Dimension® Family of Integrated Chemistry Systems – Weekly Maintenance: Cleaning the HM Wash Probes and the R2 Reagent Probe Video
This video provides instructions for performing weekly maintenance on the Dimension® Family of Integrated Chemistry Systems, including cleaning the HM was probes and the R2 reagent probe.
Dimension family of integrated chemistry systems, weekly maintenance, cleaning the HM Wash probes and the R2 reagent probe video. The HMM wash robes and our two reagent probes should be cleaned weekly to remove any residue. Cleaning the HM Wash robes. To prevent any automatic priming of the HM Wash Robes, navigate to the, Umm, prime screen. With the instrument in standby, select F4, system re. Def 7 pump prime. F6 HM. Raise the IMT and reagent lads. Use a moist cotton swab dipped in water and beginning at the top of the probe, wipe down the outside of both HM Wash probes. Cleaning the are two reagent pro. Turn the gold bronze colored spline shaft on the R2 reagent arm until the R2 probe rises out of the R2 reagent drain. Move the arm counterclockwise until the are two probe can be accessed. Use a cotton swab dipped in 0.1 normal sodium hydroxide and scrubbed the nut at the top of the probe. Beginning at the top, wipe down the outside of the R2 reagent pro. Once the cleaning is complete, select F1 HM Wash pump to Prime the HM Wash pumps. If documenting using the air logs from the operators guide, document the maintenance and the weekly Slash monthly maintenance log sheet or instrument log sheet. Or, if documenting using the electronic logs in the dimension software, document the maintenance in the instrument log screen.
11 Dimension® Family of Integrated Chemistry Systems- Weekly Maintenance: Cleaning the HM Wash Probes and the R2 Reagent Probe Video SIEMENS . Standby Sampler Idle IMT OK 09/14/22 21:23 Dimension Exit Enter labmanager en OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert Supplies At any time, press HELP to get help, or press EXIT to exit a program. QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM CHECK F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS Healthineers 01/12/23 07:43 SYSTEM PREPARATION MENU F1: INVENTORY F2: REAGENT PREP F3: IMT F5: TEMP MONITOR F6: SYS COUNTERS F7: PUMP PRIME F8: DAILY MAINT. Diagnostics 01/18/23 11:59 PUMP PRIME CYCLES: 3 SAMPLER REAGENT 1 REAGENT 2 DIL PUMP REAGENT 3 F1: PRIME WATER F2: SAMPLER F3: REAGENT 1 F4: REAGENT 2 F5: IMT DIL PUMP F6: HM ... F7: REAGENT 3 Dimension EXL 2 Dime Dimension EXL 200 SIEMENS 778849.001-B Cleaning the R2 Reagent Probe WP2 01/18/23 12:03 HM PUMP PRIME HM REAGENT WASH CLEANER F1: HM WASH PUMP F2: REAG CLEANER F3: SAMP CLEANER F4: F5: F6: F7: F8: Priming pumps ... MAINTENANCE LOGS Table 11-5: Weekly/Monthly Maintenance Year: 20XX Clean R3 Operator Drain ** Initials Weekly/Monthly Maintenance Clean Clot Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthcare GmbH 2023 Health
- Dimension
- dimension
- Dimension System
- HM wash probe
- R2 Probe
- R2 probe
- weekly maintenance
- Dimension weekly maintenance
- hm wash probe
- hm
- HM