Sensis Vibe VD15C Respiration rate and EtCO2 (end-tidalCO2)

Learn how to configure and visualize the respiration rate and EtCO2 value.

Target group: Basic users
Recommended to be viewed on the following devices: All (incl. tablet, smartphone)

SIEMENS Healthineers Respiration rate and EtCO2 (end-tidal CO2 ) Sensis Vibe VD15C Template Effective Date: 10 Aug 2021 | HILS 2223 1 Job Aid Effective Date: 09 Sep 2024 | QR700010823 Unrestricted I Published by Siemens Healthineers AG I © Siemens Healthineers, 2024 Table of contents SIEMENS Healthineers Method 3 Configuration 4 Visualization 6 Visualization configuration 7 MicroPod TM */NanoPod * measurement TM 9 SIEMENS 2 * The product names and/or brands referred to are the property of their respective Effective Date: 09 Sep 2024 | QR700010823 trademark holders. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 Method SIEMENS Healthineers The patient’s respiration rate and EtCO2 can be measured by using the MicroPod * or TM NanoPod TM * provided by Medtronic. Oridion etCO, Microstream2 Medtronic Microstream CO2 Microstream TM MicroPod TM Microstream TM NanoPod TM External EtCO2 module External EtCO2 module 60 EtCO2 32 30 Resp 25 3 * The product names and/or brands referred to are the property of their respective Effective Date: 09 Sep 2024 | QR700010823 trademark holders. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 Configuration SIEMENS Healthineers No breath timeout can be configured in the Vital Signs tab card of the #f Sensis Vibe - Configuration Panel X File Event Configuration dialog box. × Before starting a study, you can override the system default. These changes will be applied as the new system default, even after restarting Auto Delete Patient Certificate Transfer Patient Software Browser Handler Renistration Distribution the system. Events Configuration Tachycardia Bradycardi Vital Signs tal Signs Al ... Mark Inherit CO Parameters Events Ablation To configure No breath timeout, go to Options > Configuration, and open HR (averaged) Configuration Configuration SpO2 the Events Configuration dialog: NBP Respiration Rate IBP EtCO2 On the Vital Signs tab, select the relevant Respiration Rate Source. IBP Channel 12 Lead ECG Stimulation • P1 P2 P3 . P4 Snapshot Configuration • Confirm the setting. When Vital Signs Event is created, prompt for: Consciousness Muscle Function Pain Score Respiration Rate Source Respiration Circulation 02 Saturation Sedation Assessment No breath timeout For neonatal 15 s For adult 30 s Measure vital signs: Repeatedly Log interval (min) 10 Which license to use for calculating respiration rate NBP By default, only use neonatal mode . CO2 Module ECG Respiration Min age to use adult mode (months) 24 Note that you should not use NBP adult mode on neonates or patients who weigh less than 8.2 kg (18 pounds). • No breath timeout: If the system does not detect any breath for When examining emergency patient, use: the period specified here, the system will treat this as no breath NBP neonatal mode . NBP adult mode and display the respiration rate as — instead of 0. Respiration Rate Source No breath timeout For neonatal 15 s For adult 30 s • CO2 Module: Use the MicroPod TM * or NanoPod TM * provided by Which license to use for calculating respiration rate . CO2 Module . ECG Respiration Medtronic. OK Cancel 4 * The product names and/or brands referred to are the property of their respective Effective Date: 09 Sep 2024 | QR700010823 trademark holders. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 Configuration SIEMENS Healthineers No breath timeout can be modified during the study, and Vital Signs Settings × Vital Signs Alarms it is valid until the study is closed. Setting for current examination Effective directly when pressing OK. Logging Parameters To modify No breath timeout, open the Vital Signs HR (averaged) SpO2 Settings dialog, and do the following: NBP Respiration rate IBP EtCO2 IBP Channel 0 P1 P2 P3 P4 1. On the Vital Signs tab, select the relevant Respiration When Vital Signs Event is created, prompt for : Rate Source. Consciousness Muscle Function Pain Score Respiration Circulation 02 Saturation Sedation Assessment 2. Confirm the setting. Log interval (min) 10 min NBP For the current examination, use: Respiration Rate Source O NBP neonatal mode NBP adult mode No breath timeout For neonatal For adult 30 s Note that you should not use NBP adult mode on neonates or patients who weigh less than 8.2 kg (18 pounds). License to use for calculating respiration rate Respiration Rate Source CO2 Module ECG Respiration No breath timeout For neonatal For adult 30 s @ CO2 Module O Lead I O Lead Il License to use for calculating respiration rate CO2 Module ECG Respiration o CO2 Module O Lead | O Lead Il OK Cancel 5 Effective Date: 09 Sep 2024 | QR700010823 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 Visualization SIEMENS Healthineers The respiration rate is displayed on the top EtCO2 32 30 2 Resp 25 1 HF 1 80 ............. or – – on the right side of the LIVE display. 