Dimension® EXL™ with LM Integrated Chemistry System and Dimension® EXL™ 200 Integrated Chemistry Systems - Replacing the LOCI® Vacuum Cup Video

This video provides instructions on how to replace the LOCI® vacuum cup on the Dimension® EXL™ 200 and Dimension® EXL™ with LM Integrated Chemistry Systems.

With the system in standby, enter the Loci diagnostic screen by selecting F7 Diagnostics, F1 Electromechanical, F2 Sample System, F7 Loci Module. Select F6 inspect to rotate the low site arm out above the HM wheel. Pull down on the vacuum cup to remove it. Press the new vacuum cup onto the arm in reverse motion, making sure it is attached firmly to the underside of the arm from the touch screen. Exit to the main operating screen by selecting the home key. Then. Select Reset on the keyboard. To. Return. The loci arm to. The loci chamber. Next, realign the loci arm. Refer to the Dimension System Operator's Guide for specific details on this process. After the procedure is complete, run QC on all loci methods if documenting using the paper logs from the Dimension System Operator's Guide. Document the maintenance in the weekly monthly maintenance log sheet or instrument log sheet. Or, if documenting using the electronic logs in the Dimension system software, document the maintenance in the instrument log screen.

212 10 2467 100 0, 1111 Replacing the LOCI® Vacuum Cup on the Integrated Chemistry Systems SIEMENS . Standby Sampler Idle IMT OK 09/14/22 21:23 Dimension Exit labmanager Enter en OPERATING MENU STAT Status PLEASE SELECT AN OPTION USING THE FUNCTION KEYS: Sample Alert At any time, press HELP to get help, or press EXIT to exit a program. Supplies QC Alert Calib Alert To Shut down or Change Operator, press Sign Out or Alt+EXIT. Run Home F1: ENTER DATA F2: SAMPLE STATUS F3: TEST RESULTS F4: SYSTEM PREP F5: PROCESS CTRL F6: SYSTEM CONFIG F7: DIAGNOSTICS F8: SEGMENT STATUS 07/07/23 02:54 DIAGNOSTIC MENU Press EXIT to exit a program. F1: ELECTRO/MECH F2: FILM LOAD F3: ALIGNMENTS F4: F5: CHECK WINDOWS F6: ERROR LOG F7: KINETICS F8: OPEN CHANNELS Diagnostics ELECTRO/MECHANICAL DIAGNOSTICS SELECT PROGRAM USING FUNCTION KEYS F1: REAGENT F2: SAMPLE SYSTEM F3: IMT F4: CUVETTE F5: PHOTOMETER F6: WATER WORKS F7: POWER SYSTEM F8: ULTRASONICS SAMPLE SYSTEM DIAGNOSTICS MENU F1: Level Sense F2: Sampler F3: Samp Wheel F4: Clot Check F5: HM Module F6: F7: LOCI Module F8: LOCI MODULE DIAGNOSTIC: CYCLE: 1 ( 0) Position Encoder Bump Pressure: 0.0000 ROTATION [ 0] VERTICAL DETECTOR READ Temperature 0.00 PM 0.00 chamber Preread 0 PM 0.0 contamination 0.0 illumination Read init Read PM 0 total Read VF Gain 0 VF 0.000 relative Shutter 0.000 open 0.000 close 0.000 stability F1: INITIALIZE F2: PLUS F3: MINUS F5: HOME F6: INSPECT F7: Statistics F8: CYCLE HM SIEMENS 07/07/23 02:23 Position Encoder 0} FPSA F1 F2 F4 F7 F8 Delete Alarm Caps Lock Control Reset LYTES Tab MAINTENANCE LOGS Year: 20XX Clean R3 Operator Disclaimer Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively "Operator Manual") issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively "Functionality") may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Siemens Healthcare GmbH 2023 Healthineers

  • Dimension
  • dimension
  • exl
  • EXL
  • lm
  • LM
  • LOCI
  • loci
  • losigh
  • maintenance
  • dimension maintenance
  • Dimension System Maintenance
  • vacuum cup