CN-3000 and CN-6000 Systems Replace Consumables Job Aid

CN-3000 and CN-6000 Systems Replace Consumables Job Aid

This job aid describes how to replace consumables on the CN-3000 and CN-6000 Systems.

CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace Consumables Job Aid Replenish the Rinse Water 1. Remove the rinse tank cap 2. Verify there are no impurities in the tank • If impurities are present, wash the inside of tank and float switch with tap water, followed by rinse water 3. Fill the tank with rinse water 4. Reattach the cap 5. Verify tubing is not kinked 6. Resume analysis if it was suspended • Verify the instrument status is Int. Ready then press the Start button 1 Effective Date: 18-Jun-2021| HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace Consumables Job Aid Sysmex CN-COAGWASHER 1300A2700649 2Lx1 15 CE Replace CN-COAGWASHER 1. Remove the CN-COAGWASHER connected to the main unit from the fixture apparatus 2. Remove the cap of the new CN-COAGWASHER 3. Remove the cap of the old CN-COAGWASHER and take out n is used up DDIDI PT B2 the tubing ATBM OVB 4. Place the tubing into the new CN-COAGWASHER and B4 OVB reattach the cap DOIDI ATB 5. Scan the upper barcode on the new CN-COAGWASHER with O the 2D barcode reader - 6. Press Register 7. Set the CN-COAGWASHER into the fixture apparatus 8. Verify the tubing is not kinked 9. Resume analysis if it was suspended ROWUSAS • Verify the instrument status is Int. Ready then press the CN- COAGWASHER Start button Note: To manually register consumable information, press sysmex Reagent/consumables > Settings > Register consumables info. CN-COAGWASHER Enter the required consumable information, then press Register. 2 Note: CN-COAGWASHER is displayed as CN-Wash on the software. Effective Date: 18-Jun-2021| HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace Consumables Job Aid Replace CA CLEAN II 1. Lift the fixture apparatus and pull forward the CA CLEAN II connected to the main unit 2. Remove the cap of the new CA CLEAN II 3. Remove the cap of the old CA CLEAN II and take out the tubing 4. Place the tubing into the new CA CLEAN II and reattach the cap 5. Scan the upper barcode on the new CA CLEAN II with the 2D ACLEAN'S barcode reader .............. 6. Press Register 7. Set the CA CLEAN II into the fixture apparatus CONSULTE CON EL MÉDICO. INGERIDO/SI HAY CONTACTO CO LAVE BIEN CON AGUA, EN CASO D' SI HAY CONTACTO CON LA PIEL! EVITAR EL CONTACTO CON LA PIEL Y ADVERTENCIAS Y PRECAUCIONES: NO "" GLI OCCHI: CONSULTARE UN MEDICO. IN CASO DI INGESTIONE E/O CONTATTO CO LAVARE IMMEDIATAMENTE CON ACQUA. IN CASO DI CONTATTO EVITARE IL CONTATTO CON OCCHIEO EPIDERMIS AVVERTENZE E PRECAUZIONI: NON INGERIRE CONSULTER SANS TARDER UN MÉDECIN. D'INGESTION OU DE CONTACT AVEC LES YEUX; LAVER ABONDAMMENT ÀL'EAU, EN CAS LA PEAU OU LES YEUX, EN CAS DE CONTACT: UNDMORENASSE NE PAS AVALER EVITER TOUT CONTACT AVEC AVERTISSEMENTS ET MESURES DE PRÉCAUTION: UND/ODER AUGENKONTAKT: ARZT HINZUZIEHEN. 8. Verify tubing is not kinked MIT VIEL WASSER SPÜLEN. NACH VERSCHLUCKEN AUGEN VERMEIDEN. NACH KONTAKT SOFORT NICHT EINNEHMEN, KONTAKT MIT HAUT UND WARNHINWEISE UND VORSICHTSMASSNAHMEN: IN CASE OF INGESTION AND/OR EYES CONTACT. WATER IMMEDIATELY. CONSULT WITH PHYSICIAN IN CASE OF CONTACT; FLUSH WITH PLENTY OF DON'T INGEST. AVOID SKIN AND EYES CONTACT. WARNINGS AND PRECAUTIONS: 9. Resume analysis if it was suspended • Verify the instrument status is Int. Ready then press the Start button Note: To manually register consumable information, press Reagent/consumables > Settings > Register consumables info. Enter the required consumable information, then press Register. Note: CA CLEAN II is displayed as Clean II on the 3 software. Effective Date: 18-Jun-2021| HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Replace Consumables Job Aid Add Cuvettes 1. Open the cuvette feeder cover on top of the main unit 2. Add cuvettes up to the fill line • Do not overfill cuvettes or a jam may occur 3. Close the cuvette feeder cover oc et 4 Effective Date: 18-Jun-2021| HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Innovin and all related marks are trademarks of Siemens Healthineers. Any other trademarks are the property of their respective owners. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 5 Effective Date: 18-Jun-2021| HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021

  • cn-3000
  • cn-6000
  • sysmex
  • cn