SOMATOM On.site System Checkup and Daily QA Video

In this video, you will learn about the system checkup and daily QA necessary for the SOMATOM On.site scanner.

Target group: Radiographers, physicists.
Audio: Yes.

Recommended to be viewed on the following devices: Tablet, laptop, desktop computer (sufficiently large display required).

Somatom on site system checkup and daily QA To ensure that the image quality is stable even under variable environmental conditions, you must perform quality tests regularly. It's important that you perform a system checkup once you start the system and that you run a daily quality test. The daily quality test is not mandatory in all countries. The system checkup will only take approximately 5 to 6 minutes. This calibration is important when you have powered off the system, for example, overnight or between working shifts. To start the system checkup, select Checkup from the Settings menu. The system check window opens. Tap the Go button. When prompted, press the Start key on the gantry operator panel. Scanning is initiated and the checkup procedure starts. Complete the calibration when the room temperature has changed by more than five degrees or when the room humidity has changed by more than 5%. The the daily quality test is automated. You can perform the whole test according to the guidance on the screen. Press the cover open key on the gantry operator panel to move the front cover to the patient load position. This gives you more space to place the phantom set. Mount the phantom set on the support column with the appropriate phantom holder bracket. Tighten the single wing nut lever to hold the Phantom securely in place. Once the Phantom is securely attached to the holding bracket, press and hold the Cover close key on the gantry operator panel to move the front cover back to the scan start position. To start the daily quality test, tap the Settings icon and from the menu select Daily QA. The the Daily QA window is displayed. Tap Go to initiate the daily quality test. The homogeneity, water, and noise tests are respectively performed. The pixel noise of the images is calculated as a standard deviation. You can see the test results in the content area of the Daily QA window.

1% 14% 28% 42% 56% 60% 65% 72 72% 74% 76% 78% 80% 82% 23 111 25% 03/14/2024 250.0 500 100 50% 75% SIEMENS Healthineers SOMATOM On.site System Checkup and Daily QA Effective Date: 28 Nov 2024 | QR700012537 Unrestricted @ Siemens Healthineers, 2024 SIEMENS . Examination Touch UI System Check Restart Lock Daily QA Configuration Panel Administration Portal Administration Portal [EN] Clinical Configuration Checkup Resume Calibration syngo Remote Assist Screen recorder SaveLog Remote Service Exam Designer Create Screenshot Fast Contact File Browser Load Update Pack Direct Laser Check myNeedle Laser Check Scan Recon Starting Application ... Resuming was successful. EN 7:10 PM Close Confirm GO Cancel For safety reasons, please stand still in front of this control panel. Run the Checkup procedure to ensure optimum image quality and system safety. Do not move away from the control panel until the Checkup procedure is completed. X-rays will be produced during the Checkup procedure. Checkup has started Repeat Press the GO button to continue. Checkup @ System Check in progress ... Workflow 'Checkup @ System Check' started. System Config Check @ Checkup @ System Check in progress ... Function 'System Config Check @ Checkup @ System Check' completed. Pixel Quality Equalization @ Checkup @ System Check in progress ... Function 'Pixel Quality Equalization @ Checkup @ System Check' started. Gantry accelerating 120 kV, 10 mA, 1 s, 32x0.75 mm (A) Sn120 kV, 20 mA, 2 s, 6x0.75 mm (A) Safety System Self-Test @ Checkup @ System Check in progress ... Function 'Safety System Self-Test @ Checkup @ System Check' started. 120 kV, 13 mA, 0.8 s, 32x0.75 mm (A) Scan Checkup succeeded. Function Tube Event Listener @ Checkup @ System Check' completed. Workflow 'Checkup @ System Check' completed with status OK. Next 몸목 5,6 k Savelog Recon Daily QA has started. Position system phantom at the gantry center. The inner lightmarker must hit the reference position of the phantom. Click GO when ready. Service, Local Phantom Position Check @ Daily QA @ System Check in progress ... Function 'Phantom Position Check @ Daily QA @ System Check' started. Exit the scan room. The table will move automatically. X-rays will be produced during the Daily QA procedure. 1 0 Scan Re Siemens Service Service 11602012_175605 syngo CT VB10A 2:13:56.77 PM ?AC 101 10cm KV 120 GT 0.0 an Recon Siemens Quality 2:14:06.04 PM 102 IMA 3 TP 9.1 5cm TP 0.9 Quality 11602012_175605 HFS 2:14:16.29 PM 104 IMA 5 220.0 0/0 03/1 4/2024 2:14:23.90 PM 105 IMA 3 Daily QA succeeded. Function 'Noise @ Daily QA @ System Check' completed. Workflow 'Daily QA @ System Check' completed with status OK. Please note this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of utility model or design, are reserved. Unrestricted | Published by Siemens Healthineers AG | C Siemens Healthineers AG, 2024 Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com

  • ct
  • vb10
  • onsite
  • site
  • system
  • checkup
  • daily
  • qa
  • quality
  • test
  • phantom
  • calibration