CN-3000 and CN-6000 Systems Process STAT Samples Job Aid
This job aid describes how to perform STAT analysis on the CN-3000 and CN-6000 Systems.
CN-3000 and CN-6000 Systems SIEMENS Healthineers Process STAT Samples Job Aid Perform STAT Analysis 1. Verify the STAT sample holder cover status display LED is solid green 2. Open the STAT sample holder cover 3. Set the sample in the STAT sample holder with the barcode facing out 4. Close the cover CN-6000 5. Press the Start button • The analysis of normal samples is interrupted and STAT analysis begins • Analysis progress and analysis results can be reviewed on the Joblist screen 6. Wait until the STAT sample holder cover status display LED returns to solid green 7. Open the STAT sample holder cover 8. Remove the analyzed sample and close the cover • The analysis of normal samples resumes Note: If the STAT order has not be obtained from the host computer, manually register the order by pressing STAT on the Toolbar. 1 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Innovin and all related marks are trademarks of Siemens Healthineers. Any other trademarks are the property of their respective owners. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 2 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021
- cn-3000
- cn-6000
- sysmex
- cn