Creating Calibration Orders Manually on the Atellica® Solution Job Aid Version 1.26 and Above
This job aid provides instructions on how to create calibration orders manually on the Atellica® Solution.
Creating Calibration Orders Manually on the Atellica® Solution Job Aid SIEMENS Healthineers Creating Calibration Orders Manually – Reagent Overview ‹ Â Reagent > Reagent Overview 6 abManager Screen 2 3 E All No Filters Reagent Details 1. Navigate to Reagent Overview 00260 00160 O Analyzer Reagent Inventory OBS Sequence Cal Status Cal Exp Cal Type In Use QC Status Lot Cal Eligibility Name Aic_3 120d 10h 00260 / 00160 120 18 h Pack 122312 Lot & Pack Aic 3 Paired Pack CH-O Filters 2. Use the Filters to select to view reagent information for 4 Lot Counts Remaining O CH-01 ALP 2c 250 120d 10h 00261 / -- Pack 421551 Lot & Pack 122312 0 a selected analyzer Pack Location Onboard Stability 0 CH-01 10260 / 10160 12d 18 h Pack 532316 Lot & Pack CH Reagent Compartment 1 120d 10h ALT 400 120d 10h Position Lot Expiration IM-01 HBsl 10 10 d 8h 10261 12d 18 h Pack 531321 Pack 10 6/23/2021 111:59 PM 3. To sort, select a column heading Calibration Expiration IM-01 VB12 Empty 6/23/2021 111:59 PM 80 Od 12h 10262 12d 18 h Lot 631322 Pack Sequence Number IM-01 CKMB 200 120d 10h 10263 2d 18 h Lot 721524 Pack 00260 ➢ Cal Status: to bring reagents that have an invalid IM-01 aHAVM 12d 18 h 721526 Pack Well 1 180 10d 10h 10264 Lot calibration or have not been calibrated to the top Test Count patusored O IM-01 aHCV 180 100d 2h 10265 12d 18 h Pack 781526 Lot & Pack O Time Punctured of the screen 0 IM-01 BNP 80 100d 2h 10266 12d 18 h Pack 621523 Lot & Pack 3/23/2021 10:20 AM O IM-01 hCG 0 Oh 10267 12d 18 h Pack 621524 Lot & Pack Calibrate Lot Calibrate Pack ➢ Cal Interval: to bring the reagents that have a 0 CI-01 eHIV 0 100d 2h 10268 12d 18 h Pack 621524 Lot & Pack Well 2 calibration that will expire soon to the top of the Test Count Status CI-01 AFP 160 100d 2h 10268 2d 18 h Pack 621524 Lot & Pack O Punctured 0 1-01 AAG 90 100d 2h 10269 / 11121 12d 18 h Pack 621524 Lot & Pack Calibrate Lot Calibrate Pack screen 0 01-01 CREA 44 100d 2h 10270 2d 18 h Pack 621524 Lot & Pack Submit 4. Select the desired reagent pack Group by Reagent Disable Reagent Pack Disable Reagent Lot Unload * This procedure is continued on the next slide 1 Template Effective Date: 10 Aug 2021 | HILS 2223 Effective Date: 11 FEB 2022| HOOD05162003253245 Unrestricted © Siemens Healthineers, 2022 Creating Calibration Orders Manually on the Atellica® Solution Job Aid SIEMENS Healthineers Creating Calibration Orders Manually – Reagent Overview ‹ Â abManager Reagent > Reagent Overview 6 Screen (continued) 5 EID AII No Filters Reagent Details In the Reagent Details area of the screen, select the Tx 00260 00160 5. O Analyzer Reagent Inventory OBS Sequence Cal Status Cal Exp Cal Type In Use QC Status Lot Cal Eligibility Name Paired Pack appropriate button to order either a pack calibration or Aic 3 CH-01 Aic_3 120d 10h 00260 / 00160 2d 18 h Pack 122312 Lot & Pack Filter Lot Counts Remaining 421551 Lot & Pack 122312 0 a lot calibration 0 CH-01 ALP 2c 250 120d 10h 00261 / -- Pack Pack Location 120d 10h CH-01 CH Reagent Compartment 1 Onboard Stability ALT 120d 10h 10260 / 10160 12d 18 h Pack 532316 Lot & Pack 400 Position Lot Expiration 6/23/2021 111:59 PM ➢ If there are multiple wells in the pack, there will be IM-01 10 HBsl 10 10 d 8h 10261 12d 18 h Pack 531321 Pack Calibration Expiration IM-01 Od 12h 10262 12d 18 h Lot 631322 Pack Empty 6/23/2021 111:59 PM separate options to calibrate each well VB12 80 Sequence Number 00260 IM-01 CKMB 200 120d 10h 10263 2d 18 h Lot 721524 Pack ➢ The options available will depend on what type of IM-01 aHAVM 180 10d 10h 10264 12d 18 h Lot 721526 Pack Well 1 Test Count patusred 0 O IM-01 aHCV 180 100d 2h 10265 12d 18 h Pack 781526 Lot & Pack calibration the pack is eligible for (pack/lot) Time Punctured 0 IM-01 BNP 80 100d 2h 10266 12d 18 h Pack 621523 Lot & Pack 3/23/2021 10:20 AM Oh 10267 12d 18 h Pack 621524 Lot & Pack Calibrate Lot Calibrate Pack 6. If the reagent includes multiple assays, the system will O IM-01 hCG 0 0 CI-01 eHIV 0 100d 2h 10268 12d 18 h Pack 621524 Lot & Pack Well 2 display a menu listing the analytes. Select the Test Count Status CI-01 AFP 160 100d 2h 10268 2d 18 h Pack 621524 Lot & Pack O Punctured