FlexLab™ X Total Lab Automation Daily Maintenance on the Centrifuge Module Job Aid

This job aid provides instructions on how to complete daily maintenance on the Centrifuge Module. 

FlexLab TM X Total Lab Automation SIEMENS Healthineers Daily Maintenance on the Centrifuge Module Job Aid 5 Inspect Bucket Inserts on the Centrifuge Module Refer to the FlexLab X Total Lab Automation Operators Manual when performing maintenance procedures. 1. Set the Centrifuge Module Off-line 2. Lift up the Centrifuge Module cover 3. Remove the bucket inserts from the input area and the output area 4. Press the Stop/Open button to open the Centrifuge lid 5. Remove all bucket inserts from the rotor 6. Check all bucket inserts for cracks or corrosion. If any cracks or corrosion are observed, discontinue use of bucket inserts. Ensure the silicone tabs holding 6 sample tubes are not worn or broken 1 7. Return the bucket inserts into each rotor bucket paying attention to the orientation 8. Close the Centrifuge lid 9. Place a bucket insert in each location of the input area and output area, except the “O” and “A” locations 10. Close the cover Template Effective Date: 10 Aug 2021 | HILS 2223 1 11. Set the Centrifuge Module to Online Effective Date: 28 Mar 2024| QR700003068 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 SIEMENS Healthineers Please note this learning material is for training purposes only. FlexLab TM X Automation is manufactured by Inpeco and distributed by Siemens Healthcare Diagnostics Inc. FlexLab is a trademark of Inpeco SA. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com 2 Effective Date: 28 Mar 2024| QR700003068 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024

  • automation
  • flexlab
  • track