CN-3000 and CN-6000 Systems Order a Calibration Job Aid

CN-3000 and CN-6000 Systems Order a Calibration Job Aid

This job aid describes how to order a calibration on the CN-3000 and CN-6000 Systems.

CN-3000 and CN-6000 Systems SIEMENS Healthineers Order a Calibration Job Aid Order a Calibration 1. Press Order calib. on the Toolbar Order 2. Select the method of calibration curve creation: • Create by using a new reagent lot: Create a calibration curve. Select the method of creation. • Use a new reagent lot to analyze all points New Edit • Use when a reagent lot is replaced • Create by selecting an assay group: Full calibration Replace points • Use an analysis parameter from an existing Create calibration curve by analyzing Analyze calibration curve points of the specified calibration calibration curve, and re-analyze all points all calibration curve points. curve and replace them. • Use when performing calibration curve analysis again • Replace points Create by using a new reagent lot Correct zero point • From [New] or [Current], specify the point to be Create the calibration curve for reagent lots with [No calib. curve]. Create the calibration curve with dilution ratio 0/1 point a replaced, and analyze only the concerned point correction criteria for a validated calibration curve. • Correct zero point Create by selecting an assay group • Create a calibration curve from [Current] or [Old]. Select the assay group and create a calibration curve. Analyze only 1 point with the dilution ratio of 0/1 from an existing, validated calibration curve. Create a calibration curve parallel to the existing Cancel calibration curve 1 *Procedure continued on the next slide Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Order a Calibration Job Aid Order a Calibration (continued) Order New lot 1 Select calibrator 2 Register order 3. Follow the Order dialog boxes and press Register to complete the order Enter the information used for creating the calibration curve and touch the [Next] button. Read the calibrator barcode if available. 4. Set the calibrator into the sample rack according to the dialog box image • If the calibrator is set onto the reagent table, proceed to step 6 Order Assay group Reagent lot group Parameter Calibrator Lot no Exp. date Assay sheet value PT INN 5. Set the sample rack into the right sampler pool PT INN Cal 549733 PT IC INR PTMult1 6. Press the Start button PT INN Cal PT INN 549733 PT IC INR PTMult2 • Calibration curve analysis starts. To view the V PT INN Cal PT INN 549733 PT IC INR PTMult3 calibration status and progress, press Joblist V PT INN Cal PT INN 549733 PT IC INR PTMult4 PT INN Cal PT INN 549733 PT IC INR PTMult5 V PT INN Cal PT INN 549733 PT IC INR PTMult6 Show all reagent lot groups with [No calib. curve]. Next Cancel 2 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 CN-3000 and CN-6000 Systems SIEMENS Healthineers Order a Calibration Job Aid Edit calibration curve PT % PT R. PT INR PT DFbg Load value from reagent lot master Normal value: 12.0 7 8 9 Calculate normal value from calibration curve 4 5 6 Enter the ISI and Normal Value 1 3 Enter O 1. Press Calib. curve on the Menu 2. Select the desired analysis parameter 3. Select the reagent lot group 4. Press Edit in the Toolbar Exp. date: 5. If loading the normal value/ISI value from the Reagent lot master, press Load value from reagent lot master Load a calibration curve OK Cancel • Verify the PT lot and ISI value has been entered into the Edit calibration curve Reagent lot master with the 2D barcode from the PT ISI PT THS~% PT THS~R. PT THS~INR DFbgTHS C. value sheet XY point setting: Normal value: 6. To manually enter the normal value/ISI value: No. % sec 7 8 If creating a ratio calibration curve, enter your 25. 28.0 • 50.0 18.0 ISI value: 100.0 13.0 4 5 6 laboratory’s established normal value 4 If creating an ISI input calibration curve, enter your 5 3 • laboratory’s established normal value and ISI value Enter 8 O 9 7. Press OK 10 11 12 Note: ISI value with 2D barcode is found on the table Exp. date: of assigned values sheet for Innovin® reagents. PT normal patient mean is established by laboratory. Load a calibration curve OK Cancel These are required for every new lot of Innovin 3 reagent. Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Innovin and all related marks are trademarks of Siemens Healthineers. Any other trademarks are the property of their respective owners. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 4 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021

  • cn-3000
  • cn-6000
  • sysmex
  • cn