118 / 78 AO 93 The EtCO2 curve is displayed in the horizontal parameter --- / 2 NBP area. Il 10 200 The EtCO2 may only be displayed in the horizontal parameter area AO 200 B T1 36.8 ºc 180 across the top of the screen. If the EtCO2 value cannot be 160 calculated, an icon is displayed, showing one of the following 140 reasons: 120 25 100 The MicroPod TM */NanoPod * is in self-maintenance status. TM 80 50 1 C The MicroPod /NanoPod TM TM is being purged. 40 20 The MicroPod /NanoPod TM TM is disconnected. The MicroPod /NanoPod TM TM is not functioning properly. 6 * The product names and/or brands referred to are the property of their respective Effective Date: 09 Sep 2024 | QR700010823 trademark holders. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 Visualization configuration SIEMENS Healthineers Configuration of displaying the respiration rate on the EtCO2 32 Resp 25 HF 80 30 ............... LIVE display O f. Sensis Vibe - Configuration Panel X 118 / 78 Eile X Before starting a study, do the following: AO 93 Auto Delete Patient Certificate Transfer Patient Browser Software Handler Registration Distribution 1. From the main menu Options > Configuration, Image Text Regional and EP Catheter Setups Events Ablation Configuration Setup NBP open the Setup dialog and select the Setup to Label HEMO_P1 AO Displayed Waveforms Parameters Analog Outputs Sp02 97 % configure the visualization. ** Displayed Parameters Vertical Parameters On the Parameters tab, select Insert for either Horizontal Parameters 200 2. Insert. ype Timeout Time eart Rate (averaged) Type Insert. 4 s 180 Edit P1:AO (averaged) 4 s Vertical Parameters or Horizontal Parameters. NBP 5 min Edit. 25 SpO2 4 s - Move Up Insert Vital Sign Parameter X Move Left 160 Then, choose Respiration Rate from the list. Move Down Move Right 140 Remove Parameter Cycle Length Remove 3. Confirm the setting. Timeout Respiration Rate 120 Algorithm T2 100 -Audible Parameters O beat-to-beat Repeat the steps for configuring the subsequent 80 QRS tone OK Cancel Volume 60 setups. SpO2 level on QRS (requires SpO2 in Displayed Parameters list) Min Max 40 20 OK Cancel Apply Set as default Rename Setup Copy to New Delete Setup 7 Effective Date: 09 Sep 2024 | QR700010823 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 Visualization configuration SIEMENS Healthineers Configuration of displaying EtCO2 on the LIVE display 50 EtCO2 32 30 Resp 25 HF 80 0 Before starting a study, do the following: #f. Sensis Vibe - Configur X 118 / 78 File X 1. From the main menu Options > Configuration, AO 93 Auto Delete Patient Certificate Transfer Patient Browser Software Handler Registration Distribution open the Setup dialog and select the Setup to Image Text Regional and EP Catheter Setups Events Ablation Configuration Setup NBP configure the visualization. Label HEMO_P1 AO Displayed Waveforms Parameters Analog Outputs Sp02 97 % On the Parameters tab, select Insert for Horizontal ** 2. -Displayed Parameters Vertical Parameters Horizontal Parameters 200 Parameters. Choose EtCO2 from the list. Insert. Type Timeout Type Time Insert. Heart Rate (averaged) 4 s P1:AO (averaged) 180 4 s 3. Confirm the setting. Edit NBP 5 min Edit SpO2 4 s - Move Up Insert Vital Sign Parameter Move Left 160 x Move Down Move Right 140 Remove Parameter Remove Repeat the steps for configuring the subsequent setups. Cycle Length CO (thermo) meout 120 EtCO2 gonthm T1 100 -Audible Parameters T2 Dea 80 QRS tone OK Cancel Volume 60 Spo2 level on QRS (requires SpO2 in Displayed Parameters list) Min Max 40 20 0 OK Cancel Apply Set as default Rename Setup Copy to New Delete Setup 8 Effective Date: 09 Sep 2024 | QR700010823 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 MicroPod TM */NanoPod * measurement TM SIEMENS Healthineers The respiration kit contains the following items: Microstream® MicroPod TM Microstream® NanoPod TM • The MicroPod /NanoPod TM TM which can be External EtCO2 module External EtCO2 module mounted on an accessory/IV pole or bed-rail • Microstream® CO₂ sampling lines to the MicroPod /NanoPod TM TM • Connector cable from the MicroPod /NanoPod TM TM tco, to the MicroPod /NanoPod TM TM connector of the Medtronic Microstream CO2 Sensis Vibe signal input box (HemoBox or HemoBox ComboBox) ComboBox ICEG CEG P1/P2 ECG For more information on the system setup, please refer Sp02 AUX to the Respiration Rate Measurement chapter in the P3/P4 SpO2 Operator Manual. ECG MicroPod" O NIBP CO HEMOMED EtCO2 32 ............................................... STIM NBP Resp 25 A 0 IPX1 T1/T2 EtCO 2 Microstream * x 9 * The product names and/or brands referred to are the property of their respective Effective Date: 09 Sep 2024 | QR700010823 trademark holder. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 SIEMENS Healthineers Please note that this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com 10 Effective Date: 09 Sep 2024 | QR700010823 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024