Direct Stopping Power Ratio configuration to syngo.via Rapid Results
This job aid shows the steps for creating a Direct Stopping Power Ratio calculation series as a Rapid Results in syngo.via, from a dual energy CT scan. The calculation is then sent to the radiation therapy treatment planning system for use.
Direct Stopping Power Ratio configuration to syngo.via Rapid Results job aid syngo.via VB40/VB60/VB80/VC10; SOMATOM Force, Drive, Definition Flash, Confidence, Definition AS Open, SOMATOM go., SOMATOM X., and NAEOTOM Alpha class CT scanners Note: This is not available in DE Workflow but only as Rapid Results. It is only selectable as a data role at Stopping Power Ratio (AA) in Scan Protocol Assistant (SPA) and Exam Designer for Dual Source, Twin Spiral, and Dual Spiral scanners before VB10. No Twin Beam Dual Energy can be used for Direct SPR. The Stopping Power Ratio result will be visible when exported to a treatment planning system (TPS), but it is not viewable in syngo.via. Part 1: Configuration on syngo.via 1. Log in to syngo.via as a clinical administrator and navigate to the configuration panel. # meduser Configuration Panel Server: syngo Primary role: Reading Physician Administration Portal Additional roles: Clinical Administrator, Flight Recorder Dual Energy Administrator, IT- Administrator Client-Server Connection Test Open Print Open with DICOM Import Export Correct Delete Q/R Open Print Open with DICOM Import Export Q/R Correct Delete Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers 2. Open the Configuration Panel, and under the Advanced Visualization tab, navigate to CT Dual Energy Configuration. a. The needed value can be set in the Application Classes tabcard for the RhoZ appli- cation. b. Under Dual Energy Spectra, select the scanner to be configured to syngo.via. c. Select the Automatic Parallel Ranges tab. SIEMENS .. Healthineers R Configuration Panel Client Settings Advanced Visualization Display And Behavior Export OpenApps CT Dual Energy Configuration Application Profiles Application Classes Application Class Rho/Z Base Application Class Rho/Z Create New Delete Application Subclasses Rho/Z Dual Energy Spectra Drive/Flash/DS ... 80/140kV Visualization Algorithm Material Definitions Presets Preprocessing ROI Properties Automatic Parameters Options Parallel Ranges Automatic Archiving of Thin Slice (Original) Parallel Range 1 Parallel Range 2 Display Type MPR Thick Display Type MPR Thick Orientation Axial Orientation Axial Slice Thickness [mm] 3.04 Slice Thickness [mm] 1.54 Increment [mm] 3.04 Increment [mm] 1.0 Rho Rho V Fused Fused OK Apply Cancel Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 2 of 11 3. In the boxes for Parallel Range 1 (and 2 if desired), select the parameters and desired out- put series. a. At the bottom, you will also note that the RRT selection option for RhoZ also applies to SPR. The default of this value will be Dual Energy only and should be changed to in All Workflows to include RTIS. b. Select Apply in the lower left corner. c. Select OK when finished. SIEMENS ... Healthineers Configuration Panel Client Settings Advanced Visualization Display And Behavior Export OpenApps Dual Energy Spectra Drive/Flash/DS ... 