Siemens Healthineers Academy

Multitom Rax - Detector Calibration Job Aid

In this PDF, you will learn about how to calibrate the detector of the Multitom Rax system.

 

Target group: All users
Audio: Yes
Recommended to be viewed on the following devices: Laptop, desktop computer (sufficiently large display required)

SIEMENS Healthineers Multitom Rax - 1 Detector 2 Calibration (Fluorospot .... Compact) Template Effective Date: 26 Mar 2025 | QR700020784 1 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Multitom Rax - Detector Calibration SIEMENS Healthineers (Fluorospot Compact) Preconditions for successfull detector calibration 3 Step-by-Step guide: Detector calibration portable detector 7 Step-by-Step guide: Detector calibration ceiling mounted detector 19 Template Effective Date: 26 Mar 2025 | QR700020784 2 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Preconditions for successfull detector calibration Template Effective Date: 26 Mar 2025 | QR700020784 3 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 General detector calibration information SIEMENS Healthineers • Always follow the instructions on the FLC exactly. • The detector temperature must be constant (this is checked by the system) before the calibration is • All mobile detectors of the same type (MAX wi-D started. or MAX mini) have to be disconnected except the one to be calibrated. Combinations allowed: 1 MAX • All mobile detectors have to be calibrated in free wi-D or 1 MAX mini or 1 MAX wi-D + 1 MAX mini. positions. • No objects may be in the beam path. • After successful calibration of the detector, the complete system including the imaging system has to be rebooted. Template Effective Date: 26 Mar 2025 | QR700020784 4 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Preconditions for successfull detector calibration SIEMENS Healthineers Portable detectors • The portable detector must be warmed up for at least 1 hour (battery inserted). • The detector may not have been exposed to radiation for at least 7 hours prior to calibration. • The portable detector must be free of any soiling. • The detector must be calibrated outside the trays with an SID of 150 cm in landscape or portrait format. SIEMENS MAX mini The detector edges must be overexposed slightly SIEMENS • Heutthineers (about 2 cm; check with the light field). Template Effective Date: 26 Mar 2025 | QR700020784 5 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Preconditions for successfull detector calibration SIEMENS Healthineers RAX detector • The RAX detector must be warmed up at least 4.5 hours (system switched on). • The detector may not have been exposed to radiation for at least 7 hours prior to calibration. This also applies for previous calibration procedures. • If calibration fails, a time period of at least 7 hours must expire before calibration can be attempted again. -- Template Effective Date: 26 Mar 2025 | QR700020784 6 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Detector calibration: Portable detector Template Effective Date: 26 Mar 2025 | QR700020784 7 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector – System setup SIEMENS Healthineers Template Effective Date: 26 Mar 2025 | QR700020784 8 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector – Service user SIEMENS Healthineers • Enter the service by clicking on the icon. @ siemens Healthcare Service Software - Internet Explorer X Tune Up Home Logout Tube Load Component Adjust FD Calibration TUI 1 Enter the service as service level 7 user. Stand Select the detector for calibration • Centering Detector System Adjust Calibrate FD Detector Serial Number Last calibration Inverter MAX wi-D 2013C6 29-07-2022 Tube mAs Integrated Detector in ceiling stand SN1748C2 18-05-2020 O o Dose lontomat • Navigate to Tune-Up > Detector FD-Calibration 1 Control Fluoro Additional Tools Calibration. Manual Calibration 1 Error Suppress List • Select the particular detector and start the Calibration menu with Next. 2 2 After selection, click 'Next" Next | Exit Template Effective Date: 26 Mar 2025 | QR700020784 9 