CN-3000 and CN-6000 Systems Replace the Piercer Job Aid

CN-3000 and CN-6000 Systems Replace the Piercer Job Aid

This job aid describes how to peform the as needed maintenance task, replace the piercer, on the CN-3000 and CN-6000 Systems.

CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace the Piercer Job Aid Replace the Piercer – As Needed Maintenance 6 00 00 1. Refill the rinse tank and discard any waste fluid Menu Reagent Calib. curve QC chart Order calib. Order QC Order STAT Joblist Maintenance Shutdown Rinse probes Execute Manual maint. Comment 2. Verify the operation cover status display LED is 2021/04 solid green and open the operation cover Routine maintenance Replace the piercer Sat Sun Mon 3. Set Owren’s Veronal buffer into holder 6 of the Daily/(weekly) 17 18 19 20 Rinse probes Press the [Execute] button to move the piercer to the replacement position. buffer table Discard used cuvettes 4. Close the buffer table cover and operation cover Discard waste fluids . Verify the instrument status is Warming Up, Check and discard the trap chamber fluid V Buffer has been added to holder B-6 of the buffer table. 5. Shutdown Ready or Not Ready and all covers to the main 7 Clean the instrument (weekly) The operation cover, reagent table cover, buffer table cover, unit are closed Clean the rinse tank (weekly) dispensing table cover, and STAT cover have been closed. 6. Press Maintenance > Execute > Replace the Monthly Piercer Clean the filters 8 Execute Cancel 7. Confirm the previous steps by adding checkmarks to the Replace the piercer dialog box 8. Press Execute > OK Note: Replace the piercer when the number of piercing operations reaches 120,000. *Procedure continued on the next slide Note: Siemens recommends to always have one piercer (Probe (PM) No. 6) in stock. 1 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace the Piercer Job Aid Replace the Piercer (continued) – As Needed Maintenance 14. Attach the piercer protection cover to the piercer 9. Turn off the main power switch 15. Disconnect the piercer’s cord connector 10. Open the operation cover 16. Loosen the tube connection screw and remove the tube 11. Remove the sample arm cover 12. Attach the piercer maintenance jig to sample arm A 15 13. Secure the piercer maintenance jig 13 16 PUSH *Procedure continued on the next slide 2 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace the Piercer Job Aid Replace the Piercer (continued) – As Needed Maintenance 19. Remove the piercer 17. Using a screwdriver, remove the piercer fixing screws 20. Discard the piercer and piercer protection cover per 18. Remove the stopper laboratory protocol 18 19 *Procedure continued on the next slide 3 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace the Piercer Job Aid Replace the Piercer (continued) – As Needed Maintenance 23 21. Push the new piercer into the locking unit 22. Reattach the stopper 23. Using a screwdriver, reattach the piercer fixing screws 24. Reinstall the tube and tighten the tube connection screw 25. Reconnect the piercer’s connector 26. Remove and discard the new piercer’s protection GA cover per laboratory protocol 27. Remove the piercer maintenance jig 28. Reattach the sample arm cover 29. Close the operation cover *Procedure continued on the next slide 4 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace the Piercer Job Aid Replace the Piercer (continued) – As Needed Maintenance 00 00 30. Close the operation cover Menu Reagent Calib. curve QC chart Order calib. Order QC Order STAT Joblist Maintenance Shutdown Rinse probes Execute Manual maint. Comment 31. Refill the cuvettes and discard any used 2021/04 cuvettes 32. Start the instrument Routine maintenance Replace the piercer Sat Sun Mon Daily/(weekly) 19 33. Verify all procedures have been performed Rinse probes 33 Check that all procedures have been completed according to 17 18 NE the checklist included in the piecer replacement kit. . O by adding a checkmark to the Replace the Discard used cuvettes . Discard waste fluids piercer dialog box Close the operation cover, reagent table cover, buffer table cover, dispensing Check and discard the trap chamber fluid table cover, and STAT cover. Touch the [Execute] button to adjust the piercer 34. Press Execute Shutdown position. Clean the instrument (weekly ) To replace the piercer again, touch the [Power off] button and replace the • Piercer position adjustment is performed Clean the rinse tank (weekly) 35. Press Yes piercer. Monthly • Maintenance is automatically Clean the filters documented and piercer cycle counts are 34 Execute Power off reset 5 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 6 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021

  • cn-3000
  • cn-6000
  • sysmex
  • cn