System Cleanup and Shutdown on the Biograph Vision, mCT and mCT Flow VG7x Job Aid

This step by step procedure will assist the user in activating system cleanup and shutdown on the Biograph Vision, mCT and mCT Flow (VG7x) scanners. This procedure should be completed once weekly to clean up and defragment disks. This procedure may take up to 2-3 hours but can be cancelled at any time.

System Cleanup and Shutdown Biograph Vision, mCT and mCT Flow VG7x Description: This step by step procedure will assist the user in activating system cleanup and shutdown on the Biograph Vision, mCT and mCT Flow (VG7x) scanners. This procedure should be completed once weekly to clean up and defragment disks. This procedure may take up to 2-3 hours but can be cancelled at any time. 1. Exit all examinations and applications and make sure system is in an idle state. 2. Remove all media from the syngo MI Acquisition Workplace and syngo MI Workplace computers. 3. Shut down the syngo MI Workplace computer, if present. a. Select System > End. b. Select Shutdown System. End Session X Log Off ... Shutdown System .. Restart System ... Restart Application .. Standby Cancel Help c. Click Yes to the confirmation dialog. 4. On the syngo Acquisition Workplace console, select System > Run > System Cleanup and Shutdown. Effective Date: 08/04/2018 | HOOD05162002923738 SIEMENS siemens.com/healthineers Healthineers . System Cleanup and Shutdown Biograph mCT and mCT Flow Effective Date: 08/04/2018 | HOOD05162002923738 5. After several seconds, the following window will display. System Clean Up and Shutdown The system clean up may take a few hours. After clean up the system wil automaticaly shut down. You may stop the system cleanup at any time by using the "Cancel" button. The system will then shut down. Remember to turn off the power for the complete system at the control box buton and then proceed with the normal start up procedure. Start Cancel 6. Click the Start button to start the cleanup and shutdown procedure. 7. After the cleanup is complete, the CT scanner, syngo Acquisition Workplace, IRS, and MARS will shutdown. 8. After the scanner is turned off, wait 5 minutes. 9. Turn on the power to the CT scanner and computers by pressing the System On button on the rear of the control box. Siemens Healthineers Services Page 2 of 4 System Cleanup and Shutdown Biograph mCT and mCT Flow Effective Date: 08/04/2018 | HOOD05162002923738 10. The system is ready for scanning when all the disk icons in the lower right corner are visible and show adequate space remaining. 73% 26% 11. Proceed with daily quality control procedures. Siemens Healthineers Services Page 3 of 4 System Cleanup and Shutdown Biograph mCT and mCT Flow Effective Date: 08/04/2018 | HOOD05162002923738 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively ‘‘Operator Manual’’) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively ‘‘Functionality’’) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2018 ………………………………………………………… Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers Siemens Healthineers Services Page 4 of 4