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector – Admin user SIEMENS Healthineers Siemens Healthcare Service Software - Internet Explorer X • Enter the service as clinical admin user. Tune Up Home | Logout User Password Component Adjust FD Calibration Centering System Adjust • Navigate to Repair > Tune-Up > Detector FD Calibration 1 Select the detector for calibration 3D Calibration Calibrate FD Detector Serial Number Last calibration Calibration. Projection Geometry 1 O MAX wi-D 2013C6 29-07-2022 Integrated Detector in ceiling stand SN1748C2 18-05-2020 • Select the particular detector and start the Calibration menu with Next. 2 2 After selection, click 'Next" Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 10 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers The calibration switch must be switched on. For the Multitom RAX systems the calibration switch can be found on the back of the RCC. 1 O CAL 1 Template Effective Date: 26 Mar 2025 | QR700020784 11 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers • All mobile detectors must be calibrated on the floor. (Do not place the detector directly on the floor!) • Center the tube unit to the detector and set the SID specified by the UIS. Calibration of the portable detector is performed in landscape or portrait format. It is mandatory to overexpose the detector (circa 2 cm / 0.8 inch). 1 1 Template Effective Date: 26 Mar 2025 | QR700020784 12 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers • Follow the instructions on the imaging system exactly. Once a condition is fulfilled, the red X changes to a green checkmark. 1 • If all conditions are met, the Next button appears in the status bar. 2 @ Siemens Healthcare Service Software - Internet Explorer X SERVICE Siemens Healthcare Service Software - Internet Explorer 0 SERVICE X Tune Up Home | Event Log | Reports | Help | Logout uGy*m2 Tune Up Home | Event Log | Reports | Help | Logout Service Service O uGy*m2 Tube Load FD Calibration 0.00 Tube Load FD Calibration 0.00 Component Adjust Component Adjust TUI 1 TUI 1 Stand Detector: MAX wi-D Serial number: 2013C6 Stand Detector: MAX wi-D Serial number: 2013C6 Centering Centering Detector Detector System Adjust Step 1 of 4 - check FD parameters for calibration System Adjust Step 1 of 4 - check FD parameters for calibration Untersuchung Untersuchung Inverter Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! Inverter Tube Service Tube Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! mAs Please wait while systems verifies detector parameters Einstell mAs Please wait while systems verifies detector parameters Einstell. Service Dose lontomat Dose lontomat FD-Calibration FD-Calibration Control Fluoro Parameter Ready Current value Remarks Control Fluoro Parameter Ready Current value Remarks Additional Tools Additional Tools Manual Calibration Manual Calibration Error Suppress List SID X 114.4 cm Please set SID to 150,0 cm Error Suppress List SID V 150,0 cm Nachverarbeitung Nachverarbeitung DO Grid 1 V Grid 1 V DC Temperature constant X 35.6 ºC Please wait ... (00:23) Temperature constant 35.7 ºC Dokumentation Dokumentation 2 System is verifying detector parameters ... Manual Back Exit Please click 'Next' to select the calibration mode Manual Back Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 13 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers X SERVICE • A status dialog appears when Next is @ siemens Healthcare Service Software - Internet Explorer 0 Tune Up Home Event Log Reports Help Logout uGy*m2 Service selected. Tube Load FD Calibration 0.00 Component Adjust TUI 1 Stand Detector: MAX wi-D Serial number: 2013C6 Centering Detector 1 System Adjust Step 2 of 4 - select the calibration mode(s) • Select the modes to be calibrated. Inverter Instructions: Please select the mode you want to calibrate Untersuchung Tube mAs Mode Select Status Progress Reset Einstell. Service Dose lontomat FD-Calibration Control Fluoro DR Radiography • To start the calibration, select Go. Last calibration: 23-6-2022 2 Additional Tools 1 Manual Calibration DR Tomography / Long Time Exposure Last calibration: 23-6-2022 LX Error Suppress List liti Nachverarbeitung DO • Follow the instructions on the imaging system. 