How to Create a New Energy Window for Symbia Pro.specta™ Systems

This job aid is designed to assist the end user in creating a New Energy Window Set on VA10 Symbia Pro.specta™ Systems.

How to Create a New Energy Window Symbia Pro.specta™ Systems VA10 This job aid is designed to assist the end user in creating a New Energy Window Set on the Symbia Pro.specta™ System. Creating New Energy Window Sets 1. From the top right corner of the display, hover the cursor over the Settings icon a. Select the Clinical Configuration option from the drop-down menu syadmin Configuration Panel Administration Portal Administration Portal [EN] Clinical Configuration Check up MI System Check Daily QA Calibration Resume syngo Remote Assist Screen recorder SaveLog Remote Service Exam Designer Create Screenshot Fast Contact File Browser Gantry Display Load Service Pack Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 SIEMENS Master Template Effective Date: 21 Apr 2023 | HILS 2224 Job Aid Effective Date: 22 Nov 2023 | QR700003040 Healthineers Job Aid | How to Create a New Energy Window Effective Date: 22 Nov 2023 | QR700003040 2. Select the Clinical Configuration tab at the top 3. Select the MI Clinical Configuration tab in the left menu 4. Select the Energy Window Sets tab to display the list of presets E Scan 13 = Recon 1 O Clinical Configuration SIEMENS Healthineers HSCCHARLIEBROWN / syadmin Free (Restricted Remote Access) Diagnose Installation ª Technical Configuration Quality Assurance ª Clinical Configuration MI Clinical Configuration Energy Window Sets Quality Assurance Configuration Isotope Show All Dose Activity Unit Preset Created By Modified Date Energy Window Sets Tc99m_Siemens Siemens 01/01/2018 View MI Series Description Co57_Siemens Siemens 01/01/2018 View Examination Configuration Gd153_Siemens Siemens 01/01/2018 View Data Preprocessing TI201_Siemens Siemens 01/01/2018 View Patient Browser Prior Study Search Lu177_Siemens Siemens 01/01/2018 View Reading 1123_Siemens Siemens 01/01/2018 View DICOM Data Handling 1131_Siemens Siemens 01/01/2018 View Licensing Se75_Siemens Siemens 01/01/2018 View Show All Ba-133 BiCo 5. From the Isotope drop-down menu, select New. A new isotope Cu-67 window will appear Ga-67 Gd-153 Ho-166 1-123 1-125 1-131 In-111 Kr-81m Kr-85 Lu-177 P-32 Ra-223 Rb-81 Re-186 Se-75 Sm-153 Sn-113 Sr-89 TI-201 Xe-127 Xe-133 Y-90 Yb-169 New SIEMENS Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date 21 Apr 2023 | HILS 2224 Healthineers Page 2 of 5 Job Aid | How to Create a New Energy Window Effective Date: 22 Nov 2023 | QR700003040 6. Enter the New Isotope details: a. New Isotope name b. A default DICOM Code will populate, but can be changed c. Enter the Half-life for the Isotope and select the unit of time from the drop-down menu d. The Bremsstrahlung field is a check option for the user to turn on or keep off e. Enter the desired window Width % for the Isotope f. Enter the Center (kev) g. Add Window gives the user the option to add additional photopeaks New Isotope X Isotope Test225 DICOM Code Test225_Code * Half-life 9.00 Hours Bremsstrahlung ID Type Width% Center (keV) 1 Photopeak 15.00 158.000 Add Window Save Cancel 7. Once the information is entered, select Save to add the new preset to the list Please Note: Make sure the Window Width % does not overlap energies for an added Isotope, a message will display indicating the Isotope is invalid. -- SIEMENS Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date 21 Apr 2023 | HILS 2224 Healthineers Page 3 of 5 Job Aid | How to Create a New Energy Window Effective Date: 22 Nov 2023 | QR700003040 8. Once the new Isotope is created and saved, Energy Window Sets the user will have the option to Add Energy Isotope test225 Window Sets Preset Created By Add Energy Window Set 9. The Create Preset window will open, and the user can select Add Window a. This will enable a drop-down menu to choose LowerScatter, UpperScatter, OffPeakLower, OffPeakUpper, and Co57 Marker, if needed Create Preset X Isotope TEST Preset Name New ID Type Parent Width% Offset% Center (keV) 1 Photopeak 15.00 0.00 166 2 LowerScatter 15.00 Delete LowerScatter UpperScatter Add OffPeakLower OffPeakUpper Co57 Marker Save Cancel SIEMENS Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date 21 Apr 2023 | HILS 2224 Healthineers Page 4 of 5 Job Aid | How to Create a New Energy Window Effective Date: 22 Nov 2023 | QR700003040 Please note this learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (here- inafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. © Siemens Healthcare GmbH 2023 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Phone: +49 9131 84-0 siemens-healthineers.com SIEMENS Unrestricted | Published by Siemens Healthcare GmbH © Siemens Healthcare GmbH, 2023 Master Template Effective Date 21 Apr 2023 | HILS 2224 Healthineers Page 5 of 5

  • energy window
  • va10
  • symbia
  • symbia pro.specta
  • pro specta
  • prospecta
  • energy window sets