FlexLab™ X Total Lab Automation Emergency Release of Sample Tubes From the Centrifuge Module Job Aid
This job aid provides instructions on how to remove sample tubes remaining in the centrifuge following a power failure.
FlexLab TM X Total Lab Automation SIEMENS Healthineers Emergency Release of Sample Tubes From the Centrifuge Module Job Aid Emergency Release of Sample Tubes From the Centrifuge Module During a power failure, the centrifuge lid cannot be unlocked by motor. An emergency release must be executed to remove any sample tubes remaining in the centrifuge. 1. Lift up the Centrifuge Module cover 2. Look into the hole of the Centrifuge lid using a lamp. Make sure the Centrifuge is not spinning. Proceed with next steps only if it is possible to see bucket inserts and rotor at standstill. START 3. Insert the hexagonal wrench key into the bore hole and carefully rotate half a turn in * clockwise direction until the lid opens 3 OPEN 4. Open the Centrifuge lid 5. Manually remove sample tubes 6. Close the Centrifuge lid 7. Carefully rotate the hexagonal wrench key half a turn clockwise until the lid locks; then, extract the key 8. Close the cover 9. In the User Interface, select the Centrifuge card Template Effective Date: 10 Aug 2021 | HILS 2223 1 10. Set module Offline and select the Clear Tubes button Effective Date: 28 Mar 2024 | QR700002750 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 SIEMENS Healthineers Please note this learning material is for training purposes only. FlexLab TM X Automation is manufactured by Inpeco and distributed by Siemens Healthcare Diagnostics Inc. FlexLab is a trademark of Inpeco SA. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com 2 Effective Date: 28 Mar 2024 | QR700002750 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024
- automation
- flexlab
- track