3 Dokumentation Message from webpage X X-ray will be automatically released during calibration. Ensure that nobody is in the examination room. You can stop the calibration at any time by clicking "Abort' 3 OK 2 Click 'Go' to start the calibration Manual Back Go Exit ES Template Effective Date: 26 Mar 2025 | QR700020784 14 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers Follow the instructions on the imaging system. Siemens Healthcare Service Software - Internet Explorer X SERVICE Tune Up Home Event Log Reports Help Logout Service Insert the 0.6 mm Cu Filter and press OK. uGy*m2 1 Tube Load Component Adjust FD Calibration 0.00 O O calib fd4343rc 30 table TUI 1 Stand Detector: MAX wi-D Serial number: 2013C6 Centering The Cu Filter can be found in the blue Detector System Adjust Step 3 of 4 - system calibrates selected mode(s) Untersuchung Inverter Instructions: Please wait while the system performs acquisitions Tube The system then will process the acquired images accessory folder. mAs Dose lontomat Mode Select Status Progress Reset FD-Calibration Control Fluoro Additional Tools Manual Calibration DR Radiography Started Preparation, please wait ... 46.9 cm × 46.9 cm Error Suppress List DR Tomography / Long Time Exposure Last calibration: 23-6-2022 130 cm Ist-Wert Soll-Wert Nachverarbeitung ohne ohne Message from webpage X KV 70 ? Is the following filter inserted: - 0.6 mm Cu mAs 1 Dokumentation Click 'Cancel' to abort the calibration ms 100 Dosis 1 OK Cancel S SIEMENS mm Cl Fokus % 80 NounVO Bild Extras 1 - System is performing acquisitions ... Abort Template Effective Date: 26 Mar 2025 | QR700020784 15 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer SERVICE • Wait until the calibration has X Tune Up Home Event Log Reports Help Logout uGy*m2 Service finished. Do not enter the room or 4 User Password FD Calibration 3346.0 O Tube Load calib fd3543ezh 31 free Component Adjust interrupt the calibration! At the end TUI 1 Detector: MAX wi-D Serial number: 2013C6 TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) of the calibration, an indicator is Stand Centering Please wait while the system performs acquisitions Detector Instructions: The system then will process the acquired images Examination System Adjust Fluoro Rad displayed showing or not the Inverter Mode Select Status Progress Reset Tube mAs calibration was successful. Dose lontomat 1 FD-Calibration DR Radiography M Accepted Finished (not yet saved) Dose DEXI Water Equivalent DR Tomography / Long Time Exposure Started 53 % processing # 47.0 cm x 47.0 cm Control Fluoro 8 150 cm Collimation Postprocessing 3D Calibration Current Target Projection Geometry none • Press OK to save the calibration Reference 10 2D-Topo Calibration Message from webpage Reference 10 1 × results. 2 Additional Tools Manual Calibration Error Suppress List ? The calibration has finished successfully. KV 70 Save the calibration results now? mAs Documentation Click 'Cancel' to discard all calibration results or save them later. ms Dose 2 OK Cance mm C 0.0 Focus % 80 Image Tools System is processing acquired images ... Abort S/N:2013C6 Ready 2013C6 O 7%HU Template Effective Date: 26 Mar 2025 | QR700020784 16 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer X SERVICE Tune Up Home Event Log Reports Help Logout uGy*m2 Service Wait until the results have been saved. O • 4 1723.0 User Password FD Calibration Tube Load calib fd3543ezh 30 free The progress area will be updated Component Adjust TUI 1 Detector: MAX wi-D Serial number: 2013C6 TUI 2 TUI 3 accordingly. Step 3 of 4 - system calibrates selected mode(s) 1 Stand Instructions: 1 Centering Examination Detector System Adjust Mode Select Status Progress Reset Settings Service Inverter Tube mAs Confirm with OK. DR Radiography Successful Saved • 2 Dose lontomat FD-Calibration DR Tomography / Long Time Exposure Successful Saved Dose DEXI Water Equivalent Control Fluoro Postprocessing Collimation 3D Calibration • Press Exit to leave the detector Projection Geometry Reference 10 2D-Topo Calibration calibration. Reference 10 Additional Tools Message from webpage X Manual Calibration Error Suppress List Calibration results have been saved sucessfully Documentation • Disengage the calibration switch and reboot the system. 