appropriate assay name in the displayed list 0 01-01 AAG 90 100d 2h 10269 / 11121 12d 18 h Pack 621524 Lot & Pack Calibrate Lot Calibrate Pack 0 CI-01 CREA 44 100d 2h 10270 2d 18 h Pack 621524 Lot & Pack ➢ Submit For example, if the operator selects HBsII the Group by Reagent Disable Reagent Pack Disable Reagent Lot Unload displayed menu includes HBsII and CONF. The operator selects HBsII or CONF as appropriate 7. Select Yes in the confirmation window NOTE: When calibrating reagent packs or wells directly from Reagent Overview, the system requires 1 active calibrator definition. If more than 1 calibrator definition exists on the system for a reagent, the system defaults to using the oldest calibrator material and definition to calibrate the reagent. 2 Effective Date: 11 FEB 2022| HOOD05162003253245 Unrestricted © Siemens Healthineers, 2022 Creating Calibration Orders Manually on the Atellica® Solution Job Aid SIEMENS Healthineers Creating Calibration Orders Manually – Calibration Orders Screen Orders > Calibration Orders 8 SH1-SCI1 LabManager 1. Navigate to Calibration Orders 5 6 Select B Select Reagents 2 Assay Assay Reagent Pack ID Inventory Cal Exp In Use Type to Run Lot Cal Exp 2. Select by Assay or Material O Material Alb 323456-11112:well 1 1 Lo Oh 3 Select Analyzer * 3. Select an analyzer CH-01 Alb 323456-11112:well 2 100 Lot Oh Select Assay(s) for Calibration 4. Select the assay/material(s) as appropriate Q Search Select All 5. Select the reagent/material(s) as appropriate 4 6AM AAG Alb ALT 6. If desired, change the type of calibration Albp Amylas Select Material (Lot/Pack/C0 Adjust)* Amp300 Amp1000 Bnz 200 Bnz300 Material Name Material Status Lot Material ID Lot Expiration Assay 7. In the Select Material area of the screen, you can expand the C3 Ca CHEMCAL Needed EC8989 77 12/4/2021 Alb Carb CI Level 1 EC8989 12/4/2021 rows to view the material status for each calibrator level 7 Onboard Chol2 DHDL DMX ETOH . Level 2 Onboard EC8989 12/4/2021 8. Select Next GluH 3 Level 3 Needed EC8989 12/4/2021 Lac Op 200 . Level 4 Onboard EC8989 12/4/2021 PROC TACR . Level 5 Onboard EC8989 12/4/2021 UPro Xtc300 * This procedure is continued on the next slide Related Worklist Callbration Due Callbration Results Callbrator Definitions 8 Next Clear *NOTE: The calibration type listed is the default. You don’t have to specify lot or pack unless you want to. The system will always do a lot calibration over a pack calibration if possible. 3 Effective Date: 11 FEB 2022| HOOD05162003253245 Unrestricted © Siemens Healthineers, 2022 Creating Calibration Orders Manually on the Atellica® Solution Job Aid SIEMENS Healthineers Creating Calibration Orders Manually – Calibration Orders Screen Orders > Callbration Orders 8 SH1-SCI1 LabManager (continued) Calibration Orders A B 9. Review the calibration orders for accuracy X Analyzer ID Assay Type Reagent Pack ID Cal Metrial Name Cal Lot QC Material Name Replicates X CH-01 Alb Lot 323456-11112:well 1 CHEMICAL EC8331 ALB_QUAL Replicates A. If there is QC material listed in the QC Material Name column, this is QC that will be automatically ordered with the calibration*. If the QC is stored onboard, it will automatically run when it is ordered B. If desired, change the number of replicates • This is the number of calibrator replicates. The number of replicates can only be increased from the number defined in the test definition, you cannot decrease the replicates below that number 10. Select Place Order 10 Related Worklist Callbration Due Calibration Results Callbrator Definitions Back Place Order Cancel *NOTE: In this case, the Order QC with Calibration setting is enabled on the Calibration Setup screen. 4 Effective Date: 11 FEB 2022| HOOD05162003253245 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Atellica, and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc. or its affiliates. All other trademarks and brands are the property of their respective owners. Product availability varies by country and is subject to varying regulatory requirements. Please contact your local representative for availability. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 5 91052 Erlangen, Germany Effective Date: 11 FEB 2022| HOOD05162003253245 Telephone: +49 9131 84-0 Unrestricted © Siemens Healthineers, 2022 siemens-healthineers.com
- atelica
- attellica
- attelica