Using Dual Energy SPP Slider Bar in Postprocessing
This job aid discusses how to use the slider bar to select an optimum image on a Dual Energy scan.
Dual Energy Reconstructions: Using the Slider Bar to select desired energy SOMATOM go.Sim, go.Open Pro; VA30/VA40 There are several methods for selecting the desired energy for dual energy reconstructions. One of these methods is described below: 1. Log in to the system as a SuperUser. 2. Select Exam Designer from the Settings (cog wheel) menu. 3. Select the Factory Default protocols under the RTP folder for either the Abdomen or Head & Neck, as the process is the same. 4. Click the Modify Protocols button on the bottom right of the screen. 5. Make the necessary adjustments to the scan and reconstruction parameters as needed. Note: In the VA30 software version, the default Head & Neck protocol has two reconstruc- tions already available, and the default Abdomen protocol does not. 6. If needed, right mouse click on the Abdomen reconstruction and select Repeat. 7. Rename the reconstructions for better clarity. 8. Adjust one of the reconstructions for 120kV. The below can be the base Dual Energy sec- tion of the final protocol. Scan/Recon Scan Favorites Scan/Recon Scan Mode KV KV (L/H) Quality ref. mAs Quality ref, MAS Quality ref. mAS (H) m CTDIvol [mGy] [mGy"cm] Acquisition [mm] Pitch Kot Time RTP Head and Neck DE [factory] (Adult) Topogram Scanplanning TopoAdultSn Sn 100 80 SPR (32 cm) 0.01 6 × 0.6 Head and Neck for dose calculation RoutineSpiralAdultNeck 156 (32 cm) 10:43 335.0 64 × 0.6 210 (32 cm) 6:25 206 0.00 Head and Neck for contouring MultienergySpiralAdultNeckTwinspiral (max 100/Sn140 149 64 × 0.6 0.80 1.00 0.50 RTP Abdomen DE [factory] (Adult) Topogram Scanplanning TopoAdultSn Sn 100 80 SPR (32 cm) 0.01| 0.5 6 x 0.6 Abdomen MultienergySpiralAdultAbdomen Twinspiral ( ... 00/Sn140 146 163 (32 cm) 4.38 1445 64 × 0.6 0.80 0.35 Scan/Recon Scan Favorites Scan/Recon Scan Mode KV kV (L/H) Quality ref. mAs Quality ref. mas Quality ref. mAs mA CTDIvol [mGy] DLP [mGy*cm] Acquisition [mm] Pitch Kot Time RTP Head and Neck DE [factory] (Adult) Scanplanning TopoAdultSn Sn 100 80 SPR (32 cm) 0.01 0.5 6x0.6 Head and Neck for dose calculation RoutineSpiralAdultNeck 156 (32 cm) 10:43 335.0 64 × 0.6 0.80 1.00 Head and Neck for contouring MultienergySpiralAdultNeckTwinspiral (max. 100/Sn 140 149 210 (32 cm) 6,25 206.5 64 × 0.6 0.80 0.50 RTP Abdomen DE [factory] (Adult) Topogram Scanplanning TopoAdultSn Sn 100 80 SPR (32 cm) 0.01 05 6 x 0.6 Abdomen for dose calculation RoutineSpiralAdultAbdomen 120 143 (32 cm) 6.76 217 1 64 x 0.6 0.80 0.50 Abdomen for contouring MultienergySpiralAdultAbdomen Twinspiral ( .. 100/Sn140 146 163 (32 cm) 4 38 144 5 64 × 0,6 0 80 0. 35 SIEMENS Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2022 Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Healthineers Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 9. Select the Recon&GO tab. CHI Protocol Scan Favorites General Scan Dose Timing Config Physio Scan Contrast Recon Favorites General Recon Recon&GO Image Impression Recon Box Physio Recon Auto Tasking Inline Options Find/Replace Scan/Recon Recon&GO × Scan/Recon Spectral Recon Spectral Recon keV Inline Results Multi Recons Fix Axial RTP Head and Neck DE [factory] (Adult) Topogram Head and Neck for dose calculation Head and Neck for dose calculation 1.50 None None On Head and Neck for contouring Head and Neck for contouring 1.50 Qr40 Monoenergetic Plus 40 None None On None Monoenergetic Plus L3D H3D 10. On the reconstruction for contouring, hover over the drop-down arrow on Spectral Recon and select None. 11. On the Inline Results tab, hover over the drop-down arrow next to Multi Energy and select Spectral Postprocessing (SPP). CH Protocol Scan Favorites General Scan Dose Timing Config Physio Scan Contrast Recon Favorites General Recon Recon&GO Image Impression Recon Box Physio Recon Auto Tasking Inline Options Find/Replace Scan/Recon Recon&GO X Scan/Recon Spectral Recon Inline Results Multi Recons Fix Axial RTP Head and Neck DE [factory] (Adult) Topogram Head and Neck for dose calculation Head and Neck for dose calculation 1.50 None None On Head and Neck for contouring Head and Neck for contouring 1.50 Qr40 None None None On None Parallel Ranges Radial Ranges Applications Multi Energy DE Parallel Ranges DE Radial Ranges SPP L&H RT Automatic Contouring Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 2 of 11 12. Select the Inline Options Tab. The default monoenergetic KeV is 70. You can change to a different starting energy if desired. Back to Exam Designer Scan Protocols 1. Select Protocols 2. Modify Protocols 3. Confirm Changes Protocol Scan Favorites General Scan Dose Timing Config Physio Scar Contrast Recon Favorites/ General Recon Recon&GO Image Impression Recon Box Physio Recon Auto Tasking Inline Options Find/Replace Scan/Recon Recon&GO x Inline Options x Scan/Recon Spectral Recon Inline Results Multi Recons Results Applicatio Energy n Class In kev RIP Abdomen DE Demonstration (Adult) Topogram Abdomen Abdomen 1.50 Qr40 A3 13. Rename the protocol per facility specifications. 