80/140kV Visualization Algorithm Material Definitions Presets Preprocessing ROI Properties Automatic Parameters Options Parallel Ranges Automatic Archiving of Thin Slice (Original) Parallel Range 1 Parallel Range 2 Display type MPR Thick Display Type MPR Thick Orientata Axial Orientation Axia Slice Thickness [mm] 3.0 ¢ Slice Thickness [mm] 1.5 ¢ Increment [mm] 3.0¢ Increment [mm] 1.04 Rho Rho ¥ Fused Fused High Resolution High Resolution Include Reference Image Include Reference Image Rapid Results Techonology will be available in All Workflows The same parameters will also be used for the Reset to Default CT Liver Analysis MM Oncology Configuration XA Interventional Reading RT Automatic Contouring Configuration OK Apply Cancel Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 3 of 11 Part 2: Configuration on the CT scanner – SOMATOM Drive/Definition Flash 1. Start the Scan Protocol Assistant by navigating to Options>Configuration>Scan Protocol Assistant. Patient Applications Edit Insert View Setup Image Options system Help License ... File Browser ... Switch to External Simulator Hide Simulation Taskcard Maintenance ... P. S Configuration ... X File Notes Patient Patient Patient Transfer Browser Registratic CT.Log Expert-i ImageTex Scan Protocol E-Logbook Editor Assistant Configuration Ready NUN 2. Select syngo.via under Scan protocols. Select the at the bottom to proceed to the next step. Note: If the syngo.via bullet is not in the menu of options, and all appropriate licenses are present, contact local service to configure. Scan Protocol Assistant 1 2 3 4 5 Scan protocols . Manipulate scan protocols (cut / copy / paste / delete) . Change parameters . Import user scan protocols . moort from teamplay . Export scan protocols Restore scan protocols to Siemens default Contrast protocols Define contrast protocols · List/ import / export contrast protocols Update customer protocols Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 4 of 11 3. In the Scan Protocol Assistant, select the applicable RT scan protocol(s), and at the bottom to proceed to the next step. Scan Protocol Assistant X 2 3 4 5 Connect to syngo.via o Adult · Child Select all protocols Axial recon jobs Protocols with ECG Interventional protocols Customized scan protocols Select invalid protocols 3D recon jobs Protocols with Dual Source Protocols with Dual Energy Siemens default protocols ADMIRE recon jobs Respiratory protocols Protocols for workflow Protocols for data role Head Spine C_Spine_IR_Vol Vascular Dual Energy Head XCARE HeadAngio_X ... DE Head_Brai Head IR C_Spine_IR HeadAngio_IR DE Head Brai Head_XCARE C_Spine_Vol HeadAngio_X .. TBDE_Thorax_ ... HeadRoutine C_Spine_Sn HeadAngioRou. TBDE_Thorax_ HeadNeuro_X. C_Spine HeadAngio_Vol TBDE Thorax HeadNeuro Spine_IR_Vol NeuroDSACT ..... Neck Thorax Specials PET Neck IR Thorax_XCAR. Trauma NeckRoutine Thorax_IR TraumaThin NeckThorax Thorax XCARE PolyTrauma ThoraxRoutine HeadTrauma Shoulder ThoraxHR_IR HeadTraumaS ShoulderRoutine Shoulder_Sn Abdomen RT Cardiac Abdomen_IR RT_Head CaScore AbdomenRoutine RT Head IR CaScoreSeq Upper Extremities AbdMultiPhase FAST_Cardio WristZUHR AbdomenSeq RT DE Head CoroCTA_Ada ExtrZUHR_IR Colonography CoroCTA_Ada ExtrRoutineZUHR Colonography RT_HeadNeck. CoronaryCTA_IR ....... ...... Lower Extremities Pelvis KneeUHR Pelvis_IR FootUHR PelvisRoutine ExtrZUHR_IR Hip ExtrRoutineZUHR Hip_IMAR ..... ExtrUHR ...... Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 5 of 11 4. In the bottom of the Change parameters workflow tab, select Workflow: MM RT Image Suite and Data role: StoppingPowerRatio (AA). a Scan Protocol Assistant [X] 2 3 5 Change parameters Protocol name Requested Procedure Workflow Recor Dues role DE_Thorax_Analysis Dual Energy / DE_Thorax_Analysis (Adulo) DE_Thorax 5.0 831f DE_Thorax 5.0 B80f DE IPP DE_Thorax 20 D331 Select al Scopirice scena _Spiral scans I Musscons IL Adycove CD spiny econ Axial recon jobs I ADMIRE I .Spawnce M ISpra VM RT Image Suite Workflow Data role RICVIZ(AA) SPP StoppingPowerRatio AA . Aspect Score AA . Lung CAD AA . Lung CAD COOV AA - Lung CAD COOY Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 6 of 11 5. Select Yes to save the changes and the lower right red button to exit Scan Protocol Assistant. Scan Protocol Assistant x 2 3 4 5 ..... ..... Confirmation Scan