Replacing the HM Pump Heads on the Wash Station for the Dimension® Family of Integrated Chemistry Systems Video
The HM Pump Heads on the Wash Station for the Dimension® Family of Integrated Chemistry Systems are customer replaceable parts. This video will demonstrate how to replace the HM Pump Heads.
To prevent any automatic priming of the HM Wash probes, navigate to the pump prime screen. With the system in standby, access the HM Pump prime screen from the operating menu by selecting F4 system RE. Select F7 pump prime. Select F6 HM. Raise the IMT and reagent lids to access the pump head. The aspirate and dispense pump heads are located on the wash station frame. Disconnect the tubing from the pump head at its connectors. Using your fingers, squeeze the two tabs on the top and bottom of the pump head and pull it off the pump base, then discard it into a biohazard waste container. To install the new pump head on the motor, insert the shipping plug or T plug that came with the pump head completely into the front of the cassette. Slide the pump head on to the shaft of the motor, allowing the shipping plug to fall as the new pump head is installed. Snap the pump head into place. Reattach the tubing to the pump head. Repeat these steps to replace pump head wash pump #1. To prime the HMM wash pump, set the cycles field to three and press enter. Then select F1 HMM wash pump. Close all instrument lids, then perform a system check. Refer to the operators guide for specifics on this procedure. If documenting using the paper logs from the dimension system operators guide, document the maintenance and the weekly slash monthly maintenance log sheet or instrument log sheet. Or if documenting using the electronic logs in the dimension system software, document the maintenance and the instrument log screen.
Replacing the HM Pump Heads on the Wash Station for the Dimension® Family of Integrated Chemistry Systems Video SIEMENS Standby Sampler Idle IMT OK 09/14/22 21:23 Dimension Exit Enter labmanager en OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert Supplies At any time, press HELP to get help, or press EXIT to exit a program. QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM CHECK F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS 01/12/23 07:43 SYSTEM PREPARATION MENU F1: INVENTORY F2: REAGENT PREP F3: IMT F5: TEMP MONITOR F6: SYS COUNTERS F7: PUMP PRIME F8: DAILY MAINT. Diagnostics PUMP PRIME CYCLES: 10 IMT SAMPLER REAGENT 1 REAGENT 2 DIL PUMP REAGENT 3 F1: PRIME WATER F2: SAMPLER F3: REAGENT 1 F4: REAGENT 2 F5: IMT DIL PUMP F6: HM ... F7: REAGENT 3 Dimension EXL 20 Dimension E Dimensi Dimension EXL 200 WASH PROU Pump Head Wash Pump 2 ASHY Warning: the pump head is a biohazard. Use laboratory safe biohazard waste disposal procedures when discarding. HOME SENSOR WASH PROB CLAMP PC 01/12/23 07:58 HM PUMP PRIME CYCLES: HM REAGENT WASH CLEANER F1: HM WASH PUMP F2: REAG CLEANER F3: SAMP CLEANER F4: F5: F6: F7: F8: Priming pumps ... Dimen Dim MAINTENANCE LOGS Year: 20XX Clean R3 Operator Drain ** Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. C Siemens Healthcare GmbH 2023 Healthineers
- 10472702
- Dimension
- dimension
- HM
- hm
- HM wash probe
- hm wash probe
- maintenance
- Maintenance
- Dimension Family