14. Click on Confirm Changes and click on Save to save the protocol. 15. You are now ready to scan. If a contouring and a 120kV reconstruction were not created, at this point you can load a routine 120kV protocol and add the base Dual Energy. Protocol Selection Protocol Parameters Browse Protocols Physiological Signal Requested Procedure ECG Demo Patient Type all Protocol: RTP Head and Neck DE Demonstration Contrast Protocol all -Indications: radiotherapy planning of the head, Topogram syngo.via all neck and shoulder region dose calculation and in addition monoenergetic plus images for Head and Neck for dose calculation contouring wo CM -Scan modes: RoutineSpiralAdultNeck, Head and Neck for contouring MultienergySpiralAdultNeckTwinspiral (max. 100 Adult Child cm) Protocol: RTP Head and Neck [factory] -Indications: radiotherapy planning of the head, neck , shoulder region, w/wo CM Topogran -Scan modes: RoutineSpiralAdultNeck Head and Neck Trauma Protocol: RTP Head and Neck DE [factory] -Indications: radiotherapy planning of the head, Topogram RTP neck and shoulder region dose calculation and in addition monoenergetic plus images for Head and Neck for dose calculation contouring wo CM Head and Neck for contouring Intervention -Scan modes: RoutineSpiralAdultNeck, MultienergySpiralAdultNeckTwinspiral (max. 100 Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 3 of 11 16. On the Auto Tasking Tab, you can set the reconstruction to automatically load in View&GO or you can manually load the reconstruction from the Patient Browser into View&GO. KV 100 Ef. mAs 133 KY 8n100 Quality ref. mas 148 ma 80 A T1 0.35 TI: 32 ................................................. ....................................... GT 0.0 SL 4.064x0.6/p0.8 512.0 0/0 350 Simulator Ester Sn 50 800.0 0/D Simulator Recon Favorites | General Recon Recon&GO Image Impression Recon Box Physio Recon Auto Tasking Inline Options Scan/Recor Auto Recon Recon Ref. Requested Procedure Lines Body Part Nodes Auto View&GO View&GO Dataroles Auto Local Printing Applicati ... syngo.via Examined Syngo.via Datarole Abdomen 1.50 Qr40 A3 On ABDOMEN On None Off None None Auto L 15.15 17. On the SPP reconstruction, you will see Start SPP. Press Start SPP. In the VA40 software version, the SPP is done during the reconstruction process so this step will be skipped and show you the KeV energy below. When finished, the Start SPP will change to Mono+ and the selected KeV energy (70 as shown in this case). DE Test, 1 SOMATOM 90. OPERADA play *01/01/1981, 61Y CURRENT 09/19/2022 Abdomen 1.50 Qr40 A3 SPP_ME70 [2] 2 IMA~100 5:39: 18.68 PM SP F 1287.3 E KV 100/5n 140 ME 70ke / Fiter Sn 18. Scroll through image slices to find an appropriate area to evaluate. Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 4 of 11 19. Right mouse click on the KeV number and a Slider Bar pop-up box will appear. SIEMENS Healthineers 19 o 12 º21/1972 (30) play CT VIENSGO CT View&GO CT View&GO Seack 342 Stack Aus Follow Up 2TP DE ADO. play SOMATOM go. Open Pro 09/21/1972, 50Y VA40A MPR VET NPEMP MPK Comrpar CURRENT 09/21/2022 Abdomen 1.50 Qr40 A3 SPP_ME70 [2] 2 IMA~1 1:42:37.02 PM SP F1188.0 Favorite Tools C MPR SL Parallel O Distance ROX Cicle Mono+ Farges kev 70 0cm Snapthet SimAGO Export F KV 100/Sn140 400 ME 7CkeV / Filter Sn 20. With the cursor on the slider bar, move to the desired KeV for best viewing. In this example, 43KeV was chosen. Dual Energy Mono+ Properties ? x Cancell Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 5 of 11 Create Axial Series at desired KeV There are 2 ways to create the series. Start by leaving the slider bar at the desired energy level. 