Replacing a Reagent Probe Tip on the Dimension® Family of Integrated Chemistry Systems Video
The probe tips on the Dimension® Systems are customer replaceable parts. This video provides instructions on how to replace a probe tip, using the R2 probe tip replacement as an example.
As the reagent probe tips for the R1R2 and R3 probes are the same part, this replacement process will be demonstrated on the R2 probe tip. With the instrument in standby, select pause to turn the samplers off, raise the IMT and reagent lids. Turn the bottom gold bronze colored spline shaft on the reagent arm to raise the reagent probe up and out of the reagent drain. Move the R2 arm clockwise until the probe tip is easily accessible. Place a paper towel under the probe to prevent anything from falling into a reagent cartridge while removing the probe tip. Use A5 sixteenths open end wrench on the flat part of the probe body to stabilize it. Slide the 1/4 end of the offset wrench from the bottom up the reagent probe and onto the probe tip nut while stabilizing the probe body. Turn the offset wrench to the left to loosen the probe. Once the nut is loose, unscrew it by hand to remove both the probe tip and the nut. Next, insert a new reagent probe tip into a new probe tip nut. Screw the new reagent probe tip onto the reagent probe body finger tight using the five sixteenths open end wrench on the reagent probe body to stabilize and the 1/4 offset wrench to turn. Completely tighten the new reagent probe tip. close the instrument lids. And select Pause to restart the sampling system. From the operating menu, select F4 System Prep, F6 System counters. Move the cursor box to the R2 Probe tip field and select Enter to change the field from No to Yes. Select F1 to store changes. So do you want to store the new system? Information message box will appear to update the system counter. Select Yes and the screen will update to 0 cycles with the current date. From the operating menu, display the Pump Prime screen by selecting F4 System Prep F7 Pump Prime. Move the cursor to the Cycles field and set the cycle count to three. Select Enter. Then press F4 to prime the R2 pump three times. Follow the operator's guide instructions to perform the probe alignments. Then perform daily QC before running patient samples. If documenting using the paper logs from the Dimension System Operators Guide, document the maintenance in the weekly slash, monthly maintenance log sheet or instrument log sheet. Or, if documenting using the electronic logs in the dimension system software, document the maintenance in the instrument log screen.
1742 13 20000 22 12000 268 137 48000 5214 14551 36000 8:11 Replacing a Reagent Probe Tip on the Dimension® Family of Integrated Chemistry Systems PASSED TESTING SIEMENS . Standby Sampler Idle IMT OK 09/14/22 21:23 Dimension Exit labmanager Enter en OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert At any time, press HELP to get help, or press EXIT to exit a program. Supplies QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM PREP F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS FERR CTN FPSA F1 F2 F4 F7 Alarm Control Pause PBNP TRNF F5 PHOS Caps Lock Samplers Off 07/08/23 07:15 Healthineers Dimension EXL 20 Dimensio Dimer Dimension EXL 200 SIEMENS Tray Photometric Sampler FRONT FRON Photometri Sample Dimen Dim F6 F8 - Backspace Tab F4: SYSTEM CHECK 07/10/23 01:24 SYSTEM PREPARATION MENU F1: INVENTORY F2: REAGENT PREP F3: IMT F5: TEMP MONITOR F6: SYS COUNTERS F7: PUMP PRIME F8: DAILY MAINT. 07/08/23 07:37 SYSTEM COUNTERS Dirty windows: Item Cycles Clean at Last Cleaned Sample Probe: 23-MAR-2023 20:46 Pos: Cycles Replace at Last Replaced Replace Cuvette Cartridge: 12-APR-2023 4:32 Diaphragm: 5-AUG-2020 11:29 NO Sample Probe Tip: 137 48000 R2 Probe Tip: F2: DIAPH CHANGE F3: FILM LOAD F4: CLEAN PROBE F6: HM COUNTERS F7: RMS COUNTERS Installed: 5-AUG-2020 Bootups: 15 Cuvettes: 298 Tests: F1: STORE CHANGES F5: F8: YES Installed: 5-AUG-2020 Bootups: 15 Cuvettes: ce Do you want to store the NEW system information? (y/n) R1 Probe Tip: 8-JUL-2023 8:11 Parameters have been stored. 07/13/23 01:38 Diagnostics PUMP PRIME CYCLES: 10 IMT SAMPLER REAGENT 1 REAGENT 2 DIL PUMP REAGENT 3 F1: PRIME WATER F2: SAMPLER F3: REAGENT 1 F4: REAGENT 2 F5: IMT DIL PUMP F6: HM ... F7: REAGENT 3 CYCLES: Select F4: Reagent 2 to prime the R2 pump a total of 3 times Reagent pump cycling. Press (Enter) to cancel. ING Note: The following probe alignments are for the R2 Reagent Arm and Probe only. Follow the Dimension® System Operator's Guide for alignments for other Reagent Arms and Probes. R2 Reagent Arm and Probe to Drain Reagent R2 Reagent Arm and Probe to Target (Cuvette) to Reagent Tray 24V/30 to HM Incubate Wheel MAINTENANCE LOGS Year: 20Xx Clean R3 Operator Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthcare GmbH 2023
- 10455793
- dimension
- Dimension
- Maintenance
- EXL
- exl
- lm
- LM
- reagent probe tip
- Reagent probe
- reagent
- probe