Service Plug

The video shows you how to use the service plug.
Target group: Basic users.
Duration: 4 min | Audio included.

Perform the following test tools. Start the service plug test for LC plug one and LC plug 2. Open the lid underneath the patient table and take out the service plug. Connect the service. Plug Tequil socket 1. Start the LC plug one test with OK. Retract the upper part is shown to unlock the service plug. Notice damage of the quill connection. Do not disconnect a coil plug without releasing the locking mechanism. Retract the upper part as shown to unlock the service plug. Disconnect the service plug from coil socket 1. Connect the service plug to quail socket 2. Start the LC plug two test with OK. Unlock and disconnect the service plug from Quill socket 2 as shown before. Store the service plug underneath the patient table. Confirm the finished test status with OK.

SIEMENS I s Healthineers T Assu Assu Final Steps SIEMENS System Status Overview Customer TestTools - General Final Steps Healthineers SIEMENS Healthineers Please note that the learning material is for training purposes only. Plug ce02:55 0K ce02:55 '32347 0K Control System ox 0K For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute Magnet the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version Of the software and hardware available at the time Of the training. 1 Test Test 2 Test NOTIC The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. SIEMENS Plug Co" 1 Plug 1 Plug 2 Plug Co" 2: PI it to Patient if fnish a" it to Patient fnish a" to socket it to it to if to socket if Transmit System System Please note: Some functions shown in this material are optional and might not be part of your system. u PR s u 2 u 10 s u s Demage of the Coil Connection! Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Healthineers Do not disconnect a Coil Plug without releasing the locking mechanism. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders @ Siemens Healthcare GmbH, 2021 - Retract the upper part as shown to unlock the Service Plug. System Check myExam Cockpit Examination Lock will be liable for damages. Siemens Healthcare GmbH provides this documentation "as is" without the assumption of any liability All names and data of patients, parameters and configuration dependent designations are fictional and examples only. under any theory of law. The content described herein shall be used by qualified personnel who are All rights, including rights created by patent grant or registration of a utility model or design, are reserved. employed by Siemens Healthcare GmbH or one of its affiliates or who are otherwise authorized by Siemens Healthcare GmbH or its affiliates to use such documents. Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens Healthcare GmbH or one of its affiliates are not entitled to use this documentation without prior written authority. Disconnect the Service Plug from Coil Socket 1. - Retract the upper part as shown to unlock the Service Plug. Unlock and disconnect the Service Plug from Coil Socket 2 Open the lid underneath the Patient Table and Connect the Service Plug to Coil Socket 1. Start the LC Plug 2 Test with OK. Start the Service Plug Test: take out the Service Plug. Store the Service Plug underneath the Patient Table. Confirm the finished test status with OK. Perform the following Test Tools. - LC Plug 2 - LC Plug Make sure you use the latest version. take out the Service Plug. as shown before. 10005 : 501 10005: sol Customer 10005: sol 10003051 10005: 501 10002:551 2 Plug 2 0K 0K

  • free.max
  • freemax
  • free max
  • service plug
  • plug freemax
  • plug max
  • mri service plug
  • mr plug
  • service support
  • coils problem
  • coils plug
  • coil plug
  • coil socket
  • service
  • freeMax service
  • remote diagnosis
  • FreeMax Service Plug
  • coil socket issue