MR CSI Scanning Workflow - Numaris X Job Aid
Multi Voxel Spectroscopy Scanning Workflow (Numaris X) Step by Step
MR CSI Scanning Workflow Numaris X Job Aid MR Multi-Voxel Spectroscopy (CSI) Scanning workflow - Numaris X Job Aid 1. Images acquired in all 3 planes are used to position the voxel (VOI). a. Routine sequences (e.g., T1,T2, etc.) are used to position the voxel. (i.e., localizers can be used to position the voxel, but routine sequences have better image quality) b. Acquire series with Full Brain coverage in all three (3) planes. c. Table Positioning Strategy is Local Range. d. Hint: series can be acquired orthogonal, oblique, or double oblique. 2. Click on the myExam Cockpit icon to open myExam Cockpit. a. Choose any Multi-Voxel Spectroscopy (CSI) sequence. SIEMENS >> spectroscopy >> head_1H > csi localizer 00:09 center on brain Table Positioning Strategy localizer_5@center 00:40 loc_in-plane_csi@center 01:26 t1_mprage_sag_p2 03:14 csi_se 30 07:12 Spectroscopy Auto Start MR Spectro Advanced Application csi_se_135 07:12 Spectroscopy csi_se_270 07:12 Spectroscopy csi3d_se_135 09:30 Spectroscopy csi_st_20 07:12 Spectroscopy csi st 135 07:12 Spectroscopy Csi_st_270 07:24 Spectroscopy 3. Load all three (3) orientations into the Graphic Positioning Segments (GSP). a. Left-click and drag each of the 3 imaging planes into the GSP segments. HAL HAL AHR 2024.01.03-22:27:29-STD-Trauma 2024.01.03-22:27:29 STD-Trauma 2024.01.03-22:27:29-STD-1.3.12.2.1107.5.2.50 2024.01.03-22:27:29-STD-1.3.12.2.1107.5.2.50 2024.01.03-22:27:29. 2024.01.03-22:27:29- -03-22:27:29-STD-1.3.12.2.1107.6.2.50 10 IMA 1 FRM 171/267 B IMA 1 FRM 48/1 92 15 IMA 1 FRM 1 5/27 RPH 1cm ARF RPH 1cm 1cm A TP O TP 0 SP P31.5 SP R44.4 SP H7.6 FoV 173-240 FoV 240 240 FoV 220*220 MAGNINAVAILIRESA or-Sap(-3)> Tra(-1) WWWJURESAMPIMUNORKIND Sag-Cor(3)>Tra(1) 11_mprage_sag_p2_75 AMURESAMPIXUNORD Tra>Cor(-10]>Sag(-1) 685 825 W 1269 355 11_mprage_sag_p2_Iso_ND 388 2_qtse_tra_p3_ND Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 2/28/2024 | QR700005768 Healthineers 4. To Position the CSI multi-voxel. a. When the sequence is opened and ready to be planned, a Spectroscopy Guidance window is displayed. The step-by-step guidance is used throughout the planning process. b. Hint: make sure the voxel is not placed in Bone, Air, or Fluid. This can be challenging depending on the location of the lesion. Brain Assist LOC . Y HAL AHR 2024.01 03-22: 27 29-STD-1.3.12. 2.1107 5 2.50 2024.01.03-22:27:29-STD.Trauma 03-22:27 29-STD- 1.312 2 1107.5 2 50 2024.01.03-22:27:29-STD-Trauma 2024.01 03-22:27:29-STD-1 3:12.2.1107 5 2 t1_qtse_dark-fluid_sag_p2 17 IMA1 FRM 51/192 1010 IMA 1 FRM 199/267 11 IMA 1 FRM 11/27 3 t2_qtse_tra_p3 t2_qtse_dark-fluid_tra 5 resolve_3scan_trace_tra_p2_1 ... 6 t1_mprage_sag_p2_iso 7 MPR planning 1cm 8 11 fizd tra 9 X csi_se_135 A E 00:00 TP O SP R41.7 SP P47.5 SP F13.2 FoV 240-240 FoV 185-243 FoV 220-220 GO Cancel 92 MAGN/AUNORNUDIS2D 11_mprage_sag_p2_/so >Con(3)>Tra(1) MAGNIDIS2D Cor 02 MAGNUMNORMDIS20 Cor(-10)-Sap(-1) 837 515 378 MPR cor 0.9mm Range 245 12_qtse_tra_p3 @ 7:12 min + Auto @ 10.7x10.7x15.0 mm @ 1.00 Spectroscopy| Parameters Spectroscopy steps can only be planned on non-distortion corrected images. Please do the following steps for examination planning: drag all senes needed for planning into GSP preselect the region of spectroscopy examination by placing a ROI on one series press "Convert images" to generate non-distortion corrected images for planning the converted images will be automatically loaded into the GSP plan your spectroscopy examination For multiple spectroscopy measurements with the same position and voxel size, please use "Copy Parameter" with "Measurement Parameters" or "Everything" after conversion of the images. Convert image 3:20 min | Auto 20x20x20 mm2 1.00 Spectroscopy Parameters Spectroscopy steps can only be planned on non-distortion corrected images. Please do the following steps for examination planning: drag all series needed for planning into GSP preselect the region of spectroscopy examination by placing a ROI on one series press "Convert images" to generate non-distortion corrected images for planning . the converted images will be automatically loaded into the GSP . plan your spectroscopy examination For multiple spectroscopy measurements with the same position and voxel size, please use "Copy Parameter" with "Measurement Parameters" or "Everything" after conversion of the images. Convert images Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 2/28/2024 | QR700005768 Healthineers Page 2 of 6 c. Scroll through the series (e.g., transverse) until the area that needs to be covered (lesion) is visible. • Once the area is selected, left-click to place the cursor on the lesion. (i.e., example below) AHR test 2020.03.26-14:18:54-STD-1.3.12.2.1107.5.99.3 11 IMA1 FRM 14/27 RPH 5cm TP 0 SP H24 FoV 220 220 Tra>Cor(-10)>Sag(-1) MAGN/NORM DIS2D W 1_112d_tra d. The cursor shape changes to a Red Circle on the image. • All three (3) imaging planes now have a RED cursor mark. • The series in the two (2) other image segments scroll to the nearest location of the cursor that was just placed. HAL AHR 2024.01.03-22:27:29-STD-Trauma 2024.01.03-22:27:29-STD-Trauma 2024.01.03-22:27:29-STD-Trauma 2024.01.03-22:27:29-STD-1.3.12.2.1107.5.2.50 2024.01.03-22:27:29-STD-1.3 12 2.1107 5 2 50 2024.01 .03-22:27:29-STD-1 3:12.2.1107 5:250 17 MA1 FRM 51/192 1010 IMA 1 FRM 189/267 11 IMA1 FRM 11/27 ARF RPH 1cm 1 cm 1cm L F TP O TP O SP R41.7 SP P47.5 8P F13.2 FoV 240-240 FoV 185-243 FoV 220-220 P2 MAGNININORNUDIS20 Sag>Con(3)>Tra(1) MAGNIDIS2D Cor Tra>Cor(-10)-Sap(-1) 11_mprage_sag_p2_Iso MPR cor 0.9mm Range 515 02 MAGNUMNORMDIS 20 245 12_qtse_tra_p3 1418 621 NE Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 2/28/2024 | QR700005768 Healthineers Page 3 of 6 e. Below the step-by-step guidance, the Convert images button is now active. • Click on the Convert images button to convert all of the series to Non- Distortion Corrected images needed for planning the voxel 0 3:20 min | Auto @ 20x20x20 mm3 1.00 Spectroscopy Parameters Spectroscopy steps can only be planned on non-distortion corrected images. Please do the following steps for examination planning: drag all series needed for planning into GSP preselect the region of spectroscopy examination by placing a ROI on one series press "Convert images" to generate non-distortion corrected images for planning the converted images will be automatically loaded into the GSP plan your spectroscopy examination For multiple spectroscopy measurements with the same position and voxel size, please use "Copy Parameter" with "Measurement Parameters" or "Everything" after conversion of the images. Convert images 5. The CSI multi-voxel is now shown in the all the selected image segments. a. The Voxel Area is White and is surrounded by 4 saturation bands. • Reposition the saturation bands to cover bone, air, or fluid (i.e., used to clean up the spectrum). b. If Saturation Bands are added, enter a Delta Frequecy of -3.4 in the Geometry Parameter card. HAL HAR AHR test test test 2020.03.26-14:18:54 STD-1.3.12 2.1107.5.99.3 2020.03.26-14:18:54-STD-1.3.12.2.1107.5 2020.03:26-14:18:54-STD-1.3.12.2.1107-5.99.3/ 14 IMA1 FRM 64/192 16 JMA 1.FRM 86/160 17 JM.A.1.FRM 14/27 ARF 5cm SPR30.0 FOV 2401240 SP P15.1 SP.H2 4 FOV 230*230 FoV 220 220 S3- Cor(3) Tra(1) Cor Sag(-3)>Tra(-1) Tra Cor(-10)>Sag(-1) INAVAJURESANPINORMUND 800 357 MAGNINAVA UIRESAMPIND" 854- 388 LINAEURESAMP/NORMIND - 1178. ti_mprage_sag p2 iso ND 11/ mprage sag_p2_iso_MPR_cor ND 11/1128 43 NO C 602 Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 2/28/2024 | QR700005768 Healthineers Page 4 of 6 6. The Field-of-View (FoV) can be adjusted in all 3 imaging planes on the Routine parameter card. (i.e., used to prevent wrap if the Volume-of-Interest (VOI) is not exactly centered. Routine Contrast Resolution Geometry Position LO.O PO.0 H30.0 ... Orientation Transversal ... Rotation 0.00 + deg FOV A >> P 160 = mm FOVR >> L 160 + mm Thickness F >> H 15 mm Vol A>> P 80 - mm Vol R>> L 80 - mm 7. Click Go, to start the CSI multi-voxel spectroscopy sequence. 2 00:00 Go Cancel Delete Skip Stop Pause Continue Copy& Go Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 2/28/2024 | QR700005768 Healthineers Page 5 of 6 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthineers AG 2024 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthineers AG Siemens Medical Solutions USA, Inc. Imaging & Therapy Systems Siemensstraße 3 40 Liberty Blvd. Magnetic Resonance 91301 Forcheim, Germany Malvern, PA 19355-9998 USA Henkestr. 127 Phone: +49 9191 18-0 Telephone: +1-888-826-9702 DE-91052 Erlangen siemens-healthineers.com Customer Care Telephone: Germany +1-800-888-7436 Telephone: +49 9191 18-0 Unrestricted | Published by Siemens Healthineers AG, 2024 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 2/28/2024 | QR700005768 Healthineers Page 6 of 6
- Multi Voxel Spectroscopy
- Numaris X