Testing Analyzer Water on the Atellica® IM Analyzer Job Aid Version 1.26 and Above
This job aid provides instructions on how to perform the as needed maintenance task of testing analyzer water for the Atellica® IM Analyzer.
Testing the Analyzer Water on the Atellica® IM Analyzer SIEMENS Job Aid Healthineers Testing IM Analyzer Water SIEMENS. Healthineers Atellica Solution 6 LabManager A 1. Put the IM Analyzer in the Diagnostic State: 1 Troubleshooting A. Select Troubleshooting System SH-01 SH-02 CH-01 CH-02 IM-01 CI-01 B. Select Operator Diagnostics ‹ A Troubleshooting 6 LabManager C. Select the appropriate IM analyzer module ? Operator Diagnostics Online Help D. Select Enter Diagnostics > Yes C IM-01 B Operator Diagnostics Subsystems Diagnostic Name Activity Description E. Wait for the module to enter the diagnostic state All Dispense IM Wash and Aspirate Dispense IM wash from wash displacement station and aspirate with probe 3 Perform Ticketing Cuvette Loader Dispense water and aspirate, wash displacement, and Dispense water from water displacement station and aspirate with probe 3 View Status * This procedure is continued on the next slide. Incubation Ring Dispense IM wash and aspirate, port 3, probe 4 Dispense IM wash for port 3 and aspirate with probe 4 Cancel Operator Event Luminometer Dispense IM acid and aspirate Dispense water from port 3 and aspirate with probe 4 Log Module Covers Dispense IM wash and aspirate, resuspend and waste Dispense and aspirate IM acid System Other Dispense water and aspirate, resuspend and waste Dispense IM wash from resuspend stationa nd aspirate with waist probe Messages Log Reagent Open front cover Dispense water from resuspend station and aspirate with waist probe Compartment Open back cover Unlock front cover and disable power from the sample and reagent probes LIS Log Reagent Probes Dry to wet priming Unlock back cover and Disable power from the sample and reagent probes Reagent Loader Sample Tube Cap Color Configuration Create / Extend sample the cap color groups used for sample sorting Assay Troubleshooting Sample Probe Initialize Module Move all module components to their primary positions Tip Loader Clear Thermals Clear system thermal components following a thermal error that caused components to go offline D Turn Mechanics On Enter Diagnostics Template Effective Date: 27-MAY-2019 | HOOD05162003010284 1 Effective Date: 28-JUL-2021 | HOOD05162003208944 Unrestricted © Siemens Healthineers, 2021 Testing the Analyzer Water on the Atellica® IM Analyzer SIEMENS Job Aid Healthineers Testing IM Analyzer Water (continued) 2. Open the right front door 3. Locate the water reservoir and unscrew the lid 2 4. Using a sterile pipette or syringe, pull out a 25-50mL sample of water 5. Screw the lid closed, and close the right front door 6. Follow laboratory protocol and test sample for microbes Note: If the bacteria count is > 50 CFU/mL, contact the local technical 3 support provider to disinfect the water system. 7. Select Exit Diagnostics > Yes 2 Effective Date: 28-JUL-2021 | HOOD05162003208944 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. Copyright © Siemens Healthcare GmbH 2021 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 3 siemens-healthineers.com Effective Date: 28-JUL-2021 | HOOD05162003208944 Unrestricted © Siemens Healthineers, 2021
- attelica
- attellica
- atelica