Performing Direct Laser Check

A direct laser check is performed to verify the integrity of the Siemens Healthineers external lasers attached to the CT gantry. This job aid outlines the steps to perform the Direct Laser check.

Performing Direct Laser Check SOMATOM go.Sim/go.Open Pro, VA40 A direct laser check is performed to verify the integrity of the Siemens Healthineers external lasers attached to the CT gantry. Direct Laser Check Quality Assurance needs to be performed at least monthly. It is common for sites to check them more often such as daily or weekly. 1. Place an indexing bar in a similar position on your tabletop, as different overlays have dif- ferent lettering. As some manufacturers have slightly different locking bars, it is recom- mended to use the same locking bar each time. It is helpful to place a mark on the index bar to ensure that placement is the same each time when running. Place the Direct Laser phantom on the index bar as shown. H4 H3 H1 H4 H3 H2 2. Refer to the table below for index bar placement with various vendor RTP overlays. If your vendor is not listed, refer to the picture above for placement. Index bar locating position Overlay type Index position (index bar position for positioning the phantom) Qfix overlay F1 Siemens overlay: F2 Elekta overlay: Position 3 Civco Universal Couchtop H3 (E2) Unrestricted | Published by Siemens Healthineers AG © Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 01/18/2024 | QR700004532 Healthineers 3. Turn the direct lasers on from the tablet, and align the lasers as shown, ensuring that the lasers are in the center of the notches on the front, sides, top, and back of the Direct Laser phantom. It is helpful to dim the room lights to better visualize laser placement. Emergency Patient Register Patient = - 750.0 +0+ HZ LH 4. From configuration, navigate to Direct Laser Check. ? Configuration Panel Administration Portal Administration Portal [EN] Clinical Configuration Checkup Daily QA Calibration Resume syngo Remote Assist Screen recorder SaveLog Remote Service Exam Designer Create Screenshot Fast Contact File Browser Load Service Pack Direct Laser Check myNeedle Laser Check What is New Unrestricted | Published by Siemens Healthineers AG, 2024 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 01/18/2024 | QR700004532 Healthineers Page 2 of 7 5. Follow the prompts on the screen, and select GO. SIEMENS .. Healthineers ... Direct Laser Check X System Check Direct Laser Check 1. Enter the scan room. Close 2. Place the RTP phantom on the table using a Lok-Bar. 3. Manually adjust the longitudinal table position until the Direct Laser lights are aligned with the center groove on the top and sides of the RTP phantom. 4. Manually adjust the table height until the Direct Laser lights are aligned with the grooves on both sides of the phantom. 5. Exit the scan room. 6. Press GO to start the Direct Laser Check. 7. The table will automatically move to the scan plane. 8. X-rays will be produced during the procedure. 9. The table will automatically move back to the Patient Marking Laser plane at the end of the procedure. Repeat GO Cancel 6. Then press start on the control box when indicated. Unrestricted | Published by Siemens Healthineers AG, 2024 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 01/18/2024 | QR700004532 Healthineers Page 3 of 7 7. The program will run for about two minutes and then indicate the results with the follow- ing table below. There is a +/- 1mm tolerance to get a green check result. If there are any red x results, realign the phantom and repeat. If the results still display any red x results, contact your CT Service Provider or Siemens Healthineers Customer Care Center. If fin- ished, select Next, followed by Close. Healthineers . system Check SIEMENS Healthineers Direct Laser Check Direct Laser Check results Direct Laser Check Close Close IEC Coordinate