Sensis Vibe Configuration Panel VD12 Job Aid - USA
This job aid describes the configuration panel for Sensis Vibe VD12 software.
SIEMENS Healthineers Configuration Panel Job Aid Sensis Vibe VD12 Template Effective Date: 10 Aug 2021| HILS 2223 US Effective Date: 08/30/2022 I HOOD05162003299782 1 OUS Job Aid Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Table of Contents SIEMENS Healthineers Configuration Panel 3 Events Configuration 4 Predefined Pullback Sequences 9 Pressure Signal Schemes 11 Study Type Settings 12 Print Configuration 13 12 Lead ECG Snapshot Configuration 14 Image Text Editor 15 Auto Delete Configuration 16 SIEMENS Please note: The product and/or features here mentioned may not be commercially available in all countries. Due to regulatory reasons their future availability cannot be guaranteed. Please contact your local Siemens Healthineers organization for further details. 2 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Configuration Panel SIEMENS Healthineers The Sensis Vibe Configuration Panel provides access to various system, Patient Applications Edit View Acquire Input Tools Options Help acquisition, and display settings for customizing your system. 1 Maintenance. Review Real Time End Session Calculate Configuration Pressures Service 1 From the menu bar, select Options>Configuration. Setup ody nam .. Sensis Vibe - Configuration Panel 2 x 2 Configuration Panel opens. File × L 3 Select the appropriate parameter to modify. Auto Delete Patient Certificate Transfer Patient Software Browser Handler Registration Distribution Image Text Regional and EP Catheter Setups Events Ablation Please note: The study should be closed, and no exam is running, e.g., Editor Language . Connections Configuration Configuration some settings you have only read-only access like setups. Pressure Transfer Parameters Print 12 Lead ECG Stimulation Signals Setups a ... Configuration Snapshot Configuration Mapping Workflow Predefined Study Type 3 System Co ... Support Co. Pullback Settings Sequences Predefined Pullback Sequence Configuration 3 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Events Configuration SIEMENS Healthineers Tachycardia & Bradycardia In the Events Configuration dialog box, you can define whether, when, and Sensis Vibe - Configuration Panel x how events are to be created automatically*. File Auto Delete Patient Certificate Transfer Patient tware Browser Handler Docintroti 1 1 Double-click on Events Configuration. Click on either the Tachycardia or Bradycardia tab card. The Events Configuration dialog box will Image Text Regional and EP Catheter Setups Events Ablation Editor Language ... Connections Configuration Configuration open. This is where you can set the trigger level. Pressure Transf Tachycardia Bradycardia Vital Signs Vital Sig Mark Inherit CO Signals Setups The default tachycardia or bradycardia Heart Rate trigger and 2 2 Triggers for tachycardia Duration of beats should be entered. When tachycardia or Mapping Workflo System Co. Support HR >= 150 Duration (beats) 4 bradycardia is present, the system by default will automatically create an Event in the Case Log based on the configuration setting. Tachycardia Bradycardia Vital Signs Vital Sig Mark Inherit CO 2 Triggers for bradycardia 3 Depending on how you wish to document these events: Configure different tung Do not log as an event to turn tachycardia/bradycardia logging off. HR <= 50 Duration (beats) 4 Log as an event if recording is active to have the system set a marker in an ongoing recording and save it as an event. When a tachycardia appears Log as an event and start recording to make the system Log as an event and start recording automatically start a waveform recording and create an event when Log as an event if recording is active 3 the trigger events set above are observed Do not log as an event 4 *The Clinical Administrator should define them according to the requirements in the Cath Lab. OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Events Configuration SIEMENS Healthineers Vital Signs The Vital Signs tab provides the parameters to be acquired and logged Events Configuration x 1 Tachycardia Bradycardia Vital Signs Vital Signs Al .. Mark Inherit CO