2 OK Synchronizing new calibration data, please wait .. Abort S/N:2013C6 Ready 2013C6 O 10%HL Template Effective Date: 26 Mar 2025 | QR700020784 17 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a portable detector SIEMENS Healthineers e Siemens Healthcare Service Software - Internet Explorer O SERVICE • Press Exit to leave the detector X Tune Up Home Event Log Reports Help Logout uGy*m2 Service calibration. 1 User Password FD Calibration 4 1723.0 O Tube Load calib fd3543ezh 30 free Component Adjust TUI 1 Detector: MAX wi-D Serial number: 2013C6 TUI 2 TUI 3 Step 4 of 4 - system has calibrated the selected mode(s) An information message comes up to Stand Centering Instructions: • Examination Detector System Adjust Mode Select Status Progress Reset Settings Service deactivate the calibration switch. 2 Inverter Tube mAs DR Radiography Successful Saved Dose lontomat FD-Calibration DR Tomography / Long Time Exposure Successful Saved Dose DEXI Water Equivalent • Disengage the calibration switch and Control Fluoro Postprocessing Collimation 3D Calibration reboot the system. Projection Geometry Reference 10 2D-Topo Calibration Reference 10 Additional Tools Message from webpage X Manual Calibration Error Suppress List Please deactivate the listed switch(es): Calibration Switch Documentation 2 OK 1 Click 'Next' to select another detector. Click 'Go' to repeat the calibratio Manual Next Go Exit S/N:2013C6 Ready 2013C6 10%HU O Template Effective Date: 26 Mar 2025 | QR700020784 18 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Detector calibration: Ceiling mounted detector Template Effective Date: 26 Mar 2025 | QR700020784 19 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of the ceiling mounted detector SIEMENS Healthineers System setup H 70 1 SIEMENS Template Effective Date: 26 Mar 2025 | QR700020784 20 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector – Service user SIEMENS Healthineers • Enter the service as service level 7 user @ Siemens Healthcare Service Software - Internet Explorer X Tune Up Home Logout and navigate to Repair > Tune-Up > Tube Load Component Adjust FD Calibration TUI 1 Detector Calibration. 1 Stand Select the detector for calibration Centering Detector System Adjust Calibrate FD Detector Serial Number Last calibration Inverter O MAX wi-D 2013C6 29-07-2022 Tube Integrated Detector in ceiling stand SN1748C2 18-05-2020 mAs Select the particular detector and start Dose lontomat • FD-Calibration 1 Control Fluoro the Calibration menu with Next. Additional Tools 2 Manual Calibration Error Suppress List 2 After selection, click 'Next' Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 21 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector – Admin user SIEMENS Healthineers e Siemens Healthcare Service Software - Internet Explorer X • Enter the service as clinical admin user. Tune Up Home Logout User Password Component Adjust FD Calibration Centering • Navigate to Repair > Tune-Up > Detector System Adjust Select the detector for calibration FD Calibration 1 3D Calibration Calibration. Projection Geometry Calibrate FD Detector Serial Number Last calibration 1 MAX wi-D 2013C6 29-07-2022 O Integrated Detector in ceiling stand SN1748C2 18-05-2020 O • Select the particular detector and start the Calibration menu with Next. 2 2 After selection, click 'Next' Next Exit Template Effective Date: 26 Mar 2025 | QR700020784 22 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector SIEMENS Healthineers The calibration switch must be switched on. For the Multitom Rax systems the calibration switch can be found on the back of the RCC. 1 O CAL 1 Template Effective Date: 26 Mar 2025 | QR700020784 23 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector SIEMENS Healthineers • Follow the instructions on the imaging system exactly. Once a condition is fulfilled, the red X changes to a green checkmark. 