NAEOTOM Alpha Class Spectral Recalibration - USA

NAEOTOM Alpha Class Spectral Recalibration - USA

With the VB20 software version, you can now perform the Spectral Recalibration (SRC) on your own. Step by step, you will learn in this PDF how to perform the SRC on your NAEOTOM Alpha class scanner.  

Product relevance: NAEOTOM Alpha, NAEOTOM Alpha.Peak, NAEOTOM Alpha.Pro, NAEOTOM Alpha.Prime.  
Target group: Basic user, all users.

Recommended to be viewed on the following devices: All (incl. tablet, smartphone).

SIEMENS Healthineers NAEOTOM Alpha Class Spectral Recalibration Job Aid Version 2024.12 Template Effective Date: 10 Aug 2021| HILS 2223 Job Aid Effective Date: 12 Nov 2025 | QR700028685 Unrestricted I Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Spectral Recalibration (SRC) SIEMENS Healthineers Table of Contents Introduction 4 System Check tile status overview 5 SIEMENS Healthineers Green check status 6 SRC steps 8 Yellow check status 12 O. NAEOTOM Alpha.Peak Red check status 14 2 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 SIEMENS Healthineers SIEMENS Healthineers Quantum Technology Spectral NAEOTOM Alpha.Peak Recalibration (SRC) 3 Click to add footnote Effective Date: 27 Mar 2025 | QR700018813 second line Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration SIEMENS Healthineers Introduction Spectral Recalibration (SRC) addresses the long-term drift of the energy response of individual detector modules, which can cause ring artifacts in images. This recalibration resets the spectral drift, although the material properties will start drifting again over time. Purpose SRC helps mitigate artifacts caused by spectral drift, supporting image quality over time. Impact SRC was not possible before the VB20 software version – Siemens Healthineers Customer Service Engineers (CSEs) had to perform tuning functions manually. With the VB20 version, you can now perform the Spectral Recalibration on your own. The system will prompt you to perform this recalibration, typically every four VB20 months. 4 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) SIEMENS Healthineers System Check tile status overview Nothing to do at the moment: Spectral Recalibration: Spectral Recalibration failed: No Spectral Recalibration needed. Is due. It could not restore the previous status. Perform the SRC via the configuration If you have already performed the Check- Execute again. If the problem persists, call menu. up, execute the SRC*. the Siemens Healthineers Customer Care Center**. lealthineers SDK ealthineers 50 Healthineers CH C L¿ Co CH Examination System Check Restart Lock Examination System Check Restart Lock Examination System Check Restart Lock (1) 5 * You can also perform SRC with the next checkup, immediate action not necessarily needed. Effective Date: 27 Mar 2025 | QR700018813 ** System is not blocked. You may decide how to proceed. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) SIEMENS Healthineers Green check status Nothing to do at the moment: Spectral Recalibration: Spectral Recalibration failed: No Spectral Recalibration needed. Is due. It could not restore the previous status. Perform the SRC via the configuration If you have already performed the Check- Execute again. If the problem persists, call menu. up, execute the SRC*. the Siemens Healthineers Customer Care Center**. ealthineers DE Healthineers SD Healthineers CH C L¿ Co CH Examination System Check Restart Lock Examination System Check Restart Lock Examination System Check Restart Lock (1) 6 * You can also perform SRC with the next checkup, immediate action not necessarily needed. Effective Date: 27 Mar 2025 | QR700018813 ** System is not blocked. You may decide how to proceed. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) SIEMENS Healthineers Configuration menu If Spectral Recalibration is not due and you would like to ? execute it via the Configuration menu: Configuration Panel Configuration Administration Portal Administration Portal [EN] • This is not recommended or needed. [WIP] Administration Portal [WPF] • This will update the SRC timer. [WIP] Administration Portal [WPF, NoWeb] Clinical Configuration Checkup Preconditions: Spectral Re-Calibration Resume Daily QA Minimal system uptime: 12 hours.  Before this Calibration • syngo Remote Assist period of time, you cannot execute SRC.  After this Screen recorder period of time, a message will appear, and the GO SaveLog Remote Service button will be interactive (see next page). Exam Designer Create Screenshot Fast Contact File Browser Load Update Pack Direct Laser Check myNeedle Laser Check 7 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) | Step 1 SIEMENS Healthineers A message will tell you SIEMENSSDK Healthineers Dk] O Spectral Re-Calibration x Service, Local x Scan Recon System Check when the system is ready to Spectral Re-Calibration perform SRC. Close Select GO, when you are ready to start SRC. Spectral Re-Calibration is not due. This procedure may take up to 1.5 hours. Spectral Re-Calibration is recommended every 120 days for optimal image quality. X-rays will be produced during the Spectral Re-Calibration procedure. The table will move automatically. Repeat GO Cancel Click GO when ready. EN 8:33 AM 8 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) | Step 2 SIEMENS Healthineers The SRC background SIEMENS HealthineersSDK 0 O Spectral Re-Calibration x Service, Local X Scan Recon System Check configuration process will Spectral Re-Calibration start. Close Another message will tell you to NOT select Cancel during this process. Spectral Re-Calibration background configuration processes have started. This may take up to 5 minutes, please be patie Do not click Cancel. Repeat GO Cancel