during a procedure. Additional assessments can also be prompted by the Parameters Vital Signs Event. ER (averaged) Sp02 Sedation Assessment X NBP Respiration Rate IBP Time 12:58:09 PM EtCO2 Sedation Assessment IBP Channel Consciousness 1 1: Arousal on calling Select the Parameter(s) in which you would like to acquire and log P1 P2 . P3 . P4 Respiration by checking the corresponding checkboxes. When Vital Signs Event is created, prompt for. 2: Deep breath or cough freely 02 Saturation 2 Consciousness Muscle Function Pain Score 0: < 90% despite supplemental 02 Respiration Circulation 2 Select the Assessment(s) to be prompted for from the vital signs 02 Saturation Sedation Assessment event. When you select one or several conditions, the system 2 automatically prompts you to rate the status of the patient according to these criteria every time a vital sign event is created. Pain Score 5 of 10: Discomfort - Moderate Pain OK Cancel 3 You may choose to check the By default, only for neonatal mode box NBP in order to use the system only for neonatal mode. You can also set By default, only use neonatal mode the NBP mode for emergency studies to neonatal mode. Additionally, the minimum age to use adult mode can be set. Min age to use adult mode (months) 24 3 Note that you should not use NBP adult mode on neonates or patients who weigh less than 8.2 kg (18 pounds) When examining emergency patient, use: NBP neonatal mode · NBP adult mode 5 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Events Configuration SIEMENS Healthineers Vital Signs No breath timeout: If the system does not detect any breath for the Events Configuration x 4 Tachycardia Bradycardia Vital Signs Vital Signs Al ... Mark Inherit CO specified period, the system will treat this as no breath and display the Parameters respiration rate as — instead of 0. HR (averaged) Sp02 NBP Respiration Rate IBP EtCO2 IBP Channel P3 5 Select the Respiration Rate method to be used as default: P1 P2 . P4 CO2 Module: The respiration rate is calculated with MicroPodTM When Vital Signs Event is created, prompt for Consciousness Muscle Function Pain Score Respiration Circulation 02 Saturation Sedation Assessment Measure vital signs: Repeatedly Log interval (min) 10 NBP By default, only use neonatal mode Min age to use adult mode (months) 24 4 Respiration Rate Source No breath timeout For neonatal 15 s For adult 30 s Respiration Rate Source No breath timeout For neonatal 15 s For adult 30 s Which license to use for calculating respiration rate CO2 Module OK Which license to use for calculating respiration rate 5 · CO2 Module 6 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Events Configuration SIEMENS Healthineers Vital Signs Alarm On the Vital Signs Alarm tab, alarms can be set per vital sign and patient type. Events Configuration X Tachycardia Bradycardia Vital Signs Vital Signs Al .. Mark 1 CO Vital Sign Alarm Setup 1 You can select the Parameters that need to be configured, then you can Patient Settings Patient Type permanently activate the Audible alarm and set Minimum and Maximum Adult Limits that will activate the alarm when the related vital signs exceed Parameter Audible Min Limit Max Limit M HR (average) 45 120 bpm these limits. You can also arrange the Mute Time of the alarms and M Sp02 90 100 % volume. Resp M 15 30 breaths/min 2 EtCO2 M 10 60 mmHg On the other hand, the vital sign alarm default settings can be modified by NBP Mean M 60 125 mmHg the system administrator. When you click on the Modify button, the Sensis BP Mean 60 125mmHg SIS Administrator - Login dialog box will be displayed, and you are Apply Reset to Factory Presets prompted to enter the SIS Administrator Username and Password in IBP Arterial Channel order to access the tab card. Therefore, if necessary, contact your Clinical 6 P1 O P2 O P3 O P4 Administrator if you wish to permanently change a Vital Sign Alarm Audible Alarm Settings setting. 