1 • If all conditions are met, the Next button appears in the status bar. 2 @ siemens Healthcare Service Software - Internet Explorer 0 SERVICE Siemens Healthcare Service Software - Internet Explorer SERVICE X X Tune Up Home Event Log | Reports | Help | Logout uGy*m2 Tune Up Home | Event Log Reports Help Logout uGy*m2 Service Service 0.00 O Tube Load FD Calibration Tube Load FD Calibration 0.00 O Component Adjust Component Adjust TUI 1 TUI 1 Stand Detector: Integrated Detector in ceiling stand Serial number: SN1748C2 Stand Detector: Integrated Detector in ceiling stand Serial number: SN1748C2 Centering Centering Detector Detector System Adjust Step 1 of 4 - check FD parameters for calibration System Adjust Step 1 of 4 - check FD parameters for calibration Untersuchung Inverter Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! Inverter Untersuchung Tube Tube Instructions: Please assure that no additional objects are in the beam path (e.g. filter)! mAs Please wait while systems verifies detector parameters Einstell. Service mAs Please wait while systems verifies detector parameters Einstell. Service Dose lontomat Dose lontomat FD-Calibration Calibr Control Fluoro Parameter Ready Current value Remarks Control Fluoro Parameter Ready Current value Remarks Additional Tools Additional Tools Manual Calibration Manual Calibration Error Suppress List SID 114.4 cm Please set SID to 130.0 cm Error Suppress List SID V 114.4 cm Please set SID to 130.0 cm Nachverarbeitung X Grid 1 Grid 1 Nachverarbeitung V 3. Temperature constant X 35.6 ºC Please wait ... (00:23) Temperature constant 35.6 ºC Please wait ... (00:23 2 Dokumentation Dokumentation 2 System is verifying detector parameters ... Manual Back ----- Exit System is verifying detector parameters ... Manual Back ----- Exit Template Effective Date: 26 Mar 2025 | QR700020784 24 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector SIEMENS Healthineers @ siemens Healthcare Service Software - Internet Explorer X SERVICE • A status dialog appears when Next is Tune Up Home Event Log Reports Help Logout O uGy*m2 Service selected. Tube Load FD Calibration 0.00 Component Adjust TUI 1 Detector: Integrated Detector in ceiling stand Serial number: SN1748C2 Stand Centering Detector Select the modes to be calibrated. 1 System Adjust Step 2 of 4 - select the calibration mode(s) • Instructions: Please select the mode you want to calibrate Untersuchung Inverter Tube mAs Mode Select Status Progress Reset Einstell. Service Dose lontomat FD-Calibration Control Fluoro DR Radiography • To start the calibration, select Go. Last calibration: 23-6-2022 2 Additional Tools Manual Calibration DR Tomography / Long Time Exposure 1 Last calibration: 23-6-2022 Error Suppress List 0 0 Nachverarbeitung DC • Follow the instructions on the imaging system. 3 Dokumentation Message from webpage X X-ray will be automatically released during calibration. Ensure that nobody is in the examination room. You can stop the calibration at any time by clicking 'Abort' 3 OK 2 Click 'Go' to start the calibration Manual Back Go Exit Template Effective Date: 26 Mar 2025 | QR700020784 25 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector SIEMENS Healthineers e Siemens Healthcare Service Software - Internet Explorer O X SERVICE Follow the instructions on the imaging system. Tune Up Home Event Log Reports Help Logout uGy*m2 Service Insert 0.6 mm Cu filter and press OK. FD Calibration 0.00 O O 1 Tube Load Component Adjust calib fd4343rc 30 table TUI 1 Detector: Integrated Detector in ceiling stand Serial number: SN1748C2 Stand Centering Detector The Cu filter can be found in the blue System Adjust Step 3 of 4 - system calibrates selected mode(s) Inverter Instructions: Please wait while the system performs acquisitions Untersuchung Tube The system then will process the acquired images accessory folder. mAs Dose lontomat Mode Select Status Progress Reset FD-Calibration Control Fluoro Additional Tools DR Radiography Started Preparation, please wait ... Manual Calibration U # 46.9 cm x 46.9 cm Error Suppress List DR Tomography / Long Time Exposure Last calibration: 23-6-2022 130 cm I DO Replace the 0.6 mm Cu filter with the 2.1 mm Ist-Wert Soll-Wert Nachverarbeitung ohne ohne Cu filter when requested by the system. 