Replacing the LOCI® Insert on the Dimension® EXL™ with LM and Dimension® EXL 200 Integrated Chemistry Systems Video
This video provides instructions on how to replace the LOCI® insert on the Dimension® EXL™ with LM and Dimension EXL™ 200 Integrated Chemistry Systems.
1st power down the system by following the performing a controlled power shutdown procedure in the dimension system operators guide. Open the loci access door located on the rear side of the analyzer. Lift up the Locci arm and rotate it out of the way. Remove the rubber seal. Carefully insert a pipette tip or slender object into the Reed chamber. Press the tip against the inside of the insert and draw the insert out. Hold the new insert by the top edge and carefully drop it into the low side chamber. Next, carefully reinsert the rubber seal and press firmly down around the entire surface to ensure it is inserted fully. Return the low side arm to its original position and then close the low side access door. After the door is closed, restore power to the system by following the restore power after a controlled shutdown procedure in the dimension system operators guide. With the system in standby, from the operating menu screen, select F7 diagnostics. Select F1 electromechanical. Select F2 sample system. Select F7 LOCI module. Then select F7 statistics. Select F1 reset to reset the low size statistics. When the message do really want to clear, the low size statistics file appears on the screen. Select yes to confirm. In the cycle field, type 40 as the number of cycles to perform. Select F8 cycle. After cycling is complete, run QC on all loci methods. Then exit to the main operating menu screen. If documenting using the paper logs from the dimension system operators guide, document the maintenance and the weekly slash monthly maintenance log sheet or instrument log sheet. Or, if documenting using the electronic logs in the dimension system software, document the maintenance in the instrument log screen.
14 000 212 10 2467 100 0, 0.000 (1) Replacing the LOCI® Insert on the Integrated Chemistry Systems SIEMENS . Healthineers Standby Sampler Idle IMT OK 08/16/22 23:20 Dimension Exit Enter en Change Operator STAT Status Sample Alert Do you really want to Shutdown/ Change Operator? Supplies rogram. QC Alert Calib Alert GuestUser Run Shutdown Home Cancel F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM PREP F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS 24 3D 3.15A SMP LD OVRD OFF - BRANCH SLOT VOLTAGE 09/14/22 21:23 labmanager OPERATING MENU PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: At any time, press HELP to get help, or press EXIT to exit a program. To Shut down or Change Operator, press Sign Out or Alt+EXIT. 01/24/23 15:58 DIAGNOSTIC MENU Press EXIT to exit a program. F1: ELECTRO/MECH F2: FILM LOAD F3: ALIGNMENTS F4: F5: CHECK WINDOWS F6: ERROR LOG F7: KINETICS F8: OPEN CHANNELS Diagnostics ELECTRO/MECHANICAL DIAGNOSTICS SELECT PROGRAM USING FUNCTION KEYS F1: REAGENT F2: SAMPLE SYSTEM F3: IMT F4: CUVETTE F5: PHOTOMETER F6: WATER WORKS F7: POWER SYSTEM F8: ULTRASONICS SAMPLE SYSTEM DIAGNOSTICS MENU F1: Level Sense F2: Sampler F3: Samp Wheel F4: Clot Check F5: HM Module F6: F7: LOCI Module F8: LOCI MODULE DIAGNOSTIC: CYCLE: 40 ( 1) Position Encoder Bump Pressure: 0.0000 ROTATION VERTICAL DETECTOR READ Temperature 0.00 PM 0.00 chamber Preread 0 PM 0.0 contamination 0.0 illumination Read init Read PM 0} 0 total Read VF Gain 0 VF 0.000 relative Shutter 0.000 open 0.000 close 0.000 stability F1: INITIALIZE F2: PLUS F3: MINUS F5: HOME F6: INSPECT F7: Statistics F8: CYCLE LOCI DETECTOR STATISTICS DIAGNOSTIC: Collection Rate: 240 minutes Value Mean STD Contamination Illumination Gain Read PM Gain Read VF Gain Ratio Open Shutter Ratio Close Shutter Ratio Stability Ratio Press F1: RESET to clear all Loci statistics Press F8: CYCLE to collect Loci statistics 'n' times F1: RESET F2: F3: F5: F7: Do you really want to clear the Loci statistics file? (y/n) YES NO Working ... Press any key to stop (may take up to a minute). MAINTENANCE LOGS Year: 20XX Operator Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. C Siemens Healthcare GmbH 2023 SIEMENS
- dimension
- Dimension
- LOCI
- loci
- insert
- dimension maintenance
- Dimension Maintenance
- Dimension Maintenance LOCI