This job aid outlines the steps to perform Smart Virtual Ortho exams on MULTIX Impact systems with a camera and auto-detection.

MULTIX Impact Smart Virtual Ortho (w/ Camera and Auto Detection) MULTIX Impact VA20 and Higher Prerequisites: • Smart Virtual Ortho is configured • An Ortho Long Leg or Full Spine organ program is selected • The collimator must be in the 0° lock-in-position • The tube unit is positioned in detent and centered to the detector at the correct SID • For Ortho acquisitions at the wall stand, the detector unit must be in the 0° position • Position the patient upright on the Ortho Stand or flat on the tabletop • The Ortho menu is displayed on the Tube TUI Process Steps: 1. Double-click on the proper OGP to open the tool. Ortho OGP list FullSpine ap_w180_t130 S 15 2. Measure the Table-to-Object Distance (TOD) and enter on Control Console. SIEMENS Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Healthineers Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS 3. For Region of Interest Auto-Detection, select the camera icon on the imaging system. The ROI is automatically detected and shown with a blue frame on the imaging system. Patient information jone, jone, 202011121355111 Male 6-16-1992 202011121355111 S. Device Information Dose Information Intermotion 4. If necessary, adjust the upper and lower border lines to define the imaging range on the Control Console. Collimating the left or right border lines will adjust the width of your region of interest (ROI). Performing Acquisitions and Postprocessing steps: 1. Ensure the correct exposure factors are set for each required step in the ortho series at the Control Console. 2. With the patient maintaining positioning, press and hold the acquisition switch until all exposures are completed, and the tube returns to its original position. Note: Releasing the acquisition switch before all images are acquired will result in the interruption of the series. The exam will need to be repeated in its entirety. 3. Once the series acquisition is complete, release the switch. Auto-composing of the image then occurs. 4. Measurements can be made to the composed image before saving under the View tab. Stitch Print Name: STITCH, , TEST Sex M Acquisition date 2023:02.0 ID: Test11 17:07.06 Date of birth: 1998-02-02 ACAN Operator Siemens View Age: 025Y Request ID: Accession no.: Test 11 Measurement 4 T. L R Ta AA MAs KV Reference EXI: Current EXI: DI DAP OGP name: Ortho_Finished Grid type: FOCUSED WC 3232 Abort SIEMENS Unrestricted I Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Page 2 of 4 5. Adjustments to the composing process can be made in the Composing tab. View Composing Composing controls Step 5 Transition mode Blend Transparent 6. Steps are adjusted from first to last by shifting the lower image left, right, up, or down relative to the image above. Select Next after each image. 7. Select Save once completed to finish the process and save the composed image. Acquiring a second view steps: 1. Return to the examination tab and select Add OGP. Examination Postprocessing Stitch View Print Patient Information OGP List Patient Name: 2. Double-click the proper OGP for the Ortho exam and click the add button. 3. Repeat the steps above. SIEMENS Unrestricted I Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Page 3 of 4 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2023 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthcare GmbH Siemens Medical Solutions USA, Inc. Siemensstrasse 1 Healthcare Sector Henkestr. 127 40 Liberty Blvd. Forchheim, Germany 91301 91052 Erlangen, Germany Malvern, PA, USA 19355-9998 Phone: +49 9131 84-4746 Phone: +49 9131 84-0 Telephone: +1-888-826-9702 siemens-healthineers .com/radiography siemens-healthineers.com siemens-healthineers .com/en-us SIEMENS Unrestricted I Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: XX/XX/XXXX | HOOD/HILS Page 4 of 4

  • va21
  • xray
  • othopedic