System Value Tolerance Onit Status 0.8 +/- 1.0 mm -0.7 +/- 1.0 -0.2 +/- 1.0 Pitch (X) 0.0 +/- 0.3 deg Roll () 0.0 +/- 0.3 deg Yaw (Z) 0.0 +/- 0.3 den Repeat Next Direct laser check has completed. Click Next to continue. Repeat Close 8. To access the results in Service Reports, from configuration select Clinical Configuration. Then select Service Reports from the icon in the upper right portion of the screen. ? Configuration Panel Administration Portal Administration Portal [EN] Clinical Configuration Checkup Daily QA Service Reports Calibration Resume syngo Remote Assist Screen recorder aveLog Remote Service Exam Designer Create Screenshot Fast Contact File Browser Load Service Pack Direct Laser Check myNeedle Laser Check What is New Unrestricted | Published by Siemens Healthineers AG, 2024 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 01/18/2024 | QR700004532 Healthineers Page 4 of 7 9. Select which report to view. Service Report Report Status Function Function Start Time Function Status Type Type Runout 2024-01-04 07:31:21 OK Test Direct Direct Laser OK Laser 2024-01-03 09:55:54 OK Check Check Direct Direct Laser Canceled Laser 2024-01-03 09:53:16 Canceled Check Check Calibration OK Calibration 2024-01-03 08:24:32 OK Daily QA Noise 2024-01-03 07:54:41 OK OK Homogenei 2024-01-03 07:53:56 OK Tube Event 2024-01-03 07:25:17 OOK Listener Safety System 2024-01-03 07:23:33 OK Self-Test Calibration 2024-01-03 07:20:32 OK Pixel Quality 2024-01-03 07:20:16 Equalization OK Line Impedance 2024-01-03 07:20:15 OK Generator Checkup OK Adaptation 2024-01-03 07:20:15 OK Gettering 2024-01-03 07:20:14 OK Vacuum Check 2024-01-03 07:19:52 OK Image Quality 2024-01-03 07:19:38 OOK Evaluation Filamant Unrestricted | Published by Siemens Healthineers AG, 2024 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 01/18/2024 | QR700004532 Healthineers Page 5 of 7 10. Select Show PDF, then select Show. Summary Function Start Time Result Direct Laser Check 01/03/2024 9:55:54 AM OK Back Refresh Show Verbose Show PDF Summary PDF Settings: Service Level: Free Paper Format: . A4 Letter Language: English (United States) Options: Attach Raw Data Verbose Show Back Refresh Show Verbose Hide PDF Settings 11. The report contains results and trends over time and can be exported to an external stor- age device, network drive, or configured printer. To export, hover your cursor near the bot- tom of the report and select the floppy disk or printer icon at the left of the icon bar. SIEMENS lealthineers". 1 Direct Laser Trend analysis Results for X plane [mm] Legend 2.5 24 1.5 0.5 0.0 4.5 1.5 2.0 25 $ 3.0 09/13/20123 10/13/2023 12.12 2023 12.64 2123 12/13 2823 #9/22/2023 Results for X plane (Image zoomed to 67%) Results for Y plane [mm] Legend 2.5 SOMATOM go. Open Pro 1.5 Direct Laser Check 0.5 Serial Number: 130214 0.5 -1.0 -1.5 2.0 25 $ 3.0 09/13/21123 #9.22/2023 Results for Y plane (Image zoomed to 67%) Saint Lukes Hospital of Kansas City Siemens Healthcare GmbH, 2024 / 01/03/2024 9:55:54 AM Page 3 of 7 Export 17 Unrestricted | Published by Siemens Healthineers AG, 2024 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 01/18/2024 | QR700004532 Healthineers Page 6 of 7 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthineers AG 2024 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthineers AG Siemens Medical Solutions USA, Inc. Siemens Shanghai Siemensstraße 3 91301 Forcheim, Germany 40 Liberty Blvd. Medical Equipment Ltd. Phone: +49 9131 84-0 Malvern, PA 19355-9998 USA 278 Zhou Shu Road https://www.siemens-healthineers.com Telephone: +1-888-826-9702 Shanghai 201318 Customer Care Telephone: P.R. China +1-800-888-7436 Telephone: +1 847 304 7700 Unrestricted | Published by Siemens Healthineers AG, 2024 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 SIEMENS Job Aid Effective Date: 01/18/2024 | QR700004532 Healthineers Page 7 of 7

  • quality assurance
  • qa