CN-3000 and CN-6000 Systems Register a New Lot of QC Job Aid
This job aid describes how to register a new lot of QC and set the target/limit values on the CN-3000 and CN-6000 Systems.
CN-3000 and CN-6000 Systems SIEMENS Healthineers Register a New Lot of QC Job Aid Register a New Lot of QC with the 2D EY 00 00 Barcode Reader C chart Order calib. Order QC Order STAT Joblist Maintenance Shutdown Add Edit Search Sort Barcode Reagent type Lot no.: Exp. date: Manag. ID Assay group Assay parameter ISI value MNPT 1. Press Settings > Reagent lot master on the Menu Value Units Factor 504146 2021/05/19 2. Press Barcode on the Toolbar Factor 504147 2021/05/18 Calibrator 503271 2021/05/18 3. When the dialog box appears, scan the 2D barcode Calibrator 503278 2022/01/23 Calibrator 503281 2021/12/05 from the QC Table of Assigned Values sheet Calibrator 503282 2021/09/14 Calibrator 503283 2021/07/07 • Assay parameter values associated with the lot Control 507775 2021/05/20 Control 507780 2024/00/10 number of control material are imported Control 507782 20 Barcode Reading of Reagent Lot Information Factor 547625 20 Read the 2-D barcode to register the reagent lot Factor 547699 20 information in the reagent lot master. Factor 560801 20 Factor 560802 20 The barcode can be read even if this dialog box is Reagent 506800 20 closed. Reagent 539726 20 Reagent 539764 20 Reagent 563813 20 Close Reagent 563848 20 Reagent 563851 2021/08/19 Reagent 563873 2024/11/18 Reagent 563879 2021/06/02 Control 556720 2021/09/28 Reagent 546350 2022/01/23 Reagent 546351 2021/12/11 Reagent 546354 2022/03/17 Reagent 552920 2021/11/20 Reagent 552921 2022/01/23 Reagent 557800 2021/05/13 Switch Reagent 550200 2021/05/13 K 1 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Register a New Lot of QC Job Aid Manually Register a New Lot of QC Register new lot Control name:Citrol 2 1. Press QC chart on the Menu 2. Highlight the desired QC chart Lot no.: Exp. date: 3. Press Reg new lot on the Toolbar 4. Select the appropriate QC file, if necessary 548292 28 / 04 / 2022 5. From the Register new lot dialog box, select the lot number of the QC material to be registered Target/limit settings: • If the barcode ID for the entered lot number is not listed, O Setting off scan the 2D barcode on the QC Table of Assigned Values sheet OR manually enter the new barcode ID by selecting Use setting from current Add from the Lot no. dropdown 6. Verify the expiration date is correct Load from reagent 7. Determine the Target/limit settings: lot master • Setting off – no target/limit settings will be set up • Use setting from current – target/limit settings from the O Manual entry “current” QC lot will be used • Load from reagent lot master – load the QC target/limit settings from the Reagent lot master OK Cancel • Manually entry – manually enter the QC target/limit settings 8. Press OK 2 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Register a New Lot of QC Job Aid Set the Target/Limit Values Target/limit settings Citrol 2 548292 1. Press QC chart on the Menu Assay parameter Upper Upper Target Lower Lower 2. Select the appropriate QC file stop limit limit stop 1SD 3. Press Tar/Limit on the Toolbar PT I sec 33.2 28.9 24.6 PT I INR 4. Set the target/limit values by selecting a method 3.26 2.91 2.56 PT I C sec 33.2 28.9 24.6 below: PT I C INR 3.26 2.91 2.56 APTT25 sec 48.8 43.6 38.4 • Scan – scan the 2D barcode on the QC Table of Assigned Values sheet APTT25 R. 1.95 1.74 1.53 • Reagent lot master – select the assay and press Load from reagent lot master • Manually – select the assay and press Change settings • Calculate automatically – select the assay and Change Load from reagent settings Auto calculation lot master press Auto calculation 5. Press Register Detailed conditions Register Close 3 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 4 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021
- cn-3000
- cn-6000
- sysmex
- cn