Symbia Pro.specta VA10 Copying CT Volumes
Symbia Pro.specta VA10 Copying CT Volumes Job Aid
Copying CT Volume Parameters Symbia Pro.specta™ All Symbia Pro.specta™ Systems VA10 This Job Aid is designed to assist the end user in copying the CT Volume range to subsequent reconstructions in the syngo Acquisition Workspace. Note: This procedure may be necessary when there are multiple CT reconstructions pre-built into the acquisition. Copying CT Volume Range prior to starting the topogram 1. In the syngo Acquisition Workspace, select the Reconstruction panel. SPECT-CT [factory] ( ... fa Patient position HF supine HF prone HF right HF left FF supine FF prone FF right FF left Tomo Topogram . CT-AC Reconstruction Close Reconstruction SIEMENS Healthineers Copying CT Volume Parameters 2. In the Recon Ranges section, right-click on the 1st CT recon range and select Copy Recon Volume Parameters. Recon Ranges Show Parameters Tomo Delete 2D 3D 3D AC NAC Repeat lomo M auto f + lauto n lauto n Copy All Recon Parameters Topogram Copy Recon Volume Parameters Paste All Recon Parameters Paste Recon Volume Parameters CT-AC Create Image with Reference Lines Send to CT View&GO Edit Result in CT View&GO auto auto CT-AC 3.00 Qr40 S3 Slice Thickness 3.00 Increment 3.00 Image Orientation Axial Kernel Qr40 Window@120kV Abdomen 3. Right-click on any subsequent CT recon(s) and right-click and select Paste Recon Volume Parameters. This will match the exact scan range from the first reconstruction. Recon Ranges Torna 20 O NAC Toprogram CT-AC Show Parameters Repeat Copy All Recon Parameters Copy Recon Volume Parameters Paste All Recon Parameteis Paste Recon Volume Parameters Create Image with Reference Lines Send to CT VASCO Edit Result in CT Viu&GO SIEMENS Page 2 of 5 Healthineers Copying CT Volume Parameters Copying CT Volume Range after completion of the topogram 1. For scans with multiple CT reconstruction ranges, the volume ranges may not be identical. After completion of the topogram: • Once the topogram is complete navigate to the Recon Ranges area • De-select the 3D AC Tomo Recon box, as it may be selected by default, by holding the control key while single-clicking. Recon Ranges Ctrl Tomo 3D NAC auto Topogram CT-AC auto auto auto 2. In the Recon Ranges section, right-click on the 1st CT recon range and select Copy Recon Volume Parameters. Recon Ranges Show Parameters Tomy Delete NAC Repaal Copy All Recon Parameters Copy Recon Volume Paramelels Paste All Recon Parameters Paste: Recon Volume Parameters CT-AC Croate Image with Reference Lines Send to CT VW&GO Edi Resul in CT Vc-4GO CT AC 3 00 Or40 53 Bor Thickness 3.00 Increment 3.00 And Windowd 120kV Abdomen SIEMENS Page 3 of 5 Healthineers Copying CT Volume Parameters 3. Right-click on any subsequent CT recon(s) and right-click and select Paste Recon Volume Parameters. This will match the exact scan range from the first reconstruction. Recon Ranges Toma 20 NAC aula Topogram ST-AG Show Parameters Repeat Copy All Recon Parameters Copy Recon Volume Parameters Paste All Recon Parameleis Paste Recon Volume Parameters Create Image with Reference Lines Send to CT MIAMIGO Edit Result in CT Vin&GO SIEMENS Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: 30 Mar 2023 | HOOD05162003339395 Page 4 of 5 Copying CT Volume Parameters Please note this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (here- inafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2023 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com SIEMENS Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date 10 Aug 2021 | HILS 2224 Healthineers Page 5 of 5