1 Message from webpage X k 70 ? Is the following filter inserted: - 0.6 mm Cu mAs 1 Dokumentation Click 'Cancel' to abort the calibration ms 100 DOSIS 1 OK Cancel S mm Cu 0.0 000 Fokus % 80 Bild Extras 1 System is performing acquisitions ... Abort Template Effective Date: 26 Mar 2025 | QR700020784 26 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer X SERVICE • Wait until the calibration has finished. Tune Up Home Event Log Reports Help Logout uGy*m2 Service 4 Do not enter the room or interrupt the User Password FD Calibration 3346.0 O Tube Load calib fd3543ezh 31 free Component Adjust TUI 1 Detector: MAX wi-D Serial number: 2013C6 calibration! At the end of the calibration, TUI 2 TUI 3 Step 3 of 4 - system calibrates selected mode(s) Stand an indicator is displayed showing Centering Detector Instructions: Please wait while the system performs acquisitions The system then will process the acquired images Examination System Adjust Fluoro Rad whether or not the calibration was Inverter Mode Select Status Progress Reset Tube mAs Dose lontomat successful. FD-Calibration DR Radiography M Accepted Finished (not yet saved) 1 Dose DEXI Water Equivalent DR Tomography / Long Time Exposure Started 53 % processing # 47.0 cm x 47.0 cm Control Fluoro 150 cm Collimation Postprocessing 3D Calibration Current Target Projection Geometry none • Press OK to save the calibration results. 2 Reference 10 2D-Topo Calibration Message from webpage × Reference 10 1 Additional Tools Manual Calibration Error Suppress List ? The calibration has finished successfully. KV 70 Save the calibration results now? mAs Documentation Click 'Cancel' to discard all calibration results or save them later. ms Dose 2 OK Cance mm C 0.0 Focus % 80 Image Tools System is processing acquired images ... Abort S/N:2013C6 Ready 2013C6 O 7%HU Template Effective Date: 26 Mar 2025 | QR700020784 27 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 Calibration of a ceiling mounted detector SIEMENS Healthineers @ Siemens Healthcare Service Software - Internet Explorer SERVICE X • Press Exit to leave the detector Tune Up Home Event Log Reports Help Logout uGy*m2 Service calibration. FD Calibration O 4 1723.0 User Password 1 Tube Load calib_fd3543ezh_30 free Component Adjust TUI 1 Detector: Integrated Detector in ceiling stand Serial number: SN1748C2 TUI 2 TUI 3 Step 4 of 4 - system has calibrated the selected mode(s) Stand An information message comes up to Centering Instructions: Examination • Detector System Adjust Mode Select Status Progress Reset Settings Service Inverter deactivate the calibration switch. 2 Tube mAs DR Radiography Successful Saved Dose lontomat FD-Calibration DR Tomography / Long Time Exposure Successful Saved Dose DEXI Water Equivalent Control Fluoro ilh Postprocessing • Disengage the calibration switch and Collimation 3D Calibration Projection Geometry reboot the system. Reference 10 2D-Topo Calibration Reference 10 Additional Tools Message from webpage X Manual Calibration Error Suppress List Please deactivate the listed switch(es): Calibration Switch Documentation 2 OK 1 Click 'Next' to select another detector. Click 'Go' to repeat the calibration Manual Next Go | Exit S/N:2013C6 Ready 2013C6 10%HU Template Effective Date: 26 Mar 2025 | QR700020784 28 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025 SIEMENS Healthineers Please note that this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product-related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration-dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers Headquarters Siemens Healthineers AG Siemensstr. 3 91301 Forchheim, Germany Phone: +49 9191 180 siemens-healthineers.com Template Effective Date: 26 Mar 2025 | QR700020784 29 Job Aid Effective Date: 25 Jun 2025 | QR700025013 Published by Siemens Healthineers AG I © Siemens Healthineers, 2025

  • Multitom RAX Detector Calibration