2 Mute Time 2 min Isis Administrator - Login X Max SIS Administrator User name Modify Password OK Cancel 7 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Events Configuration SIEMENS Healthineers Mark The Mark tab card allows you to assign a function key (F7 to F12) using a Events Configuration x Sensis Vibe Keyboard to create the event. Tachycardia Bradycardia Vital Signs Vital Signs Al Mark Inherit CO 1 Comment Function Key Ablation success Add 1 The Add button enables you to enter a new comment and define a AH Jump Arrhythmia start Edit shortcut key for this comment. Add Comment x Comment Ablation Start Delete 2 Through the Edit button, you can change the comment or the Edit Comment 2 Comment Ablation success shortcut key of the existing one. You can also delete existing 1 comments by Delete button. Function Key F7 OK Cancel Function Key F8 3 When the Display "Select Comment" dialog is not selected, pressing one of the assigned function keys creates a Mark event in CaseLog OK Cancel immediately and inserts the associated predefined comment. 3 Select Comment Display "Select Comment" dialog Commen Function Key Ablation start Ablation success If the Display "Select Comment" dialog is selected, the Select AH Jump Arrhythmia start Arrhythmia termination Cardioversion Echo Comment dialog box opens when any of the assigned function keys are pressed. Then a comment from the comment list can be selected. OK Cancel Ablation start OK Cancel 8 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Predefined Pullback Sequences (1/2) SIEMENS Healthineers Through Pullback Sequences, the system will calculate the pressure Sensis Vibe - Configuration P Predefined Pullback Sequences Add Predefined Pullback Sequence X File difference and gradient between the sites of each step across the entire 2 Specify the settings below to create a new pullback Add .. sequence. distance of the catheter movement. L Sweep Speed : 12.5 mm/s Auto Delete Patient Certific Edit . Browser Handl Available Sites Remove Double-click the Predefined Pullback Sequences icon on the EVEN Femo - Vein Femoral 1 2 BSVC - Superior Vena Cava Regional and EP Cath IVC - Inferior Vena Cava Configuration Panel. The Predefined Pullback Sequences dialog box Image Text Editor Language ... Connect BRA - Right Atrium OK Cancel RV - Right Ventricle will be displayed. PA - Pulmonary Artery Main B-PCW - Pulmonary Capillary Wedge PCW Left - Pulmonary Capillary Wedge Left Pressure Transfer Parameters Print 12 Lead ECG Stimulation PCW Rght - Pulmonary Capillary Wedge Right Signals Setups a ... Configuration Snapshot Configuration B-PV - Pulmonary Vein 2 To create a new Pullback Sequence, click Add. Then all the pullback -LA - Left Atrium BLV - Left Ventricle B- AO - Aorta sequences that are available on your system will be listed. You can Mapping Workflow Predefined Study Type Calculate overall gradient select desired sites from the list displayed in the Selected Sites box. 3 System Co. Support Co. Pullback Settings Sequences Selected Sites 5 Then you can define your default Sweep Speed by selecting the PCW Left 1 Move Up AO 200 Sweep Speed value from the drop-down menu. Site ... Set Pullback Move Down Predefined Pullback Sequence Configuration Set Pullback Sequence AO-LV Range Up PCW Left - PA - RV - RA - IVC 2 If Calculate overall gradient is selected in the Pullback Sequence Range Down 3 Range Hide dialog box, a pullback sequence will combine several pullback 4 Hide Range Hide Grid measurements and will calculate pressure differences and gradients Show Mean Pressure Properties for each step across the entire distance of the catheter movement. Set as Analog Output 4 To use a created Predefined Pullback Sequence during an ongoing procedure, right-click the waveform, click Set Pullback Sequence, then select a predefined sequence. 9 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Predefined Pullback Sequences (2/2) SIEMENS Healthineers 5 To modify an existing pull back sequence, select the desired Sensis Vibe - Configuration Panel Predefined Pullback Sequences X Sequence from the list and click Edit. The Edit Predefined Pullback File × Sequence pop-up window opens. You can change the default Sweep PCW Left - PA - RV - RA - IVC L Auto Delete Patient Add ... Speed for this sequence by selecting a sweep speed from the drop- Certificate Transfer Browser Handler down list. Edit Predefined Pullback Sequence 5 x Edit ... By single clicking on the selected site from Available Sites list, you Image Text Regional