R3 Reagent Probe Alignment on the Dimension® Family of Integrated Chemistry Systems

Learn how to align the R3 reagent probe on the  Dimension® EXL™ with LM and
Dimension® RxL Max® Integrated Chemistry Systems.

Welcome to the R3 Reagent Management System or RMS Reagent Probe Alignments on the Dimension System video. In this video you will learn how to align the R3 reagent probe on the dimension family of analyzers, including dimension EXL with LM and RXL Max. Please note the safety warnings when aligning dimension system reagent probes. Ensure that the instrument Lids are fully open while performing the alignments or other maintenance on the instrument. During diagnostics, be aware that the modules are moving and of moving parts. While performing diagnostics, align the R3 probe with two alignments in this order, R3 Reagent arm to drain, then R3 Reagent arm to cartridge. These alignments must be performed in this order, the same order as they appear on the RMS Reagent arm alignment screen. Whenever a procedure calls for a complete RMS Reagent arm alignment, use message prompts as they appear to help with each alignment. With the instrument in standby, raise the RMS LED. Select F7 Diagnostics, F3 Alignments, F1 Reagent, F8 RMS, then F4 Reagent to display the RMS Reagent Arm Alignment screen. To complete the R3 Reagent Arm to Drain alignment, complete the following steps in order. From the RMS Reagent arm alignment screen, select F1 Align 2 Drain. Visually check that the reagent probe tip is centered over the rear hole in the reagent drain. Use the down arrow key to move the reagent probe tip down to the drain hole. This helps determine when the tip is centered in the drain. The movement will not be saved as part of the alignment. If the alignment is correct, select exit and continue with the Reagent arm to cartridge alignment procedure. If an adjustment is needed, use F3 and F4 to move the probe tip as appropriate. When this alignment is complete, press F8 Accept. After visually performing an alignment, select F7, check Align and ensure that the component moves and returns to the new alignment position. To complete the R3 Reagent Arm to cartridge alignment, complete the following steps in order. From the RMS Reagent Arm alignment screen, select F2 Align 2 Cart. Place the Reagent Alignment gauge into the RMS Load tray and select Enter. If prompted to remove a reagent cartridge, this same cartridge must be placed into the instrument at the end of this procedure. If the probe is aligned correctly, the probe tip should be centered within the target circle on the gauge and just above the gauge. If the alignment is correct, select exit and continue with the next step. If an adjustment is needed, use F1 through F4 to move and raise or lower the probe tip as appropriate. Caution Perform the up down alignment first to avoid possible damage to the probe. When this alignment is complete, select F8 Accept. After visually performing an alignment, select F7. Check Align and ensure that the component moves and returns to the new alignment position. When the prompt displays that the Reagent Alignment gauge will be transferred back to the RMS load tray, select Enter. Remove the Reagent Alignment gauge from the RMS load tray. Return any removed Reagent cartridges to the Reagent tray. Close the RMS lid.

750317.001 001 10100402 753736.90 R3 Reagent Probe Alignment on the Dimension® Family of Integrated Chemistry Systems Video Ensure instrument lids are fully open to avoid injury while performing maintenance. During diagnostics, modules are moving. Be aware of moving parts while performing diagnostics. AUTION RMS (R3) Probe 2-PHASE SIEMENS Diagnostics Sampler Idle IMT OK 02/12/22 02:39 Healthineers Dimension Exit Enter labmanager en RMS REAGENT ARM ALIGNMENT TO DRAIN STAT Status RMS Reagent arm alignments should be done in the following order: Sample Alert Supplies F1: Align Reagent Arm to Drain F2: Align Reagent Arm to Cartridge QC Alert Calib Alert Run Home F1: ALIGN TO DRAIN F2: ALIGN TO CART. F3: F4: F5: F6: F7: F8: SEE ALIGN FILE FORWARD/BACKWARD Use the up/down arrow keys to move the probe down to the drain. Then use F3/F4 to center the probe in the drain. F1 F3: FORWARD F4: BACKWARD F7: CHECK ALIGN F8: ACCEPT Processing 01/31/22 10:30 OPERATING MENU PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: At any time, press HELP to get help, or press EXIT to exit a program. To Shut down or Change Operator, press Sign Out or Alt+EXIT. F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM PREP F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS IMT Calibrating DIAGNOSTIC MENU Press EXIT to exit a program. F1: ELECTRO/MECH F2: FILM LOAD F3: ALIGNMENTS F5: CHECK WINDOWS F6: ERROR LOG F7: KINETICS F8: OPEN CHANNELS ALIGNMENT MENU SELECT PROGRAM USING FUNCTION KEYS F1: REAGENT F2: SAMPLER F3 F4: PHOTOMETER F5: CUVETTE F6: HM MODULE F8: ALIGN FILE REAGENT ARM ALIGNMENTS Reagent alignments should be done in the following order: align R2 reagent arm to drain F2 align reagent arms and cartridge insertion to reagent tray align R2 reagent arm to target (cuvette) align R2 reagent arm to incubate wheel F7 align RMS module F1:ALIGN R2 DRAIN F2:ALIGN CARTRIDGE F3:ALIGN R2 TARGET F4:ALIGN R2 VESSEL F5:SEE ALIGN FILE F6:ALIGN R1 DRAIN F7:ALIGN R1 TARGET F8:RMS RMS SYSTEM ALIGNMENT MENU TURNTABLE ALIGNMENT ELEVATOR ALIGNMENT TURNTABLE SHUTTLE ALIGNMENT REAGENT ALIGNMENT F1: TURNTABLE F2: ELEVATOR F3: SHUTTLE F4: REAGENT RMS REAGENT ARM ALIGNMENTS 01/31/22 10:33 02/13/22 07:01 IMT Not Calib 01/05/24 00:31 Insufficient IMT Consumables RMS REAGENT ARM ALIGNMENT TO CARTRIDGE UP / DOWN Use F1 through F4 to position the reagent arm. Place alignment gauge in reagent load tray then press enter. F1: FORWARD F2: BACKWARD F3: UP F4: DOWN RMS Reagent Tray P20C er eader on Station DC 4.5V CAUTION Barco Hydration Station 2A SIEMENS . 02/13/22 07:18 01/04/24 23:38 The gauge will be transferred back to load tray. Press (Enter) to continue. Dimension EXL 200 Di Dimen Dim Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, afe reserved. Unrestricted | Published by Siemens Healthineers AG | C Siemens Healthineers AG, 2024 Siemens Healthineers HQ | Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com

  • dimension
  • probe
  • probe alignment
  • r3
  • rms
  • RMS
  • reagent management system