Cuvette/Rinse Probe > Probe Rinse Note: This process may take up to 5 minutes to complete HILS#705 Effective date 06/01/2016 © Siemens Healthcare Diagnostics Inc. 2016 All rights reserved. © Siemens AG 2013. All rights reserved. Weekly Maintenance