1. Verify the energy level by checking the footer. DE Test, lest SOMATOM go Open Pro play *01/01/1961, 61Y CURRENT 09/19/2022 Abdomen 1:50 Or40 A3 SPP_ME70 [2] MPR kV 100/$n 140 2. Navigate to the upper left corner menu and select Save As Series. This will give you the chance to name the series as desired. Tip: include the selected KeV in the name so as not to confuse it with the 120kV series. Wait until the reconstruction is complete and listed in the Series window before closing. Zoom/Pan DE Abd, test play Windowing *09/21/1972, 50Y Save as Series Rotate 3 28:44-DST-1 Series Name DE Adbomen 43Kel Image Comment Scroll O, 162Y Accept Cancel Bu Movie * 3D Reference Pointour MPR Ranges 84154 SL min Mono+ O keV 43 Save as Series Region Growing Fit to Segment 3. Select Save and Pause or Save and Send (if configured) to save the series. Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 6 of 11 The second way allows you to set a new recon volume and a different slice thickness than the one set at the original scan time. 1. Go to the upper left corner menu and select Ranges>Parallel Ranges. A new view will open. Zoom/Pan -O- Windowing Rotate 6:28:44-DST-Trauma 18-28:44-DST-1.3.12.2.1107.5.1.7 Scroll O. 162Y Scroll Il Movie * 3D Reference Pointouring 1.50 Qr40 A3 SPP_ME70 [4] Ranges 484f54 Parallel Ranges " Save as Series * Radial Sliced Rang > Open Parallel Ranges Region Growing {O)Radial Ranges Fit to Segment Curved Ranges Spine Ranges * Vascular Ranges 2. In the bottom right corner box, you can set the desired Image Thickness DE Test, test BOMATOM go Open Pro DE Test test SOMATOM go Open Pro DE Test, test SOMATOM go OFF Fre "01/01/1961, 61Y "01/01/1 951. 61Y CURRENT CURRENT CURRENT 09/19:2022 03:19/2022 09/15/ 2022 Abdomen T30 OUACAD, WIE 70151 Abdomen 1.50 Or40 A3 SPP_METO [ Abdomen 1.50 ORQUAJEPPE70 20MA-101 2 IMAna FRuna SPA134 6 SP F1202 5 ........ R 10cm MPR KV-100/9014 Sice Triciress 1 3 mm KV 100/56140 Curved Spine Vascular DE Test, lest Preview "01.01/1961, 61Y Preset Select the Preset STUDY ON/1 0/2022 Image Thickness Distance between Images 1.2 mm Number of images 256 Field of View 600.0 600.0 mm Presets Orientation Image Series start AccepE Close Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 7 of 11 3. In the upper left box, set the desired begin and end positions. H DE Test, test play SOMATOM go. Open Pro *01/01/1961, 61Y VA40A CURRENT 09/19/2022 Abdomen-1-50-Qr40-A3 SPP-ME70-[2] 1,2 2 IMAn.a. FRM-na 5:39:18.68 PM SP A134:6 R MPR SL min- 178 10cm Mono+ keV 70 A Distance between images: 1.2 mm Slice Thickness: 1.2 mm KV 100/Sn140 400 ME 70keV / Filter Sn Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 8 of 11 4. On the bottom right select Start and you will then see a Preview of the series in the lower left box. CURRENT CURRENT CURRENT 14 162822 16:2122 oral Prova Seres 5. Scroll through the Preview series. If acceptable, on the right lower box, you can rename the series, such as site-energy-slice thickness, and then select Accept. Parallel O Radial Curved Spine * Vascula Preset Select the Preset Image Thickness 2.0 mm Distance between Images 2.0 mm Number of Images 151 + Field of View 600.0 x 600.0 mm Presets Orientation Image Series Range Series Name | Addomen DE 56KeV Image Comment Include Reference Image Reject Accept Close Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 9 of 11 6. The series will now be displayed in the list. SIEMENS ealthineers CT View&GO Series DE Abd, test play SOMATOM go. Open Pro 09/21/1972, 50Y VA4DA Current CURRENT 09/21/2022 CT, 9/21/2022 Abdomen 1.50 Qr40 A3 SPP_ME70 [2] RTP Abdamen DE In 1:42:37.02 PM Favorite Tools min Monot 10cm Snapshot 97400 Exper F KV 100/Sn140 A010+ 43kBV 7. Save and Pause or Save and Send (if configured) to save the series. Note: As long as the original Dual Energy scan is available, the data can be reloaded into View&GO and the steps can be repeated to create different KeVs. Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 10 of 11 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product-related claims, or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission, or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters, and configuration-dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthcare GmbH Siemens Medical Solutions USA, Inc. Siemens Shanghai Healthcare Sector 40 Liberty Blvd. Medical Equipment Ltd. Henkestr. 127 Malvern, PA 19355-9998 USA 278 Zhou Shu Road 91052 Erlangen Germany Telephone: +1-888-826-9702 Shanghai 201318 Phone: +49 9131 84-0 Customer Care Telephone: P.R. China https://www.siemens-healthineers.com +1-800-888-7436 Telephone: +1 847 304 7700 Unrestricted | Published by Siemens Healthcare GmbH “ Siemens Healthcare GmbH, 2020 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 10/12/2022 | HOOD05162003304487 Healthineers Page 11 of 11
- post processing
- SPP
- spectral
- reconstruction