MR Shut down & Restart - Free. Systems
This job aid provides guidance on how to perform a shutdown, reboot, and place an MR System in standby for routine maintenance. It is applicable for the following MAGNETOM MR Free. Systems on Numaris X software: Free.Max and the Free.Star.
MR Shut down, Restart, & Standby – Free. Systems Job Aid MR Shut down, Restart, Standby - Free. Systems Job Aid This job aid provides guidance on how to perform a shutdown, reboot, and place an MR System in standby for routine maintenance. It is applicable for the following MAGNETOM MR Free. Systems on Numaris X software: Free.Max and the Free.Star. Siemens Healthineers MR Customer Service recommendations: • Siemens Healthineers recommends that the MR System be shut down daily. If not performed daily it is recommended to shut down and reboot every other day. • After shutting down (System Off), leave the MR System off a minimum of two (2) minutes before restarting/rebooting the MR system. System powered up (ON) • System currently has error messages that require an orderly shutdown and reboot. • System has a scheduled shutdown and restart. Shutting down the computer system (System Off) You can shut down the computer system in the software or at the intercom. Shutting down the system in the software • When the system is switched off, the host computer, and MaRS (Measurement and Reconstruction System) also shuts down. • The system can only be shut down when all jobs using local resources are completed. (e.g., if a scan is currently running, the Shutdown icon on the Home screen is dimmed). • Save all patient and postprocessing data at the syngo Acquisition Workplace. • Close all workflows, examinations, and applications. • Coils on the patient table are fully connected in the coil sockets. • Patient table must be in the home position. (i.e., table up and completely out of the bore). 1. On the access bar, click the Home icon to open the Home screen. 2. On the Home screen, click the Shut Down icon and choose Shut Down System. C CH Restart System System Check Shut Down System myExam Cockpit Examination Lock FuSa Simulator Restart Workplace Standby System Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 02/03/2024 | QR700018393 Healthineers 3. Shut down System window opens, click Shut down (shut down process displayed). a. The computer shuts down. b. Shut down turns on the power saving mode for the MR cooling system. c. Monitor goes Off (blank) Shut down System ? X Do you want to shut down the system? If any applications/studies are still open, then the system will attempt to save all studies/results and to close all open applications. Shut down Cancel Switching Off the System at the Intercom • When the system is switched off, the host computer, and MaRS (Measurement and Reconstruction System) also shuts down. • The system can only be shut down when all jobs using local resources are completed. (If a scan is currently running, the Shutdown icon on the Home screen is dimmed). • Save all patient and postprocessing data at the syngo Acquisition Workplace. • Close all workflows, examinations, and applications. • Coils on the patient table are fully connected in the coil sockets. • Patient table must be in the home position. (table up and completely out of the bore). 1. On the intercom, depress the System ON/OFF button for two (2) seconds. a. Command is followed by an acoustic signal (beep) that is audible. b. The computer shuts down. c. Shutting down turns on the power saving mode for the MR cooling system. d. Monitor goes Off (blank) 2. Perform only if the System does not respond! On the Intercom, depress the System ON/OFF button for thirteen (13) seconds. Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 02/03/2024 | QR700018393 Healthineers Page 2 of 6 Restarting the syngo MR Acquisition Workplace • If the Numaris X system is not responding or the MaRS is not working correctly, you can restart the workplace in the System On mode. • Save all patient and postprocessing data at the syngo Acquisition Workplace. • Close all workflows, examinations, and applications. • Coils on the patient table are fully connected in the coil sockets. • Patient table must be in the home position. (i.e., table up and completely out of the bore). 1. The Home Screen is open. 2. On the Home screen, select the Shut Down icon and choose Restart Workplace. C X Restart System System Check Examination Shut Down System myExam Cockpit Lock FuSa Simulator Restart Workplace Standby System 3. The Restart Workplace window opens; click Restart. The screen will display the restart process. a. syngo software on the workplace shuts down and restarts the measure- Restart Workplace ? X ment and reconstruction (MaRS) sys- tem software. ? Do you want to restart this workplace? If any applications/studies are still open, then the system will b. Operating systems are not shut down. attempt to save all studies/results and to close all open applications. c. Computer system boots automatically. When restarting (booting) the system, Restart Cancel a splash screen is displayed on the workplace. 4. Wait until the system is ready for measurement, you will hear three (3) knocks/beeps. Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 02/03/2024 | QR700018393 Healthineers Page 3 of 6 Starting Standby • If you do not intend to perform measurements for a longer time or only need to edit or evaluate images, save energy by enabling Standby mode. • MaRS shuts down, and the power-saving mode of the MR cooling system turns on. • You cannot perform measurements in Standby mode. 1. The Home screen is open. 2. To start Standby, click the Shut down icon and select Standby System. C HealthMES > CH Restart System System Check myExam Cockpit Examination Shut Down System lock FuSa Simulator Restart Workplace Standby System 3. In the Standby System window, click Standby. Standby System ? X i Do you want to activate standby mode? In standby mode, any open applications/studies are kept in the memory. The system goes into the low-power state to save energy and you can nevertheless resume work quickly. Standby Cancel Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 02/03/2024 | QR700018393 Healthineers Page 4 of 6 Ending Standby 1. To End Standby, on the Intercom, press the System ON/OFF button for two (2) seconds. -OR- 2. Click the Shutdown icon and select Wake Up System. Restart System Shut Down System FuSa Simulator Restart Workplace Wake Up System Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 02/03/2024 | QR700018393 Healthineers Page 5 of 6 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collec- tively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not neces- sarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthineers AG 2024 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthineers AG Siemens Medical Solutions USA, Inc. Imaging & Therapy Systems Siemensstraße 3 40 Liberty Blvd. Magnetic Resonance 91301 Forcheim, Germany Malvern, PA 19355-9998 USA Henkestr. 127 Phone: +49 9191 18-0 Telephone: +1-888-826-9702 DE-91052 Erlangen siemens-healthineers.com Customer Care Telephone: Germany +1-800-888-7436 Telephone: +49 9191 18-0 Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 02/03/2024 | QR700018393 Healthineers Page 6 of 6
- shutdown
- shut down
- restart
- reboot