Replacing the IMT Pump Tubing on the Dimension® Family of Integrated Chemistry Systems Video

The IMT peristaltic pump tubing on the Dimension® Family of Integrated Chemistry Systems are are customer replaceable parts. This video provides instruction on how to replace the IMT peristaltic pump tubing, which is part of monthly maintenance tasks. 

The X tubing on the. Imt peristaltic pump is replaced monthly to compensate for. Wearing and stretching. With the instrument in standby. Press pause to stop the sampling system. Raise the IMT lead. From the operating? Menu. Select F4 system Rep. Select F. 3 irmt. Select F3, align prime. To locate the IMT. Peristaltic pump. X tubing locate the IMT tubing. Diagram label on the instrument. Located on the inner side of the far left cabinet door. The IMT tubing. Diagram is also. Located in the appendix section of the dimension EXL. 200 and dimension exl. With. LM. System operators guide. Turn the pressure plates foot. Bar 1/4 turn and lift to remove. After opening the pressure plates. Remove. And replace tubing acts. Located around the top of the pump. Pull to fit the ends of the tubing into place. Discard. Use tubing into a. Biohazard. Waste. Container. Close the pressure plates. Reinsert the pressure plates foot. Bar and turn 1/4 turn to lock in the pressure. Plates. Select pause to restart the sampling system. Select F. One Rime salt bridge solution. Prime The Salt bridge solution until the solution. Exits the IMT sensor and is visible in the. X2 tubing. Select. Home. On the touch screen to return to the operating menu. You also can access. The operating menu screen by holding down the shift. Key and pressing exit. The. System. Automatically performs an IMT pump alignment and IMT. Calibration. If documenting using. The paper logs from the dimension system operators guide. Document the maintenance and the weekly slash, monthly maintenance log sheet or instrument log. Sheet. Or if documenting, using the electronic logs in the dimension. System. Software. Document the maintenance in the instrument log. Screen.

Replacing the IMT Pump Tubing on the Dimension Family of Integrated Chemistry Systems Video Repl Fam SIEMENS X Tubing SIEMENS . Standby Sampler Idle IMT OK 09/14/22 21:23 Dimension Exit labmanager en Enter OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert Supplies At any time, press HELP to get help, or press EXIT to exit a program. QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM PREP F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS Samplers Off 02/27/23 03:00 Dimension EXL 20 Dim Healthineers F4: SYSTEM CHECK SYSTEM PREPARATION MENU F1: INVENTORY F2: REAGENT PREP F3: IMT F5: TEMP MONITOR F6: SYS COUNTERS F7: PUMP PRIME F8: DAILY MAINT. IMT SETUP MENU F1: CHANGE CONSUM F2: CALIBRATION F3: ALIGN/PRIME F4: COND/DILCHK F5: IMT OVERVIEW F6: REVIEW DILCHK F7: IMT CONFIG F8: IMT SYS CLEAN IMT Paused FLUIDS PRIME / PUMP ALIGNMENT Press a function key to select an option. Automatic pump alignment mode: ON IMT pumping rate (uL/revolution): 73.4 F1: PRIME SALT F2: F4: F5: PRIME STD A F6: PRIME STD B F7: PRIME FLUSH F8: PRIME DILUENT Fluidics for QuikLYTE® Module Standard A X1 PART OF PUMP PANEL F2 PN Warning: IMT pump tubing X is a biohazard. Follow laboratory specific safe biohazard waste disposal procedures when discarding this tubing. no 02/27/23 03:04 F3: ALIGN PUMP Priming Salt Bridge Solution. Processing IMT Not Calib IMT Calibration has been scheduled. MAINTENANCE LOGS Year: 20XX Clean R3 Operator Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. C Siemens Healthcare GmbH 2023 Health

  • 11075682
  • dimension
  • exl
  • exl with lm
  • rxl max
  • xpand
  • imt
  • IMT
  • peristaltic pump
  • tubing
  • IMT tubing
  • replace IMT tubing
  • replace imt tubing on dimension