MR Cardiac Flow - XA31 Job Aid
MR Cardiac Flow - XA31 Job Aid
MR Cardiac Flow Numaris X MR Cardiac Flow – Numaris X The MR Flow Quantification step is used to determine flow and velocity parameters. 1. Patient Browser, load data set into MR Cardiac Analysis Workflow a. Right-Click on Patient Examination > select Assign Workflow > MR Cardiac Analysis b. Double Left-Click on Patient to Open Open Ctrl+O Open with x View as Read-Only with Batch Open X Assign Workflow Favorites Cancel Workflow Suggested Search Patient with DICOM Q/R MR Breast Biopsy * Add to Demo List MR BreVis Remove from Demo List MR Cardiac Analysis * Set Mark state MR Neuro 30 * Set Delete Protection state MR Neurology * Set Archive state MR Oncology * . Remove Correction state More 2. Cardic Reading Step – view Functional, Dynamic, Tissue Characterization and Flow Data using Common Tools 3. Select MR Flow Quantification Step MR Cardiac Analysis Q + MR Cardiac Reading Learn Create Segments Findings Overview Functional Dynamic Tissue Characterization Combined Flow MR Cardiac Function MR Flow Quantification O MR Cardiac Perfusion O Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 11/29/2021 | HOOD05162003238062 Healthineers Job Aid | MR Cardiac Flow – Numaris X Effective Date: 11/29/2021 | HOOD05162003238062 4. Perform Vessel Segmentation > Vessel Segmentation Window a. Select an Image with Flow b. Select ROI1 Icon – Activates ROI Drawing c. Select One-Click Vessel Segment Icon • Click in Vessel (or) select a Drawing Tool and Draw an ROI d. Use Vessel Segmentation Tools to Refine Segmentation e. Scroll through Phases and Check Segmentation f. Select Copy Forward (or) Copy Backward Icon g. Optional: If you want to perform other Flow Evaluations in a Slice Position, Click ROI2 to ROI4 Icons, enables drawing other Contours Vessel Segmentation O O RO11 RO12 ROI3 ROI4 O O Pen Circle Nudge Adjust One Click Segment +0 O- Delete Delete All Copy Copy Backward Forward 5. Vessel Segmentation Results – displayed as Time-Flow Graph and a Result Ticker (e.g., Lower-Left Segment) a. Time-Flow Graph – displays Flow Quantification Results over time for the selected Flow Parameter b. Select a Flow Parameter and Select One or More ROI • Click and Drag Start and End Sliders to a position to adapt the Time Interval Flow eak-Vel Net Flow Area ROH RO12 RO13 RO14 370 323 276 223 vanced MRI 183 Skyra 136 89 -retro bh Ao_TE2 43 -51 .97 REP 1cm 10 20 30 40 50 60 70 80 90 190 SL 6.0 95 190 286 381 476 571 666 762 857 952 ms TT 300 6 SP H90.2 Start End FoV 309 450 TR 24.1 92 192 TE 2.3 Tra-Sag(17)>Cor(3) 5288 Peak-Vel Forw-Volume Rev-Volume Avg-Flow/Min p2 P/RETRO/DIS2D 108 RR 1002 + 20; 16 heartbeats FROM 185.742 73.1 21.594 3.039 Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 11/29/2021 | HOOD05162003238062 Healthineers Page 2 of 5 Job Aid | MR Cardiac Flow – Numaris X Effective Date: 11/29/2021 | HOOD05162003238062 c. Flow Parameters are Displayed • Flow – Product of Mean Velocity and Surface Area • Peak-Velocity – Peak Velocity in the ROI • Velocity – Mean Velocity in the ROI • Net Flow – Difference between Forward and Reverse Flow • Area – Surface of Segmented Vessel Flow Peak-Vel Net Flow Area TRO14 ml/s 370 323 276 223 183 135 89 43 -51 10 20 30 140 50 60 70 100 95 190 236 381 476 571 666 762 857 992 ms Start End Peak-Vel Forw-Volume Rev-Volume Avg-Flow/Min ROM 85.742 73.1 21.594 3.039 6. Evaluation Options Step – optimize Evaluation Results a. VENC Anti-Aliasing Icon – if measured maximum Flow Velocity is lower than the Ac- tual Flow Velocity, there will be signal void areas in the vessel due to aliasing b. Select VENC Anti-Aliasing Icon – images are updated • Aliased pixel values are corrected, and results are updated c. Adjust Background Phase – Move Threshold slides to adapt the stationary tissue d. Select Auto Button – restores default stationary tissue mask (e.g., select only if neces- sary) e. Select Exclude Regions Icon – draw a contour around regions that shouldn’t be in- cluded into the stationary tissue mask f. Delete Exclusion Regions Icon – select if you want to delete and redraw the exclusion region (e.g., select only if necessary) g. Select Apply – corrects the pixels for the phase contrast images Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 11/29/2021 | HOOD05162003238062 Healthineers Page 3 of 5 Job Aid | MR Cardiac Flow – Numaris X Effective Date: 11/29/2021 | HOOD05162003238062 7. MR Flow Quantification Workflow Step > select Create Findings Icon a. Findings Details dialog box open – Choose Heart Rate of Patient • Default average Heart Rate is displayed • Heart Rate influences the average flow per minute MR Cardiac Analysis + MR Cardiac Reading MR Cardiac Function MR Flow Quantification Create Findings Vessel Segmentation b. Evaluation Option Area is selected – once Checkbox “After Background Phase Ad- justment” is performed c. Invert Polarity Icon – select to Invert the Velocity and Flow Values d. Create Finding – select to Create a Finding for selected ROI • Repeat Step for each ROI Created Create Findings X Patient Information Evaluation Options Patient Name Lumina_Heart Background Phase Adjustment Age 78 Years Height 70.08 inches Start- 0 ms 00.60 lb End: 859.4 ms Heart Rate 67 beats/min Venc: 140 cm/s Analysis ROI 1 ROI 2 ROI 3 ROI 4 Region Flow_ROI_1_R23.4 Create Finding Flow Velocity and Pressure Avg Flow Over Range -71.265 ml/s Average Velocity -9.649 cm/s Avg Flow Per Minute 4.103 l/min Peak Velocity -130.498 cm/s Peak Flow -344.612 ml/s Time to Peak Velocity 148.167 ms Time to Peak Flow 148.167 ms Peak Average Velocity -46.238 cm/s Forward Volume 3.192 ml Time to Peak Average Velocity 148.167 ms Reverse Volume 64.434 ml Pressure Gradient (4Vmax2) 6.812 mmHg Net Forward Volume -61.243 ml Area Regurgitant Fraction - % Average Area 7.19 cm2 Body Surface Area 2.001 m2 Minimal Area 6.297 cm2 Avg Flow per Minute/BSA -2.05 l/min/m2 Maximal Area 7.605 cm2 Net Forward Volume/BSA -30.603 ml/m2 Export Results Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 11/29/2021 | HOOD05162003238062 Healthineers Page 4 of 5 Job Aid | MR Cardiac Flow – Numaris X Effective Date: 11/29/2021 | HOOD05162003238062 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2020 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com Unrestricted 0 Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 11/29/2021 | HOOD05162003238062 Healthineers Page 5 of 5
- MR
- Basic
- syngo
- Dot
- Numaris X