EN 8: 36 AM 9 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) | Step 3 SIEMENS Healthineers Once the SRC process SIEMENS 0 0 HealthineersSDK Spectral Re-Calibration x Service, Local X Scan Recon System Check prompts you to, press the Spectral Re-Calibration Start key on the Control Close Box. SIEMENS Healthineers + es have started. This may take up to 5 minutes, please be pati Do not click Cancel. * EN 8: 36 AM 10 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) | Step 4 SIEMENS Healthineers After pressing the Start key SIEMENS Healthineers 106 12 Spectral Re-Calibration ? Scan Recipe System Check on the Control Box, the SRC Spectral Re-Calibration will start. Close Another message about it appears. A progress bar will inform you about the SRC progress in % until it is complete. Energy Threshold Calibration @ Spectral Re-Calibration @ System Check in progress ... 1% Function Energy Threshold Calibration @ Spectral Re-Calibration @ System Check started. Repeat GO Cancel Gantry accelerating 100 kV. 0 mA, 2.2 s, 288x0.2 mm (A) Resuming was successful 11 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) SIEMENS Healthineers Yellow check status Nothing to do at the moment: Spectral Recalibration: Spectral Recalibration failed: No Spectral Recalibration needed. Is due. It could not restore the previous status. Perform the SRC via the configuration If you have already performed the Check- Execute again. If the problem persists, call menu. up, execute the SRC*. the Siemens Healthineers Customer Care Center**. lealthineers SDK Healthineers SD SIEMENS CH C L¿ Co CH Examination System Check Restart Lock Examination System Check Restart Lock Examination System Check Restart Lock (1) 12 * You can also perform SRC with the next checkup, immediate action not necessarily needed. Effective Date: 27 Mar 2025 | QR700018813 ** System is not blocked. You may decide how to proceed. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) is due SIEMENS Healthineers Untersuchung Systemprüfung Neu starten Sperren SRC is due – no immediate action Select the System Check SRC was successful. necessary, but plan to perform it tile. Then select GO on in the next couple of days*. the following screen. Preconditions 12 hours of minimal system uptime and SRC last performed more than 120 days ago. 13 * The System Check tile will not turn red if you do not execute SRC. Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) SIEMENS Healthineers Red check status Nothing to do at the moment: Spectral Recalibration: Spectral Recalibration failed: No Spectral Recalibration needed. Is due. It could not restore the previous status. Perform the SRC via the configuration If you have already performed the Check- Execute again. If the problem persists, call menu. up, execute the SRC*. the Siemens Healthineers Customer Care Center**. lealthineers SDK ealthineers 50 Healthineers CH C L¿ Co CH Examination System Check Restart Lock Examination System Check Restart Lock Examination System Check Restart Lock (1) 14 * You can also perform SRC with the next checkup, immediate action not necessarily needed. Effective Date: 27 Mar 2025 | QR700018813 ** System is not blocked. You may decide how to proceed. Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration error cases – SRC fails SIEMENS Healthineers If you can restore the previous system status: Rollback  The SRC status is set back to its previous status, successful if not due, it is displayed in green, if due, it is displayed in yellow.  Perform SRC again. 1. Close the System Check. 2. Select the System Check tile. CH System Check 3. Select GO. Close the System Check. 15 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration error cases – SRC fails SIEMENS Healthineers SIEMENS Healthineers 106 12 Scan Spectral Re-Calibration System Check Rollback Spectral Re-Calibration Close fails Close the System Check. If the previous system status could not be restored: CH LC Examination System Check Spectral Re-Calibration has failed. The CT system cannot be used currently. Perform Spectral Re-Calibration again. If the problem persists, contact your Customer Care Center.  Try SRC again. Repeat GO Cancel  If the problem persists, inform the Siemens Healthineers Customer Care Center. 16 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 Spectral Recalibration (SRC) – Cancel SIEMENS Healthineers X Do NOT cancel during restore  Cancel SIEMENS Healthineers x Cancel Spectral He-Calibration button is not deactivated, but there is a message in app log while restore is running. X Spectral Re-Calibration failed. The falkwing warnings occurred during the execution of the workflow: Function 'Energy Ihreshold Callbration & Spectral Re-Calibration & System Check aborted by error. Workflow Spectral He-Lal bration & System Check aborted by error. Perform Spectral Re-Calibration agsin. If the problem persists, contact your Customer Care Center. Repeat Next Cancel 17 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025 SIEMENS Healthineers Please note that this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training won’t be updated regularly and doesn’t necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product-related claims, or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said that Functionalities in any specific country aren’t guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission, or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters, and configuration-dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com 18 Effective Date: 27 Mar 2025 | QR700018813 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2025

  • naeotom
  • neotom
  • alpha
  • alfa
  • class
  • vb20
  • src
  • spectral
  • recalibration