MAMMOMAT B.brilliant - Breast biopsy - Calibration (USA)

This video is a step by step guide on how to perform a calibration of the biopsy unit of the MAMMOMAT B.brilliant.

Target group: All users
Audio: Yes
Recommended to be viewed on the following devices: Laptop, desktop computer (sufficiently large display required)

The biopsy unit should be calibrated. Once a year a reminder message will appear on your monitor. To start the biopsy calibration, select Biopsy Calibration from the Patient drop down menu on the examination task card. The biopsy table should be mounted on the object table and plugged into the swivel arm for the calibration. 3 stereo pairs must be acquired with the calibration needle. Please note that biopsy calibration is performed without a compression plate. Make sure that no paddle and needle holder is mounted. Press the button Move to position to get the system in the start position. When the system movement is done, the message insert the calibration needle and acquire the scan appears on the acquisition workstation. Insert the needle holder and the calibration needle at the system, then acquire the scan. When the scan is acquired, the message Please adjust the reference mark with the reference point button appears. The reference mark is shown on the image as a reference. Cross click the reference point button, then right click on the image. A drop down menu appears. Select Zoom from point. Then click on the reference crosses on both images to zoom in. Press Escape to disable the zoom function before making adjustments. By clicking and holding the left mouse button, adjust the lines. Place them as accurately as possible on the reference point on both images. To complete. The step. Click confirm, right Click to open the drop down again. Select Home Zoom Pan. Place the target mark to the peak of the calibration needle on the image. Right click and select Zoom from. Target mark from the. Drop down menu. Click on the image to zoom in and fine tune the reference point to the peek. Press Escape to disable. The zoom from target mark function before making fine adjustments to the target mark. Then press confirm, check the appearing message and remove the needle holder from the system. Press. The Move to Position button. The second stereo. Pair is. Activated, Insert the needle holder and needle again, then acquire the scan. Place the target mark again in. The second scan right click select. Zoom from target mark and zoom into the images to adjust the reference crosses in both images. Press Escape to disable. The zoom from target mark function before making fine adjustments to the target mark. For adjustment, keep the mouse pressed, when done, click confirm, remove the needle holder again and click on move to position. When the position is reached, insert the needle holder and acquire the scan. Place the target mark on the top of the needle. Right. Click and choose. Zoom from target mark in the drop down menu. Press Escape to disable. The zoom from target mark function, before making fine adjustments to the target mark fine adjust the reference. Crosses. Click, confirm. A pop up appears that calibration has been successfully done. Click OK to close the examination. To finish the calibration, remove the needle holder the. System automatically moves back to the home position.

1/1/1960 26 143 14 11 MAMMOMAT B.brilliant Breast biopsy - Calibration Move to position Patient Examination Edit View Images Tools Scroll Options Help Viewing Filming Register ... Browser ... Num . Detector Calibration Biopsy Calibration Import calibration images Schedule. Save Eilming Layout. Film Task Status ... Film Preview ... Expose Film Task Ctrl+P Copy to Film Sheet Num Enter Search Open patient Biopsy Acquisition Calibration cc Procedure: Biopsy Calibration Please ensure that no paddle is mounted and the needle (respectively the needle holder) is removed. Afterwards press the Move to position button. Manual 71.0 mAs Post processing Close patient 62 mm Advanced Edit View Transfer Procedure: Patient Transfer Examination 26 k Edit View Image The needle holder is moving to position. Detector busy. Please wait Please insert the calibration needle and acquire the scan. L-CC+ S-+ mA As SIEMENS 26 KV 0.7 mGy 0 deg 74 N Image processing is in progress. Patient Transfer Examination Edit View Image 26 kV F1 Loading image (projection view. L-CC+ ST) ... Please adjust the reference mark with the Ref. Point button if needed, place the target mark to the peak of the calibration needle on the image and press Confirm. Ref. Point Confirm Transfer Examination Edit V Hide all target marks Show selected and transmitted target marks Show all target marks Image Attributes ... Direct Comparison Invert Select All Graphics Paste Mark Home Window Fit to Segment Zoom From Point Zoom/Pan Home Zoom/Pan Zoom From Target Mark Edit Comment Patient Transfer Esc Decomp Comp 300 Contact Stereo No padde Paddle 30 Exposel Transfer Examination Edit View wy Cancel mage Attributes ... Inver dvanced Tools Scroll Options H No image to display Please place the target mark to the peak of the calibration needle on the image and press Confirm. CC Tools Scroll WH AWS Calibration has been successfully ended. Please remove the needle from the needleholder if it is still inserted and after that press OK to close the calibration. OK Operator: PV sequence: Auto Film. Auto Exp. Examination successfully ended Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Unrestricted | Published by Siemens Healthineers AG | Siemens Healthineers AG, 2024 Siemens Healthineers HQ | Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com Healthineers

  • mammo
  • bbrilliant
  • b brilliant