FlexLab™ X Total Lab Automation As Needed Maintenance on the Sealer Module Job Aid

This job aid provides instructions on how to perform as needed maintenance on the Sealer Module. 

FlexLab TM X Total Lab Automation SIEMENS Healthineers As Needed Maintenance on the Sealer Module Job Aid Replace the Sealer Coil on the Sealer Module Refer to the FlexLab X Total Lab Automation Operators Manual when performing maintenance procedures. 1. Set the module Off-line 6 2. Select Flush Carriers entry and confirm 3. Prepare the Sealer Module for reel change and release the aluminum stripe from the Foil Locker: A. Select Sealer button in Overview screen B. Select Diagnostics C. Select Reel D. Select Prepare For Reel Change function button 4. Check the heater temperature. Wait until the value corresponding to the Sealer – Heater Analog sensor is 55°C/131°F or less before continuing. It will take approximately 30 minutes to cool down 5. Open the Sealer Module cover 6. With thumbs on the reel axis, grasp the Sealer coil with your fingers. Pull coil toward you while pushing your thumbs against the reel axis and remove the Sealer coil Template Effective Date: 10 Aug 2021 | HILS 2223 1 * This procedure is continued on the next slide. Effective Date: 28 Mar 2024| QR700003068 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 FlexLab TM X Total Lab Automation SIEMENS Healthineers As Needed Maintenance on the Sealer Module Job Aid Replace the Sealer Coil on the Sealer Module, A continued 7. Remove the safety tab from the new sealer coil B 8. Install the new Sealer coil by aligning the peg (A) of the reel support with the cavity (B) in the Sealer coil case. The shiny surface of the aluminum stripe should be toward the top, the dull surface toward the bottom 9. Pull the aluminum stripe and place it as shown in the image. Then, insert the stripe under the Foil Locker (C) ensuring it is positioned on the Sealer coil sensor (D), slide the aluminum D stripe under the Puller (E), and then under the Cutter (F) 10. Close the Sealer cover 11. If changing the foil capacity, set the Sealer Reel Capacity: 9 Overview > select Sealer > select Settings > select Setup > select Number of Foils in a New Reel function button 12. Reset the reel counter and initialize the new coil by cutting some pieces of foil: Overview > select Sealer > select Diagnostics > select Reel > select Init Foil Reel C E F 13. Set the Sealer Module to On-line 2 Effective Date: 28 Mar 2024| QR700003068 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024 SIEMENS Healthineers Please note this learning material is for training purposes only. FlexLab TM X Automation is manufactured by Inpeco and distributed by Siemens Healthcare Diagnostics Inc. FlexLab is a trademark of Inpeco SA. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthineers HQ I Siemens Healthineers AG Siemensstr. 3 91301 Forchheim Germany Phone: +49 9191 18-0 siemens-healthineers.com 3 Effective Date: 28 Mar 2024| QR700003068 Unrestricted I Published by Siemens Healthineers AG | © Siemens Healthineers AG, 2024

  • automation
  • flexlab
  • track