and EP Catheter Edit the settings below to update the predefined pullback Editor Language ... Connections sequence can delete the selected site form sequence. Remove 6 To reorder the selected site, use the Move Up and Move Down Sweep Speed : 12,5 mm/s 5 mm/s buttons. Pressure Transfer Parameters Available Sites 12,5 mm/s Signals Setups a ... Configuration - Normal 25 mm/s VEN - Vein 50 mm/s VEN Inno - Vein Innominate To delete an existing pullback sequence, select the desired sequence + VEN Femo - Vein Femoral 6 Mapping Workflow Predefined Stu SVC - Superior Vena Cava ..... from the list and click the Remove button. System Co ... Support Co ... Pullback S + IVC - Inferior Vena Cava Sequences + RA - Right Atrium 5 RV - Right Ventricle PA - Pulmonary Artery Main ...... B PCW - Pulmonary Capillary Wedge PCW Left - Pulmonary Capillary Wedge Left Predefined Pullback Sequence Configuration PCW Rght - Pulmonary Capillary Wedge Right B PV - Pulmonary Vein Calculate overall gradient Selected Sites PCW Left PA Move Up RV RA IVC Move Down OK Cancel 10 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Pressure Signal Schemes SIEMENS Healthineers A Pressure Signal Scheme defines signal acquisition parameters for up to HR Sensis Vibe - Configuration Panel Configure Pressure Signal Scheme four pressure signals. You can use the Pressure Signal Scheme when you File × Scheme New ... AO define setups. AOLV AO LV RV Auto Delete Patient Certificate OK AO LV RV PCW Browser Handler Four pressures Hemo 1 Double-click the Pressure Signals icon on the Configuration Panel. 1 Select the scheme that you want to modify from the drop-down list Image Text Regional and EP Catheter New Pressure Signal Scheme X OR select New and label it. Editor anguage Connections 1 Label Exp_New Pressure Signal Assign an anatomical site to the pressure by selecting the Enter Site Pressure Transfer Parameters 2 Signals Setups a .. Configuration OK Cance button and choosing it from the list. To remove a pressure signal from the study, select it from the list and click Delete Signal. Mapping Workflow Predefined System Co .. Support Co ... Pullback ... Site x Normal VEN - Vein 3 Within Lesion Index Assignments, you can assign Distal (Pressure Pressure Signal Scheme VEN Inno - Vein Innominate Exp_New Pressure Signal @ VEN Femo - Vein Femoral Wire) and Proximal (Reference Pressure) sites to the FFR 2 2 Scheme SVC - Superior Vena Cava IVC - Inferior Vena Cava Signal P1 Site Frequency Ran. Notch Refer. Wire RA - Right Atrium measurement by selecting the corresponding pressure signal scheme Not 0 - 15 Hz X Signal P1 RV - Right Ventricle P2 Not 0 - 15 Hz Site Not Defined Enter Site + PA - Pulmonary Artery Main P3 in your setup. P4 Not .. 0 - 15 Hz + PCW - Pulmonary Capillary Wedge PV - Pulmonary Vein - LA - Left Atrium Lesion Index Assignments ELV - Left Ventricle 3 @ Proximal ( Reference Pressure) AO - Aorta .... AO Asc - Aorta Ascending Distal (Pressure Wire) AO Arch - Aorta Arch AO Desc - Aorta Descending Delete Signal ART - Artery Affect All ART Brac - Artery Brachial ART Femo - Artery Femoral 2 Frequency Range 0 - 15 Hz Congenital (Hz) Coronary Notch Peripheral Not Defined OK Cancel Rename Scheme Delete Scheme OK Cancel 11 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Study Type Settings SIEMENS Healthineers I Sensis Vibe - Configuration Panel x A study type is a typical examination that is frequently performed in your Cath lab to File clarify a specific diagnostic problem. Such a typical examination usually implies that a X set of signals are acquired, and a set of measurements and calculations are performed. Auto Delete Patient Certificate Transfer Patient Software Browser Handler Registration Distribution The study type controls the preselection of a setup or workflow support program, as well as data input and reporting options for the procedure.* Image Text Regional and EP Catheter Setups Events Ablation Editor Language ... Connections Configuration Configuration 1 Pressure Transfer Parameters 12 Lead ECG Stimulation Signals Setups a ... Configuration Snapshot Configuration 1 Double-click the Study Type Settings icon on the Configuration Panel. The Study Type Settings dialog box opens. Mapping Workflow Predefined Study Type System Co Support Co ... Pullback ... Settings 2 The configured study types are displayed, as well as defined setup or workflow Study Type Settings X support programs (depending on selection). Select the study you wish to Select Study Type 2 Associate Study Type to associate. 