CN-3000 and CN-6000 Systems Discard Used Cuvettes Job Aid
This job aid describes how to peform the daily maintenance task, discard used cuvettes, on the CN-3000 and CN-6000 Systems.
CN-3000 and CN-6000 Systems SIEMENS Healthineers Discard Used Cuvettes Job Aid Discard Used Cuvettes – Daily Maintenance 1. If the instrument is on, verify the operation cover status display LED is solid green 2. Pull out the cuvette trash box 3. Discard used cuvettes 4. Clean the cuvette trash box with water, if dirty 5. Return the cuvette trash box to its original position 6. Document maintenance Note: Document manual maintenance by pressing Maintenance > Manual maint. > Register manual maintenance. 1 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021 SIEMENS Healthineers Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. Not available for sale in the U.S. Future product availability cannot be guaranteed. © Siemens Healthineers, 2021. All rights reserved. Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens.com/healthineers 2 Effective Date: 18-Jun-2021 | HOOD05162003199986 Unrestricted © Siemens Healthineers, 2021
- cn-3000
- cn-6000
- sysmex
- cn