VERSANT® kPCR Molecular System AD Run Errors Job Aid

This job aid will show you the steps on how to troubleshoot run errors on the VERSANT® kPCR Molecular System. 

VERSANT® kPCR Molecular System AD Troubleshooting SIEMENS Healthineers Run Errors Observation Error Corrective Action Power Supply Loss of power to the instrument. Before starting a new run, check all connections and verify that the kPCR AD QS5 is turned ON. If the problem persists, contact a local technical support provider or distributor. Communication lost while running an experiment Communication failure between the kPCR and the Confirm the communication icon located in the upper Siemens Molecular Workstation. right corner of the MiPLX AD Software Home screen is green. The green icon indicates active communication between the VERSANT® AD and Siemens Molecular Workstation. The VERSANT® AD and Siemens Molecular Workstation must be connected with an ethernet cable. If the problem persists, contact a local technical support provider or distributor. 1 Effective Date: 22 JUL 2022| HOOD05162003277856 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 2 Effective Date: 22 JUL 2022| HOOD05162003277856 Unrestricted © Siemens Healthineers, 2022

  • versant
  • kpcr
  • molecular
  • run
  • errors
  • troubleshooting