3 Study Typ Coronary Default Setup Diagnostic Coronary Catheterization Setup Electrophysiology Workflow Support Program 3 If you select Default Setup, the setup designated Set As Default will be associated LElectrophysiology study with that Study Type. Pediatric 4 LH Pediatric Cardiac Catheterization Lesion Index Pediatric Cardiac Intervention LH Peripheral RH 4 Making an association using Setup or Workflow Support Program, the drop- RH/LH Ped Right-Left down list will display the defined setups or programs. Select the one to be associated with the study type and click the Apply button to apply it to the Apply selected study. Close 12 *The Clinical Administrator can modify the Study Types according to the requirements in your Lab. OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 Print Configuration SIEMENS Healthineers Grid lines are used to more easily read and follow signals in the printout. In Sensis Vibe - Configuration Panel X File the Print Configuration, you can define whether grid lines are to be printed Print Configuration X by clicking On/Off in 12 Lead ECG or Other Waveform Printouts. Auto Delete Patient Browser tifica Transfer Patient Software Registration Distribution Raster 12 Lead ECG Printouts 1 When you double-click the Print icon on the Configuration Panel, the Image Text Regional and Co 1 Setups Events Ablation Editor Language ... Configuration Configuration o On Print Configuration dialog box opens. c Off 1 ~ 125 Other Waveform Printouts Pressure Signals Transfer Setups Parameters Print 2 Lead ECG Stimulation and Programs Configuration Snapshot Configuration c On 2 By selecting On in 12 Lead ECG Printouts or Other Waveform Off Printouts, you can define grid lines to be printed. Mapping Workflow Support C. Predefined Study Type System C. Settings OK Cancel 3 By selecting Off in 12 Lead ECG Printouts or Other Waveform Printouts, you can define grid lines not to be printed. Print Configuration NUM SIEMENS 1. Rest: 12 Lead ECG 6:27:56 PM-4 s SIEMENS 1 . Rest: 12 Lead ECG 6 : 14:29 PM-4 S Sensis Vibe VD124 avL 10 mm/mv 2 aVL 10 mm/mv 3 10 mmimV 1 10 mmimv -aVR 10 mm/mV -aVR 10 mmimV Il 10 mm/mV 11 10 mm/mv aVF 10 mm/mV aVF 10 mmimv Ill 10 mm/mV Il1 10 mm/mv V1 10 mm/mV IV1 10 mm/mV V2 10 mm/mV V2 10 manimV V3 10 mm/mV V3 10 mmimV V4 10 mm/mv IV4 10 mmimv V5 10 mm/mV V5 10 manimV V6 10 mm/mV vs 10 mmimv 25 mm/s Summary: 80HR Summary: 80HR 13 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 12 Lead ECG Snapshot Configuration SIEMENS Healthineers In the 12 Lead ECG Snapshot Configuration, you can define the system af Sensis Vibe - Configuration Panel X File behavior every time a 12 Lead ECG snapshot is made. Auto Delete Patient Browser Certificate Transfer Patient Software Handle Registration Distribution 1 When you click on Print and Display, the Review waveform card will be displayed with the 12 Lead ECG and a printout will be started Image Text Regional and EP Catheter Setups Events Editor Language ... Connections Configuration Ablation automatically every time a 12 Lead ECG snapshot is made. Configuration ~ 122 2 If Print is selected, a printout is made automatically when 12 Lead Pressure Transfer Setups Parameters Print 12 Lead ECG Stimulation Signals and Programs Configuration Snapshot Configuration ECG Snapshot is clicked during an examination. Mapping Workflow Predefined Study Type System C ... Support C ... Pullback ... Settings When you select Display, the review waveform card will be moved to 3 the foreground showing the 12 Lead ECG every time 12 Lead ECG Snapshot is clicked during an examination. 12 Lead ECG Snapshot Configuration x (No printout is made if this option is selected). 12 Lead ECG Snapshot C 1 @ Print and Display 2 c Print 3 Display Sweep Speed 25 OK Cancel 14 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 15 OUS Effective Date: 03 Aug 2022| HOOD05162003291996 Unrestricted © Siemens Healthineers, 2022