protocols with changes: Apply Old Value New Value RT DE Head Sn TaskflowldMeaningProt 1 MM RT Image Suite Do you want to save these changes? Yes Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 7 of 11 Part 3: Configuration on CT scanner – SOMATOM go. and SOMATOM X. 1. Under the Settings cog, navigate to Exam Designer>Scan Protocols. SIEMENS ... ? Healthineers O. Bam Designer Cor nel Exam Designer Administration Portal Administration Portal [EN] Scan Protocols Clinical Configuration Checkup Daily QA Calibration Resume syngo Remote Assist Screen recorder SaveLog Remote Service Exam Designer Create Screenshot Fast Contact File Browser Load Service Pack Direct Laser Check myNeedle Laser Check What is New Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 8 of 11 2. Under the Select Protocols tab, choose the protocol(s) to be modified and select Modify Protocols. SIEMENS Healthineers O bom Designer Scan Protocols 1. Select Protocols 2. Modify Protocols 3. Confirm Changes Search Protocols: placol name of descripe Q Filter by Patient Type all Contrast Protocol a Show: # Factory Protocols (142) Customer Protocols (2) Decision Tree all Technique all . Adult Child Upper Extremities (0) Vascular Extremities (0) Il Head (0) Il Upper Extremities (0) + Add folder + Add folder Lower Extremities (0) Trauma (0) Neck (0) Lower Extremities (0) Trauma Trauma Vascular Head (0) · RTP (2) # Shoulder (0) Vascular Head (0) Vascular Neck (0) RTP Head and Neck DE Thorax (0) Vascular Neck (0) seats Specials Vascular Thorax (0) Specials (0) Abdomen (0) Vascular Thorax (0) Boby Vascular Cardiac (0) Intervention (0) I Pelvis (0) Vascular Cardiac (0) Vascular Abdomen (0) Spine (O Vascular Abdomen (0) Reset Selection Hide Delete Export Import Modify Protocols Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 9 of 11 3. In the Modify Protocols Recon Favorites tab, select the following: a. Inline Results> L&H (low and high series) b. syngo.via Workflow>MM RT Image Suite c. syngo.via Datarole>Stopping Power Ratio d. Transfer Nodes (Auto Tasking)>syngo.via e. Auto Transfer>On SIEMENS .. Healthineers 0 O Scan Recon O. Exam Designer Back to Exam Designer Scan Protocols 1. Select Protocols 2. Modify Protocols 3. Confirm Changes Protocol Scan Favorites General Scan Dose Timing Config Recon Favorites General Recon Recon&GO Image Impression Recon Box Physio Recon Find/Replace Scan/Recon Recon Favorites X Scan/Recon Inline Results syngo.via Workflow syngo.via Datarole Transfer Nodes Auto Transfer RIP Head and Neck DE (Adult) Head and Neck for dose calculation Head and Neck for dose calculation 1.50 Qr4 ... None None Auto Head and Neck for contouring Head and Neck for contouring 1.50 Qr40 A3 L&H MM RT Image Suite StoppingPowerRatio syngo.via On 4. When finished, select Confirm Changes (lower right). Confirm Changes 5. On the next page, select Save Protocols. Save Protocols Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 10 of 11 Please note that the learning material is for training purposes only! For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Copyright © Siemens Healthineers AG 2025 Siemens Healthineers Headquarters Local Contact Information Legal Manufacturer Siemens Healthineers AG Siemens Medical Solutions USA, Inc. Siemens Shanghai Siemensstraße 3 40 Liberty Blvd. Medical Equipment Ltd. 91301 Forcheim, Germany Malvern, PA 19355-9998 USA 278 Zhou Shu Road Phone: +49 9191 18-0 Telephone: +1-888-826-9702 Shanghai 201318 siemens-healthineers.com Customer Care Telephone: P.R. China +1-800-888-7436 Telephone: +1 847 304 7700 Unrestricted | Published by Siemens Healthineers AG, 2025 SIEMENS Master Template Effective Date: 3/5/2020 | HOOD05162003074494 Job Aid Effective Date: 05/01/2025 | QR700022010 Healthineers Page 11 of 11
- rt
- via